FDGを投与された被検者からのPETセンター専属運転手に対する職業被ばく線量の検討
中村 文俊菅野 敏彦岡田 裕之吉川 悦次安藤 一郎二ツ橋 昌実新家 智美尾内 康臣鳥塚 達郎
著者情報
ジャーナル フリー

2006 年 62 巻 8 号 p. 1105-1110

詳細
抄録

With the recent increase in FDG-PET examinations, concern has mounted regarding radiation exposure to hospital staff and the general public from patients injected with FDG. Because our PET institution is located 15 km from the hospital that provides these examinations, a driver has been designated to transport patients injected with FDG. This study was designed to measure the radiation dose to the driver from these patients (n=28) and to compare it with the estimated dose. A pocket dosimeter was used to measure radiation exposure to the driver. When the distances between the driver and patient were 1.1 m and 1.9 m, mean measured doses were 7.31 μSv and 2.26 μSv, respectively, while mean estimated doses were 8.61 μSv and 2.82 μSv, respectively, per trip. It was presumed that maximum radiation exposure per year was between 3.02 mSv (1.1 m) and 0.92 mSv (1.9 m). According to our data, the measured dose was 20% lower than the estimated dose. This discrepancy may be due to the difference between the volume source (measured dose) and point source (estimated dose).