Keywords

These keywords were added by machine and not by the authors. This process is experimental and the keywords may be updated as the learning algorithm improves.

1 Introduction

Near-infrared spectroscopy (NIRS) is a non-invasive clinical monitoring method which provides real-time information on the oxygenation and haemodynamics of tissue. It is based on the absorption of near-infrared light by haemoglobin; as the absorption properties of haemoglobin change with its oxygenation status, measuring the transmitted/reflected light from tissue informs on its oxygenation state of tissue. Using NIRS to monitor absolute tissue oxygenation in the brain is clinically desirable as it can inform on the balance between cerebral blood flow and oxygen supply. NIRS can measure brain tissue oxygen saturation StO2, also known as Tissue Oxygenation Index (TOI) , which is the ratio of the concentration of oxyhaemoglobin HbO2 and total haemoglobin HbT = HbO2 + HHb, where HHb is deoxyhaemoglobin. One of the methods to measure absolute tissue oxygenation is spatially resolved spectroscopy (SRS). SRS instruments are able to measure the change in light attenuation A with distance ρ by using two or more source-detector separations. A scaled absorption coefficient a is calculated using the slope of A:

$$ \mathrm{k}{\upmu}_{\mathrm{a}}=\frac{1}{3\left(1- h\lambda \right)}{\left(\ln (10)\frac{\partial A\left(\uplambda \right)}{\partial \rho }-\frac{2}{\uprho}\right)}^2 $$

where h is the gradient of the transport scattering coefficient μs with wavelength λ; μs '=k(1-hλ). Scaled concentration of haemoglobin k[HbO2] and k[HHb] are quantified; the scaling factor k (which is unknown from the measurement) cancels itself when expressing oxygenation in terms of TOI , i.e. TOI= k[HbO2]/k[HbT] [1].

2 Method

The data in this study were previously published by Bale et al. [2]. A broadband multi-distance continuous wave NIRS system, CYRIL, CYtochrome Research Instrument and appLication (source-detector distances: 30, 25, 20, 15 mm, 770–906 nm) [3], was used to monitor the right frontal hemisphere of neonates with hypoxic-ischaemic encephalopathy during the first 4 days of life. The patients were diagnosed with severe or moderate brain injury and all were treated with therapeutic hypothermia for 48 h. Data were recorded from 55 neonates and yielded 172 separate datasets (durations from 10 min to 18 h). To minimize computational burden, datasets were cut to be 600 s long; the time window was 200–799 s. The first 199 s of the measurement were excluded to avoid set-up errors. The “Standard” SRS method (described below) produced, in some cases, TOI values which were negative or above 100%. We now attempt to correct these values by investigating the SRS algorithm and how the input parameters influence TOI . The different approaches to resolve TOI were:

  • Approach 0, “Standard ”: This SRS approach resolved for only haemoglobin chromophores from a; the wavelength dependence of the scattering coefficient was h = 0.00063 mm−1 nm−1 [1, 4]; the entire broadband spectrum of the NIRS data (770–906 nm) was used; and all four detectors were used to quantify the attenuation slope.

  • Approach 1, “Water”: This approach additionally resolved for the water concentration from a, i.e. water was treated as a third chromophore like HbO2 and HHb. It does not obtain an absolute measure of the concentrations as k is unknown.

  • Approach 2, “Scattering”: The dependence of scattering on wavelength as described by h was modified. The “Standard” approach uses h = 0.00063 mm−1 nm−1 measured by Matcher et al. on adults [2] which is used in commercial systems [4]. In this approach, two other values are used. Highton et al. measured a value of h = 0.00085 mm−1 nm−1 in adults with traumatic brain injury [5]. Kurata et al. measured optical properties of the neonatal brain [6], a value of h = 0.00048 mm−1 nm−1 was calculated from the data for term infants using the method described by Matcher et al. [4].

  • Approach 3, “Distance”: The number and combination of source-detector separations was adjusted. A minimum of two source-detector distances are necessary for SRS; increasing their number leads to a more accurate approximation of the gradient of attenuation with distance. In this approach, the effect of using less detectors and applying various combinations of the detectors was studied; there were ten possible combinations.

  • Approach 4, “Wavelengths”: The “Standard” approach uses the full broadband spectrum; however, it is possible to measure TOI with a minimum of two wavelengths. The selected wavelengths were 775, 810, 850, 905 nm or 775, 810, 850 nm which were chosen because they are used in commercial systems; NIRO 300, NIRO 200 (Hamamatsu Photonics, Japan).

Each approach was investigated independently from the others; the other parameters remained as in the “Standard” SRS approach . Approaches “Scattering”, “Distance” and “Wavelength” have several sub-approaches. In total, there were 16 approaches resulting in 16 TOI values calculated from each dataset.

The resulting TOIs were averaged over the time window and compared to the “Standard” average TOI. A TOI value was considered to be “good” if it was between 0% and 100%.Footnote 1 The number of “good” TOIs was counted for all approaches and sub-approaches. Minimum and maximum TOIs were found for all “good” TOIs out of the 172 datasets. Datasets in which all 16 TOI values were “good” were further analysed using SPSS Statistics (IBM 2016). A Sign test was applied to evaluate differences between “Standard” and every other approach, significance threshold p < 0.05. Spearman’s rho (ρs) was calculated between the “Standard” TOIs and every other approach to assess their correlation. The combination of analysing the statistical difference between approaches and their correlation can give us information about their similarity, for example:

  • if ρs is low and there is a significant difference between the approaches, then these are not similar.

  • when ρs is high and there is a significant difference between the approaches, then these are highly associated but have different absolute TOI values.

  • if ρs is high and there is not a significant difference between the approaches then these approaches are similar.

3 Results

Table 1 shows the calculated TOI values with the percentage of “good” TOI values for each SRS approach. The number of datasets which give “good” TOIs changes per approach; the most successful one is “Wavelengths (775,810,850 nm)” with 165/172 “good” TOIs and the least successful is “Distance (15,20 mm)” with 95/172.

Table 1 Number of “good” TOI values from the 172 datasets calculated using the different approaches

For 48 out of 172 datasets all 16 approaches gave “good” TOI values. Figure 1 shows a boxplot of the median TOI values with Spearman’s Rho for each approach.

Fig. 1
figure 1

Boxplot of median TOIs for the 48 datasets in which all approaches give “good” TOI values. The full line is the median of the “Standard” approach and the dotted lines are its inter-quartile range. Circles are outliers. White boxes indicate significant difference (p < 0.05) from the “Standard” approach. Spearman’s Rho indicates the correlation between the “Standard” approach and other approaches

The Sign test showed that eight approaches gave significantly different results from the “Standard” (p < 0.05). Those were: “Water”, “Scattering” (both h values), “Distance” (20,15 mm; 30,15 mm; 25,20 mm) and “Wavelengths” (both sets). Spearman’s Rho was below 0.7 in six different instances, which suggests low association with the “Standard” approach: “Water” (ρs = 0.56), “Distance” (ρs (20,15 mm) = 0.65, ρs (30,20 mm) = 0.64, ρs (30,25 mm) = 0.36, ρs (30,25,20 mm) = 0.64) and “Wavelength” (ρs (775,810,850 nm) = 0.65).

4 Discussion

The aim of this study was to investigate the SRS algorithm using a multi-distance broadband NIRS device. The parameters investigated were: resolving for water, the scattering dependence on wavelength, source-detector distance combinations and wavelength selection. We have shown that TOI values vary significantly depending on the SRS approach.

To investigate the effect of changing the SRS approach, we quantified how many times a “good” TOI value (between 0 and 100%) was estimated per dataset (Table 1). Changing the approach sometimes changes the TOI value from negative/above 100% to what we call “good” TOI (between 0% and 100%). Even within “good” TOIs , the range of achievable values is broad. This is further explored by focusing on 48 datasets in which all 16 approaches result in “good” TOI – the median values significantly differ from “Standard” in the majority of approaches (Fig. 1). It is only the “Distance” approach, which gives similar results to “Standard” (p > 0.05 in 7/10 distance combinations), which suggests that the algorithm is robust to different source-detector combinations.

The combination of the Spearman’s Rho and the statistical significance assessments demonstrated low similarity between the “Standard” approach and: “Water”, “Distance” (20, 15 mm) and “Wavelength” (775,810,850 nm) approaches. However the approaches “Distance” (30,25,15 and 30,20,15 mm) had the most similar results to the “Standard” approach.

The approach “Wavelength ” shows that the selection of wavelength has a significant impact on the resulting TOI ; this is important as different commercial brain oximeters (that apply the SRS method) use different wavelengths. For example OxyMon (Artinis Medical Systems, the Netherlands) uses 765 and 855 nm, but NIRO-200NX (Hamamatsu, Japan) uses 735, 810 and 850 nm.

Significant shifts in TOI were caused by changing h; each value of h was acquired in a different patient group: 0.00063 mm−1 nm−1 in 7 healthy adults, 0.00085 mm−1 nm−1 in 21 brain injured adults and 0.00048 mm−1 nm−1 in 8 full-term neonates. This demonstrates the importance of having a subject-appropriate estimation of wavelength dependence of scattering. It has been shown that scattering properties change with age and pathology [6,7,8]; more investigations of the scattering dependence on wavelength and its behaviour during pathology are needed.

A broadband NIRS multi-distance device can provide additional information to improve the robustness of the SRS estimation of TOI , as it gives insight into the individual parameters in the algorithm and helps quantify their importance and impact. The limitation of this study is that the brain oxygenation of the studied patients is unknown and thus there is no physiological comparison.

The next steps include studying the effect on TOI by combining several approaches at once; and doing this using a dynamic blood phantom that allow us to control oxygenation (similar to a phantom described in [9]). We will do this while simultaneously measuring chromophore concentrations and oxygenation with a time-resolved system (TR) . TR systems measure absolute concentrations of chromophores by resolving absorption and scattering [10]. This will enable us to further study and evaluate the effect of each approach and possibly identifying an optimal combination of parameters.