Polyethylene Wear in Total Hip Arthroplasty for Suboptimal Acetabular Cup Positions and for Different Polyethylene Types: Experimental Evaluation of Wear Simulation by Finite Element Analysis Using Clinical Radiostereometric Measurements

  • Chapter
  • First Online:
Tribology in Total Hip Arthroplasty
  • 1144 Accesses

Abstract

We have examined the effects of optimal and sub-optimal cup position and different types of polyethylene on the simulated polyethylene wear. Finite Element (FE) Analysis was used, and ultra high molecular weight (UHMWPE), highly cross-linked (HCL), and E vitamin stabilized (E-poly) polyethylene were examined. We found structural stress parameters to be three times higher for a steep cup compared with a cup in optimal position. There were no differences in predicted polyethylene wear for the three examined polyethylene liners.

We further evaluated the in vivo polyethylene wear of total hip arthroplasties in 14 young patients with UHMWPE liners at mean follow-up of 5 years using model-based radiostereometric analysis (RSA). A three-dimensional estimation of polyethylene wear (femoral head penetration into the liner) located the main portion of wear in the posterior-superior part of the liner. The measured clinical wear was correlated to structural stress and strain parameters in an extended FE model of the metal-on-UHMWPE bearing coupling. All structural parameters with the extended simulation model were located in the posterior-superior part of the liner and measures were within 1 standard deviation of the clinically measured RSA wear. Stain parameters gave the best correlation.

This is a preview of subscription content, log in via an institution to check access.

Access this chapter

Chapter
USD 29.95
Price excludes VAT (USA)
  • Available as PDF
  • Read on any device
  • Instant download
  • Own it forever
eBook
USD 39.99
Price excludes VAT (USA)
  • Available as EPUB and PDF
  • Read on any device
  • Instant download
  • Own it forever
Softcover Book
USD 54.99
Price excludes VAT (USA)
  • Compact, lightweight edition
  • Dispatched in 3 to 5 business days
  • Free shipping worldwide - see info

Tax calculation will be finalised at checkout

Purchases are for personal use only

Institutional subscriptions

Abbreviations

2D:

two-dimensional

3D:

three-dimensional

CoCr:

cobalt-chromium

DSS:

directional shear stress

EGS models:

elementary geometric shape models

E-poly:

E-vitamin stabilized

FE:

finite element

HCL:

highly cross-linked

PE:

polyethylene

RMSE:

root mean square error

RSA:

radiostereometric analysis

SD:

standard deviation

SI:

stress intensity

THA:

total hip arthroplasty

UHMWPE:

ultra high molecular weight polyethylene

VMS:

Von Mises stress

References

  1. Pappas, C.: Material properties of Ringlock cups, email communication (2009). Accessed 24 Apr 2009

    Google Scholar 

  2. Anonymous authors: The finite element method. http://www://en.wikipedia.org/wiki/Finite_element_method (2010). Accessed 31 Oct 2010

  3. Barbour, P.S., Barton, D.C., Fisher, J.: The influence of stress conditions on the wear of UHMWPE for total joint replacements. J. Mater. Sci. Mater. Med. 8(10), 603–611 (1997)

    Article  PubMed  CAS  Google Scholar 

  4. Borlin, N., Thien, T., Karrholm, J.: The precision of radiostereometric measurements. Manual vs. digital measurements. J. Biomech. 35(1), 69–79 (2002)

    Article  PubMed  Google Scholar 

  5. Bragdon, C.R., Estok, D.M., Malchau, H., Karrholm, J., Yuan, X., Bourne, R., Veldhoven, J., Harris, W.H.: Comparison of two digital radiostereometric analysis methods in the determination of femoral head penetration in a total hip replacement phantom. J. Orthop. Res. 22(3), 659–664 (2004)

    Article  PubMed  Google Scholar 

  6. Bragdon, C.R., O’Connor, D.O., Lowenstein, J.D., Jasty, M., Syniuta, W.D.: The importance of multidirectional motion on the wear of polyethylene. Proc. Inst. Mech. Eng. H 210(3), 157–165 (1996)

    PubMed  CAS  Google Scholar 

  7. Brown, T.D., Lundberg, H.J., Pedersen, D.R., Callaghan, J.J.: Nicolas Andry Award: clinical biomechanics of third body acceleration of total hip wear. Clin. Orthop. Relat. Res. 467(7), 1885–1897 (2009)

    Article  PubMed  Google Scholar 

  8. Brown, T.D., Stewart, K.J., Nieman, J.C., Pedersen, D.R., Callaghan, J.J.: Local head roughening as a factor contributing to variability of total hip wear: a finite element analysis. J. Biomech. Eng. 124(6), 691–698 (2002)

    Article  PubMed  Google Scholar 

  9. Cattaneo, P.M., Dalstra, M., Melsen, B.: Orthodontic tooth movement of single and multi-rooted teeth: strains and stress in the periodontium assessed by FE-analyses. 1(1), 7. 2nd Annual Meeting of the Danish Biomechanical Society, 30 Oct 2010

    Google Scholar 

  10. Cohen, J.: Catastrophic failure of the Elite Plus total hip replacement, with a Hylamer acetabulum and Zirconia ceramic femoral head. J. Bone Joint Surg. Br. 86(1), 148 (2004)

    PubMed  CAS  Google Scholar 

  11. Dowd, J.E., Sychterz, C.J., Young, A.M., Engh, C.A.: Characterization of long-term femoral-head-penetration rates. Association with and prediction of osteolysis. J. Bone Joint Surg. Am. 82-A(8), 1102–1107 (2000)

    PubMed  CAS  Google Scholar 

  12. Fisher, J., Jin, Z., Tipper, J., Stone, M., Ingham, E.: Tribology of alternative bearings. Clin. Orthop. Relat. Res. 453, 25–34 (2006)

    Article  PubMed  Google Scholar 

  13. Hernigou, P., Bahrami, T.: Zirconia and alumina ceramics in comparison with stainless-steel heads. Polyethylene wear after a minimum ten-year follow-up. J. Bone Joint Surg. Br. 85(4), 504–509 (2003)

    Article  PubMed  CAS  Google Scholar 

  14. Huddleston, J.I., Harris, A.H., Atienza, C.A., Woolson, S.T.: Hylamer vs conventional polyethylene in primary total hip arthroplasty: a long-term case-control study of wear rates and osteolysis. J. Arthroplasty 25(2), 203–207 (2009)

    Article  PubMed  Google Scholar 

  15. Hurschler, C., Seehaus, F., Emmerich, J., Kaptein, B.L., Windhagen, H.: Comparison of the model-based and marker-based roentgen stereophotogrammetry methods in a typical clinical setting. J. Arthroplasty 24(4), 594–606 (2009)

    Article  PubMed  Google Scholar 

  16. Jacobs, C.A., Christensen, C.P., Greenwald, A.S., McKellop, H.: Clinical performance of highly cross-linked polyethylenes in total hip arthroplasty. J. Bone Joint Surg. Am. 89(12), 2779–2786 (2007)

    Article  PubMed  Google Scholar 

  17. Kang, L., Galvin, A.L., Brown, T.D., Fisher, J., Jin, Z.M.: Wear simulation of ultra-high molecular weight polyethylene hip implants by incorporating the effects of cross-shear and contact pressure. Proc. Inst. Mech. Eng. H 222(7), 1049–1064 (2008)

    Article  PubMed  CAS  Google Scholar 

  18. Kang, L., Galvin, A.L., Brown, T.D., Jin, Z., Fisher, J.: Quantification of the effect of cross-shear on the wear of conventional and highly cross-linked UHMWPE. J. Biomech. 41(2), 340–346 (2008)

    Article  PubMed  Google Scholar 

  19. Kang, L., Galvin, A.L., Fisher, J., Jin, Z.: Enhanced computational prediction of polyethylene wear in hip joints by incorporating cross-shear and contact pressure in additional to load and sliding distance: effect of head diameter. J. Biomech. 42(7), 912–918 (2009)

    Article  PubMed  Google Scholar 

  20. Kaptein, B.L., Valstar, E.R., Spoor, C.W., Stoel, B.C., Rozing, P.M.: Model-based RSA of a femoral hip stem using surface and geometrical shape models. Clin. Orthop. Relat. Res. 448, 92–97 (2006)

    Article  PubMed  Google Scholar 

  21. Kerner, J., Huiskes, R., van Lenthe, G.H., Weinans, H., van Rietbergen, B., Engh, C.A., Amis, A.A.: Correlation between pre-operative periprosthetic bone density and post-operative bone loss in THA can be explained by strain-adaptive remodelling. J. Biomech. 32(7), ­695–703 (1999)

    Article  PubMed  CAS  Google Scholar 

  22. Livingston, B.J., Chmell, M.J., Spector, M., Poss, R.: Complications of total hip arthroplasty associated with the use of an acetabular component with a Hylamer liner. J. Bone Joint Surg. Am. 79(10), 1529–1538 (1997)

    PubMed  CAS  Google Scholar 

  23. Malchau, H., Herberts, P., Eisler, T., Garellick, G., Soderman, P.: The Swedish total hip replacement register. J. Bone Joint Surg. Am. 84-A(Suppl 2), 2–20 (2002)

    PubMed  Google Scholar 

  24. Maxian, T.A., Brown, T.D., Pedersen, D.R., Callaghan, J.J.: The Frank Stinchfield Award. 3-Dimensional sliding/contact computational simulation of total hip wear. Clin. Orthop. Relat. Res. 333, 41–50 (1996)

    Article  PubMed  Google Scholar 

  25. Maxian, T.A., Brown, T.D., Pedersen, D.R., McKellop, H.A., Lu, B., Callaghan, J.J.: Finite element analysis of acetabular wear. Validation, and backing and fixation effects. Clin. Orthop. Relat. Res. 344, 111–117 (1997)

    Article  PubMed  Google Scholar 

  26. Ong, K.L., Rundell, S., Liepins, I., Laurent, R., Markel, D., Kurtz, S.M.: Biomechanical modeling of acetabular component polyethylene stresses, fracture risk, and wear rate following press-fit implantation. J. Orthop. Res. 27(11), 1467–1472 (2009)

    Article  PubMed  Google Scholar 

  27. Pedersen, D.R., Brown, T.D., Hillis, S.L., Callaghan, J.J.: Prediction of long-term polyethylene wear in total hip arthroplasty, based on early wear measurements made using digital image analysis. J. Orthop. Res. 16(5), 557–563 (1998)

    Article  PubMed  CAS  Google Scholar 

  28. Pedersen, D.R., Brown, T.D., Maxian, T.A., Callaghan, J.J.: Temporal and spatial distributions of directional counterface motion at the acetabular bearing surface in total hip arthroplasty. Iowa Orthop. J. 18, 43–53 (1998)

    PubMed  CAS  Google Scholar 

  29. Shaver, S.M., Brown, T.D., Hillis, S.L., Callaghan, J.J.: Digital edge-detection measure­ment of polyethylene wear after total hip arthroplasty. J. Bone Joint Surg. Am. 79(5), 690–700 (1997)

    PubMed  CAS  Google Scholar 

  30. Sochart, D.H.: Relationship of acetabular wear to osteolysis and loosening in total hip arthroplasty. Clin. Orthop. Relat. Res. 363, 135–150 (1999)

    Article  PubMed  Google Scholar 

  31. Sochart, D.H.: True wear rates as predictors of risk for osteolysis after total hip arthroplasty. J. Bone Joint Surg. Am. 83-A(8), 1277 (2001)

    PubMed  CAS  Google Scholar 

  32. Stilling, M.: Polyethylene wear analysis. Experimental and clinical studies in total hip replacement. Acta Orthop. 80(337), 1–74 (2009)

    Article  Google Scholar 

  33. Stilling, M., Larsen, K., Andersen, N.T., Balle, K., Kold, S.X., Rahbek, O.: The final follow-up plain radiograph is sufficient for clinical evaluation of polyethylene wear in total hip arthroplasty. Acta Orthop. 81(5), 570–578 (2010)

    Article  PubMed  Google Scholar 

  34. Stilling, M., Nielsen, K.A., Soballe, K., Rahbek, O.: Clinical comparison of polyethylene wear with zirconia or cobalt-chromium femoral heads. Clin. Orthop. Relat. Res. 467(10), 2644–2650 (2009)

    Article  PubMed  Google Scholar 

  35. Stilling, M., Rahbek, O., Soballe, K.: Inferior survival of hydroxyapatite versus titanium-coated cups at 15 years. Clin. Orthop. Relat. Res. 467(11), 2872–2879 (2009)

    Article  PubMed  Google Scholar 

  36. Stilling, M., Soballe, K., Andersen, N.T., Larsen, K., Rahbek, O.: Analysis of polyethylene wear in plain radiographs. The number of radiographs influences the results. Acta Orthop. 80(6), 675–682 (2009)

    Article  PubMed  Google Scholar 

  37. Sychterz, C.J., Engh Jr., C.A., Swope, S.W., McNulty, D.E., Engh, C.A.: Analysis of prosthetic femoral heads retrieved at autopsy. Clin. Orthop. Relat. Res. 358, 223–234 (1999)

    Article  PubMed  Google Scholar 

  38. Sychterz, C.J., Engh Jr., C.A., Yang, A., Engh, C.A.: Analysis of temporal wear patterns of porous-coated acetabular components: distinguishing between true wear and so-called ­bedding-in. J. Bone Joint Surg. Am. 81(6), 821–830 (1999)

    PubMed  CAS  Google Scholar 

  39. Valstar, E.R., de Jong, F.W., Vrooman, H.A., Rozing, P.M., Reiber, J.H.: Model-based roentgen stereophotogrammetry of orthopaedic implants. J. Biomech. 34(6), 715–722 (2001)

    Article  PubMed  CAS  Google Scholar 

  40. Valstar, E.R., Vrooman, H.A., Toksvig-Larsen, S., Ryd, L., Nelissen, R.G.: Digital automated RSA compared to manually operated RSA. J. Biomech. 33(12), 1593–1599 (2000)

    Article  PubMed  CAS  Google Scholar 

  41. van Lenthe, G.H., Willems, M.M., Verdonschot, N., Waal Malefijt, M.C., Huiskes, R.: Stemmed femoral knee prostheses: effects of prosthetic design and fixation on bone loss. Acta Orthop. Scand. 73(6), 630–637 (2002)

    Article  PubMed  Google Scholar 

  42. Vrooman, H.A., Valstar, E.R., Brand, G.J., Admiraal, D.R., Rozing, P.M., Reiber, J.H.: Fast and accurate automated measurements in digitized stereophotogrammetric radiographs. J. Biomech. 31(5), 491–498 (1998)

    Article  PubMed  CAS  Google Scholar 

  43. Wong, C., Gehrchen, P.M., Kiaer, T.: Can experimental data in humans verify the finite ­element-based bone remodeling algorithm? Spine (Phila Pa 1976) 33(26), 2875–2880 (2008)

    Article  Google Scholar 

  44. Wong, C., Gehrchen, P.M., Kiaer, T.: Can additional experimental data in humans verify the finite element-based bone remodeling algorithm? Open Spine J. 2, 12–16 (2010)

    Article  Google Scholar 

  45. Wroblewski, B.M.: Direction and rate of socket wear in Charnley low-friction arthroplasty. J. Bone Joint Surg. Br. 67(5), 757–761 (1985)

    PubMed  CAS  Google Scholar 

  46. Yamaguchi, M., Bauer, T.W., Hashimoto, Y.: Three-dimensional analysis of multiple wear vectors in retrieved acetabular cups. J. Bone Joint Surg. Am. 79(10), 1539–1544 (1997)

    PubMed  CAS  Google Scholar 

  47. Yamaguchi, M., Hashimoto, Y., Akisue, T., Bauer, T.W.: Polyethylene wear vector in vivo: a three-dimensional analysis using retrieved acetabular components and radiographs. J. Orthop. Res. 17(5), 695–702 (1999)

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Christian Wong .

Editor information

Editors and Affiliations

Rights and permissions

Reprints and permissions

Copyright information

© 2011 Springer Berlin Heidelberg

About this chapter

Cite this chapter

Wong, C., Stilling, M. (2011). Polyethylene Wear in Total Hip Arthroplasty for Suboptimal Acetabular Cup Positions and for Different Polyethylene Types: Experimental Evaluation of Wear Simulation by Finite Element Analysis Using Clinical Radiostereometric Measurements. In: Knahr, K. (eds) Tribology in Total Hip Arthroplasty. Springer, Berlin, Heidelberg. https://doi.org/10.1007/978-3-642-19429-0_12

Download citation

  • DOI: https://doi.org/10.1007/978-3-642-19429-0_12

  • Published:

  • Publisher Name: Springer, Berlin, Heidelberg

  • Print ISBN: 978-3-642-19428-3

  • Online ISBN: 978-3-642-19429-0

  • eBook Packages: MedicineMedicine (R0)

Publish with us

Policies and ethics