日本放射線技術学会雑誌
Online ISSN : 1881-4883
Print ISSN : 0369-4305
ISSN-L : 0369-4305
臨床技術
脳血管IVRにおける水晶体被曝に対する防護シールドの作成と被曝低減効果
武 俊夫佐藤 香織木内 克典中澤 靖夫
著者情報
ジャーナル フリー

2007 年 63 巻 11 号 p. 1246-1252

詳細
抄録

A variety of radiation hazards resulting from interventional radiology (IVR) have been reported in recent years. Particularly affected are the skin and the crystalline lens, with their high radiation sensitivity. During neurological interventions, the radiological technologist should consider decreasing radiation exposure. We found exposure projections where the exposure dose became a radiation hazard for the crystalline lens, and examined an efficient method of cover for the exposure projections used for neurological interventions. The exposure projection for maximum crystalline lens radiation exposure was a lateral projection. In the crystalline lens the maximum exposure to radiation was on the X-ray tube side. The method of defense adopted was that of installing a lead plate of the appropriate shape on the surface of the X-ray tube collimator. In other exposure projections, this cover did not become a redundant shadow. With the cover that was created, the X-ray side crystalline lens lateral projection could be defended effectively.

著者関連情報
© 2007 公益社団法人 日本放射線技術学会
前の記事 次の記事
feedback
Top