Abstract
We have examined the effects of optimal and sub-optimal cup position and different types of polyethylene on the simulated polyethylene wear. Finite Element (FE) Analysis was used, and ultra high molecular weight (UHMWPE), highly cross-linked (HCL), and E vitamin stabilized (E-poly) polyethylene were examined. We found structural stress parameters to be three times higher for a steep cup compared with a cup in optimal position. There were no differences in predicted polyethylene wear for the three examined polyethylene liners.
We further evaluated the in vivo polyethylene wear of total hip arthroplasties in 14 young patients with UHMWPE liners at mean follow-up of 5 years using model-based radiostereometric analysis (RSA). A three-dimensional estimation of polyethylene wear (femoral head penetration into the liner) located the main portion of wear in the posterior-superior part of the liner. The measured clinical wear was correlated to structural stress and strain parameters in an extended FE model of the metal-on-UHMWPE bearing coupling. All structural parameters with the extended simulation model were located in the posterior-superior part of the liner and measures were within 1 standard deviation of the clinically measured RSA wear. Stain parameters gave the best correlation.
Access this chapter
Tax calculation will be finalised at checkout
Purchases are for personal use only
Abbreviations
- 2D:
-
two-dimensional
- 3D:
-
three-dimensional
- CoCr:
-
cobalt-chromium
- DSS:
-
directional shear stress
- EGS models:
-
elementary geometric shape models
- E-poly:
-
E-vitamin stabilized
- FE:
-
finite element
- HCL:
-
highly cross-linked
- PE:
-
polyethylene
- RMSE:
-
root mean square error
- RSA:
-
radiostereometric analysis
- SD:
-
standard deviation
- SI:
-
stress intensity
- THA:
-
total hip arthroplasty
- UHMWPE:
-
ultra high molecular weight polyethylene
- VMS:
-
Von Mises stress
References
Pappas, C.: Material properties of Ringlock cups, email communication (2009). Accessed 24 Apr 2009
Anonymous authors: The finite element method. http://www://en.wikipedia.org/wiki/Finite_element_method (2010). Accessed 31 Oct 2010
Barbour, P.S., Barton, D.C., Fisher, J.: The influence of stress conditions on the wear of UHMWPE for total joint replacements. J. Mater. Sci. Mater. Med. 8(10), 603–611 (1997)
Borlin, N., Thien, T., Karrholm, J.: The precision of radiostereometric measurements. Manual vs. digital measurements. J. Biomech. 35(1), 69–79 (2002)
Bragdon, C.R., Estok, D.M., Malchau, H., Karrholm, J., Yuan, X., Bourne, R., Veldhoven, J., Harris, W.H.: Comparison of two digital radiostereometric analysis methods in the determination of femoral head penetration in a total hip replacement phantom. J. Orthop. Res. 22(3), 659–664 (2004)
Bragdon, C.R., O’Connor, D.O., Lowenstein, J.D., Jasty, M., Syniuta, W.D.: The importance of multidirectional motion on the wear of polyethylene. Proc. Inst. Mech. Eng. H 210(3), 157–165 (1996)
Brown, T.D., Lundberg, H.J., Pedersen, D.R., Callaghan, J.J.: Nicolas Andry Award: clinical biomechanics of third body acceleration of total hip wear. Clin. Orthop. Relat. Res. 467(7), 1885–1897 (2009)
Brown, T.D., Stewart, K.J., Nieman, J.C., Pedersen, D.R., Callaghan, J.J.: Local head roughening as a factor contributing to variability of total hip wear: a finite element analysis. J. Biomech. Eng. 124(6), 691–698 (2002)
Cattaneo, P.M., Dalstra, M., Melsen, B.: Orthodontic tooth movement of single and multi-rooted teeth: strains and stress in the periodontium assessed by FE-analyses. 1(1), 7. 2nd Annual Meeting of the Danish Biomechanical Society, 30 Oct 2010
Cohen, J.: Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head. J. Bone Joint Surg. Br. 86(1), 148 (2004)
Dowd, J.E., Sychterz, C.J., Young, A.M., Engh, C.A.: Characterization of long-term femoral-head-penetration rates. Association with and prediction of osteolysis. J. Bone Joint Surg. Am. 82-A(8), 1102–1107 (2000)
Fisher, J., Jin, Z., Tipper, J., Stone, M., Ingham, E.: Tribology of alternative bearings. Clin. Orthop. Relat. Res. 453, 25–34 (2006)
Hernigou, P., Bahrami, T.: Zirconia and alumina ceramics in comparison with stainless-steel heads. Polyethylene wear after a minimum ten-year follow-up. J. Bone Joint Surg. Br. 85(4), 504–509 (2003)
Huddleston, J.I., Harris, A.H., Atienza, C.A., Woolson, S.T.: Hylamer vs conventional polyethylene in primary total hip arthroplasty: a long-term case-control study of wear rates and osteolysis. J. Arthroplasty 25(2), 203–207 (2009)
Hurschler, C., Seehaus, F., Emmerich, J., Kaptein, B.L., Windhagen, H.: Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting. J. Arthroplasty 24(4), 594–606 (2009)
Jacobs, C.A., Christensen, C.P., Greenwald, A.S., McKellop, H.: Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty. J. Bone Joint Surg. Am. 89(12), 2779–2786 (2007)
Kang, L., Galvin, A.L., Brown, T.D., Fisher, J., Jin, Z.M.: Wear simulation of ultra-high molecular weight polyethylene hip implants by incorporating the effects of cross-shear and contact pressure. Proc. Inst. Mech. Eng. H 222(7), 1049–1064 (2008)
Kang, L., Galvin, A.L., Brown, T.D., Jin, Z., Fisher, J.: Quantification of the effect of cross-shear on the wear of conventional and highly cross-linked UHMWPE. J. Biomech. 41(2), 340–346 (2008)
Kang, L., Galvin, A.L., Fisher, J., Jin, Z.: Enhanced computational prediction of polyethylene wear in hip joints by incorporating cross-shear and contact pressure in additional to load and sliding distance: effect of head diameter. J. Biomech. 42(7), 912–918 (2009)
Kaptein, B.L., Valstar, E.R., Spoor, C.W., Stoel, B.C., Rozing, P.M.: Model-based RSA of a femoral hip stem using surface and geometrical shape models. Clin. Orthop. Relat. Res. 448, 92–97 (2006)
Kerner, J., Huiskes, R., van Lenthe, G.H., Weinans, H., van Rietbergen, B., Engh, C.A., Amis, A.A.: Correlation between pre-operative periprosthetic bone density and post-operative bone loss in THA can be explained by strain-adaptive remodelling. J. Biomech. 32(7), Â695–703 (1999)
Livingston, B.J., Chmell, M.J., Spector, M., Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone Joint Surg. Am. 79(10), 1529–1538 (1997)
Malchau, H., Herberts, P., Eisler, T., Garellick, G., Soderman, P.: The Swedish total hip replacement register. J. Bone Joint Surg. Am. 84-A(Suppl 2), 2–20 (2002)
Maxian, T.A., Brown, T.D., Pedersen, D.R., Callaghan, J.J.: The Frank Stinchfield Award. 3-Dimensional sliding/contact computational simulation of total hip wear. Clin. Orthop. Relat. Res. 333, 41–50 (1996)
Maxian, T.A., Brown, T.D., Pedersen, D.R., McKellop, H.A., Lu, B., Callaghan, J.J.: Finite element analysis of acetabular wear. Validation, and backing and fixation effects. Clin. Orthop. Relat. Res. 344, 111–117 (1997)
Ong, K.L., Rundell, S., Liepins, I., Laurent, R., Markel, D., Kurtz, S.M.: Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J. Orthop. Res. 27(11), 1467–1472 (2009)
Pedersen, D.R., Brown, T.D., Hillis, S.L., Callaghan, J.J.: Prediction of long-term polyethylene wear in total hip arthroplasty, based on early wear measurements made using digital image analysis. J. Orthop. Res. 16(5), 557–563 (1998)
Pedersen, D.R., Brown, T.D., Maxian, T.A., Callaghan, J.J.: Temporal and spatial distributions of directional counterface motion at the acetabular bearing surface in total hip arthroplasty. Iowa Orthop. J. 18, 43–53 (1998)
Shaver, S.M., Brown, T.D., Hillis, S.L., Callaghan, J.J.: Digital edge-detection measureÂment of polyethylene wear after total hip arthroplasty. J. Bone Joint Surg. Am. 79(5), 690–700 (1997)
Sochart, D.H.: Relationship of acetabular wear to osteolysis and loosening in total hip arthroplasty. Clin. Orthop. Relat. Res. 363, 135–150 (1999)
Sochart, D.H.: True wear rates as predictors of risk for osteolysis after total hip arthroplasty. J. Bone Joint Surg. Am. 83-A(8), 1277 (2001)
Stilling, M.: Polyethylene wear analysis. Experimental and clinical studies in total hip replacement. Acta Orthop. 80(337), 1–74 (2009)
Stilling, M., Larsen, K., Andersen, N.T., Balle, K., Kold, S.X., Rahbek, O.: The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. Acta Orthop. 81(5), 570–578 (2010)
Stilling, M., Nielsen, K.A., Soballe, K., Rahbek, O.: Clinical comparison of polyethylene wear with zirconia or cobalt-chromium femoral heads. Clin. Orthop. Relat. Res. 467(10), 2644–2650 (2009)
Stilling, M., Rahbek, O., Soballe, K.: Inferior survival of hydroxyapatite versus titanium-coated cups at 15 years. Clin. Orthop. Relat. Res. 467(11), 2872–2879 (2009)
Stilling, M., Soballe, K., Andersen, N.T., Larsen, K., Rahbek, O.: Analysis of polyethylene wear in plain radiographs. The number of radiographs influences the results. Acta Orthop. 80(6), 675–682 (2009)
Sychterz, C.J., Engh Jr., C.A., Swope, S.W., McNulty, D.E., Engh, C.A.: Analysis of prosthetic femoral heads retrieved at autopsy. Clin. Orthop. Relat. Res. 358, 223–234 (1999)
Sychterz, C.J., Engh Jr., C.A., Yang, A., Engh, C.A.: Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called Âbedding-in. J. Bone Joint Surg. Am. 81(6), 821–830 (1999)
Valstar, E.R., de Jong, F.W., Vrooman, H.A., Rozing, P.M., Reiber, J.H.: Model-based roentgen stereophotogrammetry of orthopaedic implants. J. Biomech. 34(6), 715–722 (2001)
Valstar, E.R., Vrooman, H.A., Toksvig-Larsen, S., Ryd, L., Nelissen, R.G.: Digital automated RSA compared to manually operated RSA. J. Biomech. 33(12), 1593–1599 (2000)
van Lenthe, G.H., Willems, M.M., Verdonschot, N., Waal Malefijt, M.C., Huiskes, R.: Stemmed femoral knee prostheses: effects of prosthetic design and fixation on bone loss. Acta Orthop. Scand. 73(6), 630–637 (2002)
Vrooman, H.A., Valstar, E.R., Brand, G.J., Admiraal, D.R., Rozing, P.M., Reiber, J.H.: Fast and accurate automated measurements in digitized stereophotogrammetric radiographs. J. Biomech. 31(5), 491–498 (1998)
Wong, C., Gehrchen, P.M., Kiaer, T.: Can experimental data in humans verify the finite Âelement-based bone remodeling algorithm? Spine (Phila Pa 1976) 33(26), 2875–2880 (2008)
Wong, C., Gehrchen, P.M., Kiaer, T.: Can additional experimental data in humans verify the finite element-based bone remodeling algorithm? Open Spine J. 2, 12–16 (2010)
Wroblewski, B.M.: Direction and rate of socket wear in Charnley low-friction arthroplasty. J. Bone Joint Surg. Br. 67(5), 757–761 (1985)
Yamaguchi, M., Bauer, T.W., Hashimoto, Y.: Three-dimensional analysis of multiple wear vectors in retrieved acetabular cups. J. Bone Joint Surg. Am. 79(10), 1539–1544 (1997)
Yamaguchi, M., Hashimoto, Y., Akisue, T., Bauer, T.W.: Polyethylene wear vector in vivo: a three-dimensional analysis using retrieved acetabular components and radiographs. J. Orthop. Res. 17(5), 695–702 (1999)
Author information
Authors and Affiliations
Corresponding author
Editor information
Editors and Affiliations
Rights and permissions
Copyright information
© 2011 Springer Berlin Heidelberg
About this chapter
Cite this chapter
Wong, C., Stilling, M. (2011). Polyethylene Wear in Total Hip Arthroplasty for Suboptimal Acetabular Cup Positions and for Different Polyethylene Types: Experimental Evaluation of Wear Simulation by Finite Element Analysis Using Clinical Radiostereometric Measurements. In: Knahr, K. (eds) Tribology in Total Hip Arthroplasty. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-19429-0_12
Download citation
DOI: https://doi.org/10.1007/978-3-642-19429-0_12
Published:
Publisher Name: Springer, Berlin, Heidelberg
Print ISBN: 978-3-642-19428-3
Online ISBN: 978-3-642-19429-0
eBook Packages: MedicineMedicine (R0)