Open Access 01-12-2024 | Pacemaker | Study protocol

Effects on tricuspid regurgitation by different techniques for passing permanent pacemaker leads through the tricuspid valve: a randomized, open-label, superiority clinical trial study protocol

Authors: Jianan Hong, Zhanyi Qiu, Zhongbo Xiao, Xiaobin Ni, Yanling Fang, Shiwan Wu, Yandan Xie, Chang Chen, Yequn Chen, Muli Wu

Published in: Trials | Issue 1/2024

Login to get access

Abstract

Background

In recent years, lead-induced tricuspid regurgitation (LITR) has attracted increasing attention. At present, there are two commonly used transvalvular methods for pacing lead wires to enter the right ventricle. The first transvalvular approach involves placing the tip of the pacing lead directly through the tricuspid valve into the right ventricle, including “direct-crossing” and “drop-down.” The second transvalvular approach is to bend the pacing lead so that the reflexed lead body crosses the tricuspid valve and then enters the right ventricle, which is “prolapsing.” However, there are no clinical trials to evaluate or compare the effects of the above two different pacing lead transvalvular approaches on tricuspid regurgitation. In this study, we will perform a randomized clinical trial to understand the effect of different transvalvular lead wire transversal techniques on the incidence of tricuspid regurgitation.

Methods

Three hundred seventy-six subjects with right ventricular single-chamber pacemaker implantation or dual-chamber pacemaker implantation were recruited in the First Affiliated Hospital of Shantou University Medical College. Participants will be randomized into the direct group (“direct-crossing” and “drop-down”) or the bending group (“prolapsing”). The primary objective will be new tricuspid regurgitation or exacerbation of existing tricuspid regurgitation within 1 year of follow-up.

Discussion

This study aims to verify whether different transvalvular approaches influence the incidence of LITR, and is expected to provide an optimized method for routine pacemaker surgery and improve the long-term prognosis and quality of life of patients.

Trial registration

Chinese Clinical Trials Registry ChiCTR2100045558. Registered on April 19, 2021.
Appendix
Available only for authorised users
Literature
1.
go back to reference Delling FN, Hassan ZK, Piatkowski G, et al. Tricuspid regurgitation and mortality in patients with transvenous permanent pacemaker leads. Am J Cardiol. 2016;117(6):988–92.CrossRefPubMedPubMedCentral Delling FN, Hassan ZK, Piatkowski G, et al. Tricuspid regurgitation and mortality in patients with transvenous permanent pacemaker leads. Am J Cardiol. 2016;117(6):988–92.CrossRefPubMedPubMedCentral
2.
go back to reference Dokainish H, Elbarasi E, Masiero S, et al. Prospective study of tricuspid valve regurgitation associated with permanent leads in patients undergoing cardiac rhythm device implantation: background, rationale, and design. Glob Cardiol Sci Pract. 2015;2015(3):41.CrossRefPubMedPubMedCentral Dokainish H, Elbarasi E, Masiero S, et al. Prospective study of tricuspid valve regurgitation associated with permanent leads in patients undergoing cardiac rhythm device implantation: background, rationale, and design. Glob Cardiol Sci Pract. 2015;2015(3):41.CrossRefPubMedPubMedCentral
4.
go back to reference Arsalan M, Walther T, Smith RL 2nd, Grayburn PA. Tricuspid regurgitation diagnosis and treatment. Eur Heart J. 2017;38(9):634–8.PubMed Arsalan M, Walther T, Smith RL 2nd, Grayburn PA. Tricuspid regurgitation diagnosis and treatment. Eur Heart J. 2017;38(9):634–8.PubMed
5.
go back to reference Gelves-Meza J, Lang RM, Valderrama-Achury MD, et al. Tricuspid regurgitation related to cardiac implantable electronic devices: an integrative review. J Am Soc Echocardiogr. 2022;35(11):1107–22.CrossRefPubMed Gelves-Meza J, Lang RM, Valderrama-Achury MD, et al. Tricuspid regurgitation related to cardiac implantable electronic devices: an integrative review. J Am Soc Echocardiogr. 2022;35(11):1107–22.CrossRefPubMed
6.
go back to reference Adamo M, Chioncel O, Pagnesi M, et al. Epidemiology, pathophysiology, diagnosis and management of chronic right-sided heart failure and tricuspid regurgitation. A clinical consensus statement of the Heart Failure Association (HFA) and the European Association of Percutaneous Cardiovascular Interventions (EAPCI) of the ESC. Eur J Heart Fail. 2024;26(1):18–33.CrossRefPubMed Adamo M, Chioncel O, Pagnesi M, et al. Epidemiology, pathophysiology, diagnosis and management of chronic right-sided heart failure and tricuspid regurgitation. A clinical consensus statement of the Heart Failure Association (HFA) and the European Association of Percutaneous Cardiovascular Interventions (EAPCI) of the ESC. Eur J Heart Fail. 2024;26(1):18–33.CrossRefPubMed
7.
go back to reference Agarwal S, Tuzcu EM, Rodriguez ER, Tan CD, Rodriguez LL, Kapadia SR. Interventional cardiology perspective of functional tricuspid regurgitation. Circ Cardiovasc Interv. 2009;2(6):565–73.CrossRefPubMed Agarwal S, Tuzcu EM, Rodriguez ER, Tan CD, Rodriguez LL, Kapadia SR. Interventional cardiology perspective of functional tricuspid regurgitation. Circ Cardiovasc Interv. 2009;2(6):565–73.CrossRefPubMed
8.
go back to reference Rogers JH, Bolling SF. The tricuspid valve: current perspective and evolving management of tricuspid regurgitation. Circulation. 2009;119(20):2718–25.CrossRefPubMed Rogers JH, Bolling SF. The tricuspid valve: current perspective and evolving management of tricuspid regurgitation. Circulation. 2009;119(20):2718–25.CrossRefPubMed
9.
go back to reference Cheng Y, Gao H, Tang L, Li J, Yao L. Clinical utility of three-dimensional echocardiography in the evaluation of tricuspid regurgitation induced by implantable device leads. Echocardiography. 2016;33(11):1689–96.CrossRefPubMed Cheng Y, Gao H, Tang L, Li J, Yao L. Clinical utility of three-dimensional echocardiography in the evaluation of tricuspid regurgitation induced by implantable device leads. Echocardiography. 2016;33(11):1689–96.CrossRefPubMed
10.
go back to reference Al-Mohaissen MA, Chan KL. Prevalence and mechanism of tricuspid regurgitation following implantation of endocardial leads for pacemaker or cardioverter-defibrillator. J Am Soc Echocardiogr. 2012;25(3):245–52.CrossRefPubMed Al-Mohaissen MA, Chan KL. Prevalence and mechanism of tricuspid regurgitation following implantation of endocardial leads for pacemaker or cardioverter-defibrillator. J Am Soc Echocardiogr. 2012;25(3):245–52.CrossRefPubMed
11.
go back to reference Paniagua D, Aldrich HR, Lieberman EH, Lamas GA, Agatston AS. Increased prevalence of significant tricuspid regurgitation in patients with transvenous pacemakers leads. Am J Cardiol. 1998;82(9):1130–A9.CrossRefPubMed Paniagua D, Aldrich HR, Lieberman EH, Lamas GA, Agatston AS. Increased prevalence of significant tricuspid regurgitation in patients with transvenous pacemakers leads. Am J Cardiol. 1998;82(9):1130–A9.CrossRefPubMed
12.
go back to reference Lee RC, Friedman SE, Kono AT, Greenberg ML, Palac RT. Tricuspid regurgitation following implantation of endocardial leads: incidence and predictors. Pacing Clin Electrophysiol. 2015;38(11):1267–74.CrossRefPubMed Lee RC, Friedman SE, Kono AT, Greenberg ML, Palac RT. Tricuspid regurgitation following implantation of endocardial leads: incidence and predictors. Pacing Clin Electrophysiol. 2015;38(11):1267–74.CrossRefPubMed
13.
go back to reference Vaturi M, Kusniec J, Shapira Y, et al. Right ventricular pacing increases tricuspid regurgitation grade regardless of the mechanical interference to the valve by the electrode. Eur J Echocardiogr. 2010;11(6):550–3.CrossRefPubMed Vaturi M, Kusniec J, Shapira Y, et al. Right ventricular pacing increases tricuspid regurgitation grade regardless of the mechanical interference to the valve by the electrode. Eur J Echocardiogr. 2010;11(6):550–3.CrossRefPubMed
14.
go back to reference Seo Y, Ishizu T, Nakajima H, Sekiguchi Y, Watanabe S, Aonuma K. Clinical utility of 3-dimensional echocardiography in the evaluation of tricuspid regurgitation caused by pacemaker leads. Circ J. 2008;72(9):1465–70.CrossRefPubMed Seo Y, Ishizu T, Nakajima H, Sekiguchi Y, Watanabe S, Aonuma K. Clinical utility of 3-dimensional echocardiography in the evaluation of tricuspid regurgitation caused by pacemaker leads. Circ J. 2008;72(9):1465–70.CrossRefPubMed
15.
go back to reference Kim JB, Spevack DM, Tunick PA, et al. The effect of transvenous pacemaker and implantable cardioverter defibrillator lead placement on tricuspid valve function: an observational study. J Am Soc Echocardiogr. 2008;21(3):284–7.CrossRefPubMed Kim JB, Spevack DM, Tunick PA, et al. The effect of transvenous pacemaker and implantable cardioverter defibrillator lead placement on tricuspid valve function: an observational study. J Am Soc Echocardiogr. 2008;21(3):284–7.CrossRefPubMed
16.
17.
go back to reference Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Europace. 2022;24(1):71–164.CrossRefPubMed Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Europace. 2022;24(1):71–164.CrossRefPubMed
18.
go back to reference Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021;42(35):3427–520.CrossRefPubMed Glikson M, Nielsen JC, Kronborg MB, et al. 2021 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Eur Heart J. 2021;42(35):3427–520.CrossRefPubMed
19.
go back to reference Gibson TC, Davidson RC, DeSilvey DL. Presumptive tricuspid valve malfunction induced by a pacemaker lead: a case report and review of the literature. Pacing Clin Electrophysiol. 1980;3(1):88–95.CrossRefPubMed Gibson TC, Davidson RC, DeSilvey DL. Presumptive tricuspid valve malfunction induced by a pacemaker lead: a case report and review of the literature. Pacing Clin Electrophysiol. 1980;3(1):88–95.CrossRefPubMed
20.
go back to reference de Cock CC, Vinkers M, Van Campe LC, Verhorst PM, Visser CA. Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads: a clinical and echocardiographic study. Pacing Clin Electrophysiol. 2000;23(4 Pt 1):423–6.CrossRefPubMed de Cock CC, Vinkers M, Van Campe LC, Verhorst PM, Visser CA. Long-term outcome of patients with multiple (> or = 3) noninfected transvenous leads: a clinical and echocardiographic study. Pacing Clin Electrophysiol. 2000;23(4 Pt 1):423–6.CrossRefPubMed
21.
go back to reference Al-Bawardy R, Krishnaswamy A, Bhargava M, et al. Tricuspid regurgitation in patients with pacemakers and implantable cardiac defibrillators: a comprehensive review. Clin Cardiol. 2013;36(5):249–54.CrossRefPubMedPubMedCentral Al-Bawardy R, Krishnaswamy A, Bhargava M, et al. Tricuspid regurgitation in patients with pacemakers and implantable cardiac defibrillators: a comprehensive review. Clin Cardiol. 2013;36(5):249–54.CrossRefPubMedPubMedCentral
22.
go back to reference Klutstein M, Balkin J, Butnaru A, Ilan M, Lahad A, Rosenmann D. Tricuspid incompetence following permanent pacemaker implantation. Pacing Clin Electrophysiol. 2009;32(Suppl 1):S135–7.PubMed Klutstein M, Balkin J, Butnaru A, Ilan M, Lahad A, Rosenmann D. Tricuspid incompetence following permanent pacemaker implantation. Pacing Clin Electrophysiol. 2009;32(Suppl 1):S135–7.PubMed
23.
go back to reference Lin G, Nishimura RA, Connolly HM, Dearani JA, Sundt TM 3rd, Hayes DL. Severe symptomatic tricuspid valve regurgitation due to permanent pacemaker or implantable cardioverter-defibrillator leads. J Am Coll Cardiol. 2005;45(10):1672–5.CrossRefPubMed Lin G, Nishimura RA, Connolly HM, Dearani JA, Sundt TM 3rd, Hayes DL. Severe symptomatic tricuspid valve regurgitation due to permanent pacemaker or implantable cardioverter-defibrillator leads. J Am Coll Cardiol. 2005;45(10):1672–5.CrossRefPubMed
Metadata
Title
Effects on tricuspid regurgitation by different techniques for passing permanent pacemaker leads through the tricuspid valve: a randomized, open-label, superiority clinical trial study protocol
Authors
Jianan Hong
Zhanyi Qiu
Zhongbo Xiao
Xiaobin Ni
Yanling Fang
Shiwan Wu
Yandan Xie
Chang Chen
Yequn Chen
Muli Wu
Publication date
01-12-2024
Publisher
BioMed Central
Published in
Trials / Issue 1/2024
Electronic ISSN: 1745-6215
DOI
https://doi.org/10.1186/s13063-024-08630-8