Skip to main content
Top
Published in:

Open Access 01-12-2024 | Pelvic Ring Fractures | Research article

Experimental study of fractures of the posterior pelvic ring C1.1 using LC-II screws and internal fixation by plate

Authors: Yu Liu, Xukai Wang, Bin Tian, Hong Yao, Guangyao Liu

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2024

Login to get access

Abstract

Introduction

To compare the biomechanical outcomes of C1.1 posterior pelvic ring fractures treated with different numbers of LC-II screws and plate internal fixation.

Materials and methods

Nine adult preserved pelvic specimens were used. After being measured by bone densitometry, the specimens were randomly divided into 3 groups of 3 pelvic specimens each. The C1.1 pelvic fractures were generated by iliac osteotomies and anterior pelvic ring osteotomies. The fractures were fixed and randomly divided into three groups: (1) one LC-II screw, (2) two LC-II screws, and (3) two reconstruction plates. The anterior pelvic rings were all fixed with pubic branch screws. A biomechanical testing machine loaded all specimens vertically, recording the displacement and ultimate load of the specimens to quantify the stiffness.

Results

When loads up to 1000 N were applied, the displacement of the iliac bone was close between the one LC-II screw and the two LC-II screw group specimens; (P > 0.05); when loads such as 1000 N, 1200 N and 1400 N were applied, there was no significant difference in displacement between the plate fixation group and the two LC-II screw fixation group (P > 0.05), both of which were superior to the one LC-II screw fixation group (P < 0.05). There was no significant difference in compressive stiffness between the two LC-II screw fixation groups and the plate fixation group (P > 0.05), and both were superior to the one LC-II screw fixation group (P < 0.05) the ultimate load of both the two LC-II screws and the steel plate is greater than that of one LC-II screw (1.74 times the ultimate load of one LC-II screw for the two LC-II screws and 1.83 times the ultimate load of one LC-II screw for the steel plate).

Conclusion

For posterior pelvic ring fractures of type C1.1, placement of two LC-II screws provides comparable posterior pelvic ring stability compared to reconstructed plates. At loads below 1000 N, one LC-II screw fixation and two LC-II screws were comparable in restoring posterior pelvic ring stability.
Literature
1.
go back to reference Holtenius J, Bakhshayesh P, Enocson A. The pelvic fracture - Indicator of injury severity or lethal fracture? Injury. 2018;49(8):1568–71.CrossRefPubMed Holtenius J, Bakhshayesh P, Enocson A. The pelvic fracture - Indicator of injury severity or lethal fracture? Injury. 2018;49(8):1568–71.CrossRefPubMed
2.
go back to reference Liu ZB, Zou GY. Application of internal fixator system for anterior pelvic ring with simultaneous application of sacroiliac screw internal fixation of the posterior pelvic ring in Tile C-type unstable pelvic fractures. Vojnosanit Pregl. 2024;81(8):467–73.CrossRef Liu ZB, Zou GY. Application of internal fixator system for anterior pelvic ring with simultaneous application of sacroiliac screw internal fixation of the posterior pelvic ring in Tile C-type unstable pelvic fractures. Vojnosanit Pregl. 2024;81(8):467–73.CrossRef
3.
go back to reference Incagnoli P, Puidupin A, Ausset S, Beregi JP, Bessereau J, Bobbia X, Brun J, Brunel E, Buléon C, Choukroun J, et al. Early management of severe pelvic injury (first 24 hours). Anaesth Crit Care Pain Med. 2019;38(2):199–207.CrossRefPubMed Incagnoli P, Puidupin A, Ausset S, Beregi JP, Bessereau J, Bobbia X, Brun J, Brunel E, Buléon C, Choukroun J, et al. Early management of severe pelvic injury (first 24 hours). Anaesth Crit Care Pain Med. 2019;38(2):199–207.CrossRefPubMed
4.
go back to reference Tencer AF, Goulet JA, Hak DJ, Swiontkowski MF, Hamilton KR, Connolly JF, Chandler RW, Karunakar MA, Bosse MJ, Buckwalter JA. Rockwood & Green’s fractures in adults. Rockwood & Green’s fractures in adults; 2014. Tencer AF, Goulet JA, Hak DJ, Swiontkowski MF, Hamilton KR, Connolly JF, Chandler RW, Karunakar MA, Bosse MJ, Buckwalter JA. Rockwood & Green’s fractures in adults. Rockwood & Green’s fractures in adults; 2014.
5.
go back to reference Moed BR, Geer BL. S2 iliosacral screw fixation for disruptions of the posterior pelvic ring: a report of 49 cases. J Orthop Trauma. 2006;20(6):378–83.CrossRefPubMed Moed BR, Geer BL. S2 iliosacral screw fixation for disruptions of the posterior pelvic ring: a report of 49 cases. J Orthop Trauma. 2006;20(6):378–83.CrossRefPubMed
6.
go back to reference Ayoub MA, Gad HM, Seleem OA. Standalone percutaneous transiliac plating of vertically unstable sacral fractures: outcomes, complications, and recommendations. Eur Spine J. 2016;25(4):1153–62.CrossRefPubMed Ayoub MA, Gad HM, Seleem OA. Standalone percutaneous transiliac plating of vertically unstable sacral fractures: outcomes, complications, and recommendations. Eur Spine J. 2016;25(4):1153–62.CrossRefPubMed
7.
go back to reference Yupeng M, Tao H, Weiwei L, Huanyu H, Yong Z, Jiangtao L, Yu L. Biomechanical effects of S1 sacroiliac screws versus S2 sacroiliac screws on sacroiliac screws combined with a lumbar iliac fixation in the treatment of vertical sacral fractures: a biomechanical finite element analysis. BMC Musculoskelet Disord 2023, 24(1). Yupeng M, Tao H, Weiwei L, Huanyu H, Yong Z, Jiangtao L, Yu L. Biomechanical effects of S1 sacroiliac screws versus S2 sacroiliac screws on sacroiliac screws combined with a lumbar iliac fixation in the treatment of vertical sacral fractures: a biomechanical finite element analysis. BMC Musculoskelet Disord 2023, 24(1).
8.
go back to reference Zhou DB, Yang GJ, Wang WL, Li YJ, Pan ZP, Lin XP, Jiang WH. [Treatment of type C pelvic fracture with Stoppa approach and posterior percutaneous plate fixation]. Zhongguo Gu Shang. 2013;26(3):236–9.PubMed Zhou DB, Yang GJ, Wang WL, Li YJ, Pan ZP, Lin XP, Jiang WH. [Treatment of type C pelvic fracture with Stoppa approach and posterior percutaneous plate fixation]. Zhongguo Gu Shang. 2013;26(3):236–9.PubMed
9.
go back to reference Judet R, Judet J, Letournel E. Fractures of the acetabulum: classification and surgical approaches for open reduction: preliminary report. J Bone Joint Surg Am. 1964;46:1615–46.CrossRefPubMed Judet R, Judet J, Letournel E. Fractures of the acetabulum: classification and surgical approaches for open reduction: preliminary report. J Bone Joint Surg Am. 1964;46:1615–46.CrossRefPubMed
10.
go back to reference Routt ML Jr., Simonian PT, Swiontkowski MF. Stabilization of pelvic ring disruptions. Orthop Clin North Am. 1997;28(3):369–88. Routt ML Jr., Simonian PT, Swiontkowski MF. Stabilization of pelvic ring disruptions. Orthop Clin North Am. 1997;28(3):369–88.
11.
go back to reference Du X, Dong B, Li C, Zhang F, Ji Y, Zhang J, Yin C. Dynamic changes of α-melanocyte-stimulating hormone levels in the serum of patients with craniocerebral trauma. Exp Ther Med. 2017;14(3):2511–6.CrossRefPubMedPubMedCentral Du X, Dong B, Li C, Zhang F, Ji Y, Zhang J, Yin C. Dynamic changes of α-melanocyte-stimulating hormone levels in the serum of patients with craniocerebral trauma. Exp Ther Med. 2017;14(3):2511–6.CrossRefPubMedPubMedCentral
12.
go back to reference Mudawi A, Alzobi O, Derbas JN, Ahmed G, Abousamhadaneh M. Optimizing percutaneous reduction and fixation with guidewire modification in pelvic and acetabular fractures: surgical technique and case series. Eur J Orthop Surg Traumatol. 2024;34(4):2107–12.CrossRefPubMedPubMedCentral Mudawi A, Alzobi O, Derbas JN, Ahmed G, Abousamhadaneh M. Optimizing percutaneous reduction and fixation with guidewire modification in pelvic and acetabular fractures: surgical technique and case series. Eur J Orthop Surg Traumatol. 2024;34(4):2107–12.CrossRefPubMedPubMedCentral
13.
go back to reference Simonian PT, Routt ML Jr., Harrington RM, Tencer AF. The unstable iliac fracture: a biomechanical evaluation of internal fixation. Injury. 1997;28(7):469–75. Simonian PT, Routt ML Jr., Harrington RM, Tencer AF. The unstable iliac fracture: a biomechanical evaluation of internal fixation. Injury. 1997;28(7):469–75.
14.
go back to reference Starr AJ, Walter JC, Harris RW, Reinert CM, Jones AL. Percutaneous screw fixation of fractures of the iliac wing and fracture-dislocations of the sacro-iliac joint (OTA types 61-B2.2 and 61-B2.3, or Young-Burgess lateral compression type II pelvic fractures). J Orthop Trauma. 2002;16(2):116–23.CrossRefPubMed Starr AJ, Walter JC, Harris RW, Reinert CM, Jones AL. Percutaneous screw fixation of fractures of the iliac wing and fracture-dislocations of the sacro-iliac joint (OTA types 61-B2.2 and 61-B2.3, or Young-Burgess lateral compression type II pelvic fractures). J Orthop Trauma. 2002;16(2):116–23.CrossRefPubMed
15.
go back to reference Burgess AR, Eastridge BJ, Young JW, Ellison TS, Ellison PS Jr., Poka A, Bathon GH, Brumback RJ. Pelvic ring disruptions: effective classification system and treatment protocols. J Trauma. 1990;30(7):848–56. Burgess AR, Eastridge BJ, Young JW, Ellison TS, Ellison PS Jr., Poka A, Bathon GH, Brumback RJ. Pelvic ring disruptions: effective classification system and treatment protocols. J Trauma. 1990;30(7):848–56.
16.
go back to reference Cai L, Zhang Y, Zheng W, Wang J, Guo X, Feng Y. A novel percutaneous crossed screws fixation in treatment of day type II crescent fracture-dislocation: a finite element analysis. J Orthop Translat. 2020;20:37–46.CrossRefPubMed Cai L, Zhang Y, Zheng W, Wang J, Guo X, Feng Y. A novel percutaneous crossed screws fixation in treatment of day type II crescent fracture-dislocation: a finite element analysis. J Orthop Translat. 2020;20:37–46.CrossRefPubMed
17.
go back to reference Russ MK, Navarre P, Stevens JP. An Alternative Site for Pin Placement in External fixation of pelvic fractures: lateral posterior pelvic external Fixator Surgical technique. Front Surg. 2020;7:621125.CrossRefPubMed Russ MK, Navarre P, Stevens JP. An Alternative Site for Pin Placement in External fixation of pelvic fractures: lateral posterior pelvic external Fixator Surgical technique. Front Surg. 2020;7:621125.CrossRefPubMed
18.
go back to reference Jingqi D, Zhenhua Z, Tao L, Jiacheng L, Jiajun M, Sheqiang C, Yilan L, Peiyi L, Shicai F. A new technique for percutaneous screw fixation for treating FFP IIIa and IIIb fragility fractures of the pelvis. Sci Rep 2024, 14(1). Jingqi D, Zhenhua Z, Tao L, Jiacheng L, Jiajun M, Sheqiang C, Yilan L, Peiyi L, Shicai F. A new technique for percutaneous screw fixation for treating FFP IIIa and IIIb fragility fractures of the pelvis. Sci Rep 2024, 14(1).
19.
go back to reference Yinger K, Scalise J, Olson SA, Bay BK, Finkemeier CG. Biomechanical comparison of posterior pelvic ring fixation. J Orthop Trauma. 2003;17(7):481–7.CrossRefPubMed Yinger K, Scalise J, Olson SA, Bay BK, Finkemeier CG. Biomechanical comparison of posterior pelvic ring fixation. J Orthop Trauma. 2003;17(7):481–7.CrossRefPubMed
20.
go back to reference Zahaf S, Kebdani S, Ghalem M, Mestar A, Zina N, Aour B. Biomechanical evaluation of two posterior lumbar Intervertebral Fusion Surgical scenarios Reinforced by a rigid posterior fixation system in the vertebral column analyzed by the Finite element Method. Nano Biomed Eng. 2018;10(3):258–78.CrossRef Zahaf S, Kebdani S, Ghalem M, Mestar A, Zina N, Aour B. Biomechanical evaluation of two posterior lumbar Intervertebral Fusion Surgical scenarios Reinforced by a rigid posterior fixation system in the vertebral column analyzed by the Finite element Method. Nano Biomed Eng. 2018;10(3):258–78.CrossRef
21.
go back to reference Zahaf S, Kebdani S. Study and analysis of mechanical behavior between rigid and dynamic fixation systems analyzed by the finite element method. J Biomimetics Biomaterials Biomedical Eng. 2017;33:12–31.CrossRef Zahaf S, Kebdani S. Study and analysis of mechanical behavior between rigid and dynamic fixation systems analyzed by the finite element method. J Biomimetics Biomaterials Biomedical Eng. 2017;33:12–31.CrossRef
22.
go back to reference Zahaf S, Kebdani S. Biomechanical Study between the rigid and dynamic fixation systems of the spinal column analyzed by the Finite element Method. Nano Biomed Eng 2017, 9(2). Zahaf S, Kebdani S. Biomechanical Study between the rigid and dynamic fixation systems of the spinal column analyzed by the Finite element Method. Nano Biomed Eng 2017, 9(2).
23.
go back to reference Zahaf S, Dahmane M, Belaziz A, Bouri I, Afane N. Failure analysis of semi-elliptical crack behavior in the cement mantle of a total hip prosthesis. Mater Phys Mech 2022;48(2). Zahaf S, Dahmane M, Belaziz A, Bouri I, Afane N. Failure analysis of semi-elliptical crack behavior in the cement mantle of a total hip prosthesis. Mater Phys Mech 2022;48(2).
24.
go back to reference Mestar A, Zahaf S, Zina N, Boutaous A. Numerical study of the effect of elastomer and cement of stress absorbers on the reduction of stresses in tibia and tibial bone analysed by finite element method. Nano Biomed Eng. 2018;10(1):56–78.CrossRef Mestar A, Zahaf S, Zina N, Boutaous A. Numerical study of the effect of elastomer and cement of stress absorbers on the reduction of stresses in tibia and tibial bone analysed by finite element method. Nano Biomed Eng. 2018;10(1):56–78.CrossRef
25.
go back to reference Mestar A, Zahaf S, Zina N, Boutaous A. Development and validation of a numerical model for the mechanical behavior of knee prosthesis analyzed by the finite elements method. J Biomimetics Biomaterials Biomedical Eng. 2018;37:12–42.CrossRef Mestar A, Zahaf S, Zina N, Boutaous A. Development and validation of a numerical model for the mechanical behavior of knee prosthesis analyzed by the finite elements method. J Biomimetics Biomaterials Biomedical Eng. 2018;37:12–42.CrossRef
26.
go back to reference Yu KH, Hong JJ, Guo XS, Zhou DS. Comparison of reconstruction plate screw fixation and percutaneous cannulated screw fixation in treatment of tile B1 type pubic symphysis diastasis: a finite element analysis and 10-year clinical experience. J Orthop Surg Res. 2015;10:151.CrossRefPubMedPubMedCentral Yu KH, Hong JJ, Guo XS, Zhou DS. Comparison of reconstruction plate screw fixation and percutaneous cannulated screw fixation in treatment of tile B1 type pubic symphysis diastasis: a finite element analysis and 10-year clinical experience. J Orthop Surg Res. 2015;10:151.CrossRefPubMedPubMedCentral
27.
go back to reference Gardner MJ, Routt ML Jr. Transiliac-transsacral screws for posterior pelvic stabilization. J Orthop Trauma. 2011;25(6):378–84. Gardner MJ, Routt ML Jr. Transiliac-transsacral screws for posterior pelvic stabilization. J Orthop Trauma. 2011;25(6):378–84.
28.
go back to reference Tile M. Acute Pelvic fractures: I. Causation and classification. J Am Acad Orthop Surg. 1996;4(3):143–51.CrossRefPubMed Tile M. Acute Pelvic fractures: I. Causation and classification. J Am Acad Orthop Surg. 1996;4(3):143–51.CrossRefPubMed
29.
go back to reference Simonian PT, Routt ML, Jr., Harrington RM, Tencer AF. Anterior versus posterior provisional fixation in the unstable pelvis. A biomechanical comparison. Clin Orthop Relat Res. 1995;310:245–51. Simonian PT, Routt ML, Jr., Harrington RM, Tencer AF. Anterior versus posterior provisional fixation in the unstable pelvis. A biomechanical comparison. Clin Orthop Relat Res. 1995;310:245–51.
30.
go back to reference Simonian PT, Routt ML, Jr., Harrington RM, Mayo KA, Tencer AF. Biomechanical simulation of the anteroposterior compression injury of the pelvis. An understanding of instability and fixation. Clin Orthop Relat Res. 1994;309:245–56. Simonian PT, Routt ML, Jr., Harrington RM, Mayo KA, Tencer AF. Biomechanical simulation of the anteroposterior compression injury of the pelvis. An understanding of instability and fixation. Clin Orthop Relat Res. 1994;309:245–56.
31.
go back to reference Simonian PT, Routt ML Jr., Harrington RM, Tencer AF. Internal fixation of the unstable anterior pelvic ring: a biomechanical comparison of standard plating techniques and the retrograde medullary superior pubic ramus screw. J Orthop Trauma. 1994;8(6):476–82. Simonian PT, Routt ML Jr., Harrington RM, Tencer AF. Internal fixation of the unstable anterior pelvic ring: a biomechanical comparison of standard plating techniques and the retrograde medullary superior pubic ramus screw. J Orthop Trauma. 1994;8(6):476–82.
32.
go back to reference Stevenson AJ, Swartman B, Bucknill AT. [Percutaneous internal fixation of pelvic fractures. German Version] Unfallchirurg. 2016;119(10):825–34.CrossRefPubMed Stevenson AJ, Swartman B, Bucknill AT. [Percutaneous internal fixation of pelvic fractures. German Version] Unfallchirurg. 2016;119(10):825–34.CrossRefPubMed
33.
go back to reference Grewal IS, Starr AJ. What’s New in Percutaneous Pelvis fracture surgery? Orthop Clin North Am. 2020;51(3):317–24.CrossRefPubMed Grewal IS, Starr AJ. What’s New in Percutaneous Pelvis fracture surgery? Orthop Clin North Am. 2020;51(3):317–24.CrossRefPubMed
34.
go back to reference Starr AJ, Nakatani T, Reinert CM, Cederberg K. Superior pubic ramus fractures fixed with percutaneous screws: what predicts fixation failure? J Orthop Trauma. 2008;22(2):81–7.CrossRefPubMed Starr AJ, Nakatani T, Reinert CM, Cederberg K. Superior pubic ramus fractures fixed with percutaneous screws: what predicts fixation failure? J Orthop Trauma. 2008;22(2):81–7.CrossRefPubMed
35.
go back to reference von Glinski A, Frieler S, Yilmaz E, Ishak B, Goodmanson R, Iwanaga J, Schildhauer TA, Chapman JR, Oskouian R, Mayo K, Tubbs RS. Osteology of the ilium revised: illuminating the clinical relevance. Eur J Trauma Emerg Surg. 2021;47(6):1671–7.CrossRef von Glinski A, Frieler S, Yilmaz E, Ishak B, Goodmanson R, Iwanaga J, Schildhauer TA, Chapman JR, Oskouian R, Mayo K, Tubbs RS. Osteology of the ilium revised: illuminating the clinical relevance. Eur J Trauma Emerg Surg. 2021;47(6):1671–7.CrossRef
36.
go back to reference Guo JL, Dong WC, Zhang ZW, Zhang RP, Yin YC, Chen W, Zhang YZ, Hou ZY. An exploratory study of pelvis anatomy to revise the bony canal used for LC2 screw insertion. BMC Musculoskelet Disord. 2022;23(1):8.CrossRef Guo JL, Dong WC, Zhang ZW, Zhang RP, Yin YC, Chen W, Zhang YZ, Hou ZY. An exploratory study of pelvis anatomy to revise the bony canal used for LC2 screw insertion. BMC Musculoskelet Disord. 2022;23(1):8.CrossRef
37.
go back to reference Li YG, Sang XG, Wang ZY, Cheng L, Liu H, Qin T, Di K. Iliac screw for reconstructing posterior pelvic ring in tile type C1 pelvic fractures. Orthop Traumatol-Sur. 2018;104(6):923–8.CrossRef Li YG, Sang XG, Wang ZY, Cheng L, Liu H, Qin T, Di K. Iliac screw for reconstructing posterior pelvic ring in tile type C1 pelvic fractures. Orthop Traumatol-Sur. 2018;104(6):923–8.CrossRef
38.
go back to reference Müller F, Füchtmeier B. A systematic review of the transiliac internal fixator (TIFI) for posterior pelvic injuries. Sicot j. 2021;7:40.CrossRef Müller F, Füchtmeier B. A systematic review of the transiliac internal fixator (TIFI) for posterior pelvic injuries. Sicot j. 2021;7:40.CrossRef
39.
go back to reference Cui R, Li J, Jiang Y, Sun H, Tan Y, Duan L, Wu M. Trajectory optimisation with musculoskeletal integration features for fracture reduction orthopaedic robot. Int J Med Robot. 2022;18(2):e2372.CrossRefPubMed Cui R, Li J, Jiang Y, Sun H, Tan Y, Duan L, Wu M. Trajectory optimisation with musculoskeletal integration features for fracture reduction orthopaedic robot. Int J Med Robot. 2022;18(2):e2372.CrossRefPubMed
40.
go back to reference Bolun Z, Huixiang W, Xingguang T, Haochen S, Leo J, Xiaojun C. A bidirectional framework for fracture simulation and deformation-based restoration prediction in pelvic fracture surgical planning. Med Image Anal 2024, 97(0). Bolun Z, Huixiang W, Xingguang T, Haochen S, Leo J, Xiaojun C. A bidirectional framework for fracture simulation and deformation-based restoration prediction in pelvic fracture surgical planning. Med Image Anal 2024, 97(0).
41.
go back to reference Zhang J, Wei Y, Wang J, Yu BQ. Numerical study of pedicle screw construction and locking compression plate fixation in posterior pelvic ring injuries: analyzed by finite element method. Med (Baltim). 2024;103(20):9.CrossRef Zhang J, Wei Y, Wang J, Yu BQ. Numerical study of pedicle screw construction and locking compression plate fixation in posterior pelvic ring injuries: analyzed by finite element method. Med (Baltim). 2024;103(20):9.CrossRef
Metadata
Title
Experimental study of fractures of the posterior pelvic ring C1.1 using LC-II screws and internal fixation by plate
Authors
Yu Liu
Xukai Wang
Bin Tian
Hong Yao
Guangyao Liu
Publication date
01-12-2024
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2024
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-024-05229-2