Using a whole-body 31P birdcage transmit coil and 16-element receive array for human cardiac metabolic imaging at 7T

  • Ladislav Valkovič ,

    Roles Conceptualization, Data curation, Formal analysis, Investigation, Methodology, Project administration, Software, Supervision, Validation, Visualization, Writing – original draft

    ladislav.valkovic@cardiov.ox.ac.uk

    Affiliations Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom, Department of Imaging Methods, Institute of Measurement Science, Slovak Academy of Sciences, Bratislava, Slovakia

  • Iulius Dragonu,

    Roles Conceptualization, Investigation, Supervision, Writing – review & editing

    Affiliation Siemens Healthcare Limited, Frimley, United Kingdom

  • Salam Almujayyaz,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation MR Coils BV, Zaltbommel, Netherlands

  • Alex Batzakis,

    Roles Conceptualization, Methodology, Writing – review & editing

    Affiliation MR Coils BV, Zaltbommel, Netherlands

  • Liam A. J. Young,

    Roles Formal analysis, Investigation, Visualization, Writing – review & editing

    Affiliation Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom

  • Lucian A. B. Purvis,

    Roles Investigation, Software, Writing – review & editing

    Affiliation Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom

  • William T. Clarke,

    Roles Conceptualization, Methodology, Software, Writing – review & editing

    Affiliation Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom

  • Tobias Wichmann,

    Roles Formal analysis, Investigation, Writing – review & editing

    Affiliation Rapid Biomedical GmbH, Rimpar, Germany

  • Titus Lanz,

    Roles Conceptualization, Supervision, Writing – review & editing

    Affiliation Rapid Biomedical GmbH, Rimpar, Germany

  • Stefan Neubauer,

    Roles Conceptualization, Funding acquisition, Writing – review & editing

    Affiliation Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom

  • Matthew D. Robson,

    Roles Conceptualization, Validation, Writing – review & editing

    Affiliation Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom

  • Dennis W. J. Klomp,

    Roles Conceptualization, Investigation, Methodology, Supervision, Writing – review & editing

    Affiliations MR Coils BV, Zaltbommel, Netherlands, Department of Radiology, University Medical Center Utrecht, Utrecht, Netherlands

  • Christopher T. Rodgers

    Roles Conceptualization, Data curation, Funding acquisition, Investigation, Methodology, Project administration, Software, Supervision, Validation, Writing – review & editing

    Affiliation Oxford Centre for Clinical Magnetic Resonance Research (OCMR), BHF Centre of Research Excellence, University of Oxford, Oxford, United Kingdom

Abstract

Purpose

Cardiac phosphorus magnetic resonance spectroscopy (31P-MRS) provides unique insight into the mechanisms of heart failure. Yet, clinical applications have been hindered by the restricted sensitivity of the surface radiofrequency-coils normally used. These permit the analysis of spectra only from the interventricular septum, or large volumes of myocardium, which may not be meaningful in focal disease. Löring et al. recently presented a prototype whole-body (52 cm diameter) transmit/receive birdcage coil for 31P at 7T. We now present a new, easily-removable, whole-body 31P transmit radiofrequency-coil built into a patient-bed extension combined with a 16-element receive array for cardiac 31P-MRS.

Materials and methods

A fully-removable (55 cm diameter) birdcage transmit coil was combined with a 16-element receive array on a Magnetom 7T scanner (Siemens, Germany). Electro-magnetic field simulations and phantom tests of the setup were performed. In vivo maps of B1+, metabolite signals, and saturation-band efficiency were acquired across the torsos of eight volunteers.

Results

The combined (volume-transmit, local receive array) setup increased signal-to-noise ratio 2.6-fold 10 cm below the array (depth of the interventricular septum) compared to using the birdcage coil in transceiver mode. The simulated coefficient of variation for B1+ of the whole-body coil across the heart was 46.7% (surface coil 129.0%); and the in vivo measured value was 38.4%. Metabolite images of 2,3-diphosphoglycerate clearly resolved the ventricular blood pools, and muscle tissue was visible in phosphocreatine (PCr) maps. Amplitude-modulated saturation bands achieved 71±4% suppression of phosphocreatine PCr in chest-wall muscles. Subjects reported they were comfortable.

Conclusion

This easy-to-assemble, volume-transmit, local receive array coil combination significantly improves the homogeneity and field-of-view for metabolic imaging of the human heart at 7T.

Introduction

Phosphorus magnetic resonance spectroscopy (31P-MRS) plays an important role in the assessment of tissue metabolism, through measurement of high-energy metabolites, such as phosphocreatine (PCr) and adenosine triphosphate (ATP), in vivo [1, 2]. The PCr/ATP ratio is of particular interest in cardiovascular medicine, serving as a valuable biomarker that changes in most major heart diseases [35], and which even predicts mortality in patients with dilated cardiomyopathy [3]. Impaired cardiac PCr/ATP ratios also occur in systemic diseases, such as type-II diabetes [6] and obesity [7].

Yet 31P-MRS is still applied in clinical medicine, principally because of its comparatively long scan times, signal-to-noise ratios that are too low for single-subject comparisons, and the challenges of recording regionally-resolved spectra across the heart.

Moving to ultra-high field (i.e. 7T) from clinical MR systems (i.e. ≤3T) leads to a 2.5-fold increase in SNR [8, 9] and increased spectral resolution [10, 11]. The use of dedicated receive-arrays further improves the SNR and extends coverage to more of the heart at 7T [5]. However, so far receive array coils for cardiac 31P-MRS have used surface coils for transmission. Surface coils’ radiofrequency (RF) transmit field strength (B1+) inherently drops-off rapidly with increasing distance from the coil to the volume of interest [12, 13]. This makes spatially-resolved 31P-MRS imaging (31P-MRSI) across the heart challenging. Even with custom-built surface transmit coils, and optimised adiabatic B1+-insensitive excitation pulses, regionally-resolved measurement across the whole heart in clinically feasible times remains elusive [14].

Recently, Löring et al inserted a whole-body-sized (52 cm in diameter) 31P birdcage RF-coil into the bore of a Philips 7T MR system (Philips Healthcare, Best, Netherlands). This showed relatively uniform spectral profiles for 31P-MRSI examinations with a rectangular RF pulse in a cylindrical phantom and in a human subject [15]. However, their coil had to be inserted into the magnet bore after complete removal of the original patient bed. They also used the RF screen inside the patient tube decreasing the patient space significantly and affecting subject comfort and study inclusion criteria. An alternative design, not requiring complex preparation, and allowing fast installation/removal would be preferred. Furthermore, the previously described proof-of-concept coil operated only in transceiver mode, which led to low resolution MRSI matrixes in order to compensate for the inherently low SNR.

In this study, we report initial results of a collaborative project to design, build and test a new, easily-removable, high-pass birdcage, whole-body (55cm-diameter) 31P transmit RF-coil for use on a Magnetom 7T MR scanner (Siemens Healthcare, Erlangen, Germany), integrated into an extension of the scanner’s motorized patient bed; we use it in conjunction with a 16-element anterior receive array (Rapid Biomedical, Rimpar, Germany) for cardiac 31P-MRSI at 7T.

Materials and methods

The underlying design of our whole-body coil was similar to that previously reported [15]. However, to allow the desired easily removable setup, the design of the whole-body coil was adjusted by MR Coils (MR Coils BV, Zaltbommel, Netherlands) so that the lower rungs of the birdcage were integrated into an extension of the scanner’s motorised patient bed. This is driven onto custom-built support rails at the service-end of the magnet for subject access (Fig 1E). As the coil had to be inserted without removing the patient tube, while maximizing space for the subject, the existing RF shield integrated in the scanner’s gradient coil was used as the RF shield for the body coil. Therefore, only the rings and rods of the birdcage had to be shifted inside the bore of 58 cm, leading to a setup with inner coil diameter of 55 cm. The upper part of the birdcage coil was made detachable to ease patient positioning. The coil was tuned to 120.3 MHz, the frequency for 31P-MR on a Magnetom 7T MR scanner. The standard Siemens 8-kW RF power amplifier was used at this stage.

thumbnail
Fig 1. Design and simulations of the new whole-body RF-coil.

(a-d) Electromagnetic field modelling results for the designed whole-body coil. (a) Coronal view and (b) transversal view of the 10g local SAR maps showing hotspots in the chest wall and shoulders. (c) Predicted B1+ efficiency map. Note the uniformity of excitation across the heart and liver. (d) Direct comparison of B1+ variability across the heart between surface transmit (Tx) coil and the whole-body Tx coil. (e) A photograph of the new apparatus installed on our 7T MRI scanner. The extension of the patient table with the integrated whole-body coil rests on a custom-build support frame and connects to the existing motorized patient bed from the service end of the scanner.

https://doi.org/10.1371/journal.pone.0187153.g001

Simulations of electro-magnetic fields were performed out using CST Studio Suite 2016 (CST Computer Simulation Technology AG, Darmstadt, Germany). A whole-body coil matching the proposed design, i.e. a 40-cm long, 20-rung birdcage coil with a 55 cm diameter, was simulated inside an RF shield matching the diameter and length of the shield of the scanner. Two voxel models, “Gustav” and “Laura”, were simulated in order to identify the 10g local, global head and global body specific absorption rate (SAR) values for 1W input power. In order to get the worst case for the 10g local SAR values, one arm of the voxel model was placed in direct contact with the inner lining of the transmit coil. A rectangular (26×28 cm2) surface transmit coil, previously used in our lab for 31P excitation at 7T [5], was also simulated.

The coil performance was first tested in a two compartment phantom consisting of a 18 L container (outer dimensions 46×24×17 cm3), filled with NaCl(aq) (73 mM), and a 2cm cube, filled with KH2PO4(aq) (1.8 M), set at a 10cm depth. A series of fully relaxed (TR = 10 s), non-localized, single average 31P-MRS spectra were acquired increasing the peak voltage of a 10 ms rectangular pulse from 25 V to 400 V in 25 V steps. The SNR and peak B1+ were compared between a quadrature surface 31P RF-coil in transceiver mode (two 15 cm diameter loops, with overlap decoupling, as described in [14]), the new whole-body (55-cm diameter) birdcage coil in transceiver mode; and the combination of the whole-body coil for transmission and a 16-element receive array, consisting of 4×4 matrix of overlapping 8×5.5 cm2 flexible receive loops [16], (Rapid Biomedical) for reception. Additional tests were performed to check the compatibility of the whole-body coil with the 16-channel receive array, e.g., scattering parameters measurements, and a heating test using fibre-optic temperature probes (Neoptix Inc, Quebec, Canada) and an infrared camera (FLIR Systems Inc, Wilsonville, Oregon). Ultimately, scanning human subjects with the receive array inside the whole-body transmit coil was found to be safe.

Eight healthy volunteers (one female, mean age 28 ± 5 years, ages ranging from 21 to 35 years) were approached between October 2016 and January 2017 and all were consecutively recruited for our three in vivo experiments. A written informed consent was obtained in compliance with ethical and legal requirements. Oxford Central University Research Ethics Committee provided approval for this technical development work. No further demographic characteristics were recorded for the recruited volunteers. The individual whose photograph is shown in Fig 1 has given written informed consent (as outlined in PLOS consent form) to publish this photograph. All subjects were positioned supine inside the whole-body coil and CINE 1H FLASH images were acquired using a single 1H, transmit/receive, fractionated dipole antenna RF-coil (MR Coils) positioned over the heart. The 1H RF-coil was then replaced with the 16-channel receive array for the 31P experiments. The first experiment recorded high-quality 3D-resolved spectra using our established cardiac 31P-MRS protocol. Specifically: acquisition-weighted 3D-UTE-CSI [17] spectra were acquired over a 16×16×8 matrix covering a 500×500×400 mm3 field-of-view (FOV) in three volunteers, using a 1 ms long amplitude-modulated excitation pulse [8]. Eight averages using TR = 1 s were acquired in 23 minutes 52 seconds.

The second experiment recorded metabolite maps and tested the performance of amplitude-modulated RF “saturation bands”. Four volunteers were examined in the second experiment using two acquisition-weighted, transverse 2D-UTE-CSI experiments with a 24×24 matrix over 500×500 mm3 FOV, and 60 mm slice thickness. Slice selective 2.5 ms long sinc pulses were used for excitation. Relatively short TR = 300 ms was used allowing for 32 averages within 19 minutes 26 seconds. Two amplitude-modulated saturation bands (10 ms duration) were used to suppress signal from chest muscles in one of the acquisitions.

The third (and final) experiment recorded B1+ field maps to quantify the transmit B1+ homogeneity of the new whole-body RF-coil in vivo. Four subjects underwent B1+ field mapping using a Bloch-Siegert sequence [13] with similar 2D resolution and excitation as in the second experiment. The Fermi pulse (8 ms duration), placed at ±2 kHz from PCr, required a minimum TR = 400 ms. The number of averages was 32 or 48, leading to a scan time of 26 minutes or 38 minutes for each of the ±2 kHz Fermi pulse frequency offsets.

Signals from individual receive elements were combined using whitened singular value decomposition [18], and the combined spectra were fitted using a Matlab (MathWorks, Natick, MA) implementation of the AMARES time domain fitting routine [19]. The B1+ field maps were calculated in all voxels with sufficient SNR, as defined in [13].

Results

The results of our simulations are depicted in Fig 1A–1D. The simulated local 10 g SAR efficiency, global SAR efficiency and global head SAR efficiency of the designed coil were 3.4, 0.28 and 0.25 W/kg/μT2, respectively. Our simulated worst-case local 10 g SAR in the body was 0.145 W/kg and global body SAR was 0.013 W/kg. The mean simulated B1+ for 1 W delivered power was 0.160 ± 0.075 μT for the heart and 0.122 ± 0.068 μT for liver (mean ± standard deviation for voxels of tissue type “heart” or “liver” in the 3D results). The coefficients of variation (CV), i.e. the standard deviation divided by the mean, for simulated B1+ across the “heart” type voxels were 46.7% for the new birdcage coil (55cm-diameter) and 129.0% for the rectangular surface coil.

The results of the phantom SNR and peak B1+ comparison between the coils are given in Table 1. The SNR of the combined whole-body coil transmit and 16-channel receive was 2.6 times higher in comparison to the whole-body coil in transceiver mode. The peak B1+ of the combined whole-body transmit and local receivers at a 10 cm depth below the coil was 3.5 times lower compared to the quadrature surface coil, however, the SNR achieved was comparable.

thumbnail
Table 1. Comparison of the SNR and peak B1+ between three combinations of RF transmit and receive coils in a phantom placed 10 cm away from the surface RF-coil.

https://doi.org/10.1371/journal.pone.0187153.t001

Fig 2 depicts transverse 2D spatial distribution in vivo maps of PCr and 2,3-diphosphoglycerate (2,3-DPG) acquired using the body coil in transceiver mode and with the combined whole-body transmit and 16-channel receive setup. While the detection of signal from the heart region is challenging if the whole-body coil is used in transceiver mode, i.e. no 2,3-DPG signal visible, the heart is clearly delineated when the combined setup is used.

thumbnail
Fig 2. Metabolite maps.

Transverse in vivo maps of (a, b) PCr and (c, d) 2,3-DPG signals acquired using the body coil in (a, c) transceiver mode and (b, d)using the combined whole-body transmit and 16-channel local receive setup. All maps are overlaid on 1H localizer images registered to the CSI grid. Note that the signals outside the body are from a concentrated fiducial mounted in the centre of the receive array.

https://doi.org/10.1371/journal.pone.0187153.g002

Representative in vivo spectra acquired with the 3D-UTE-CSI protocol in human heart and liver using the combined setup are depicted in Fig 3. The applied non-adiabatic saturation bands suppressed an average of 71 ± 4% of the PCr signal in the chest muscles (target region), while suppressing an average of 19 ± 3% of the PCr signal in the heart (not targeted); the suppression efficiency is depicted in Fig 3C and 3D.

thumbnail
Fig 3. Acquired 31P-MR spectra.

Representative in vivo 31P-MR spectra acquired with the 3D-UTE-CSI protocol in (a) human heart and (b) liver using the combined coil setup. (c) PCr metabolite map in transverse view through the chest without the saturation pulses. (d) Shows the ratio of PCr signal from acquisitions with and without the saturation bands applied to reduce skeletal muscle signal.

https://doi.org/10.1371/journal.pone.0187153.g003

The B1+ field was calculated in 52 ± 8 voxels satisfying the SNR criteria per subject. The mean measured CV for B1+ was 38.4%. The resulting mean measured 31P B1+ in vivo was 10.4 ± 2.7 μT. A representative in vivo B1+ map and a histogram of B1+ variability are shown in Fig 4.

thumbnail
Fig 4. In vivo B1+ mapping.

(a) Representative Bloch-Siegert, 2D, transverse B1+ map overlaid on 1H localizer image registered to the CSI grid. (b) A histogram of measured B1+ variability across all four subjects normalized to individual subject median B1+. A fitted normal distribution function (red line) is also depicted. All voxels with sufficient SNR were used for analysis.

https://doi.org/10.1371/journal.pone.0187153.g004

Discussion

In our study, we propose a new easily removable whole-body RF transmit coil combined with a 16-element receive array for cardiac 31P-MR at 7T. This setup provides homogeneous excitation through the whole chest without the need for adiabatic excitation pulses, as demonstrated by our CST simulations and 31P B1 maps measured in vivo. In addition, the receive array increases the SNR of our setup compared to the body coil in transmit/receive mode, allowing high-resolution 31P-MRSI experiments in vivo. The increased 31P B1+ homogeneity allows the use of conventional amplitude-modulated pulses to suppress chest wall muscles instead of the SAR demanding BISTRO approach [20] that has previously been used with surface transmit coils [8].

Our new coil was designed to easily move in and out of the bore of a Magnetom 7T MR scanner using the system’s motorized patient table. Therefore, the maximal inner diameter of the transmit coil was limited by the scanner bore diameter (58 cm). Our constructed whole-body coil had an inner diameter of 55 cm. Although this means reduction of the available patient space, it is still larger than the diameter of the previously reported whole-body coil that had a 52 cm diameter [15], and since it is integrated with the extension of the patient table, this allows us to scan a range of subjects in comfort. Additionally, the upper part of the birdcage coil is easily detachable, which facilitates straightforward positioning of subjects before their scan and rapid evacuation if a patient became acutely unwell. Using a volume transmit coil improved the transmit uniformity as expected, quantified by simulations and confirmed by our Bloch-Siegert 31P measurement in vivo. Our simulated SAR efficiency values were similar to the previously reported ones [15], i.e. 3.4 vs. 3.8 (local), 0.28 vs. 0.24 (global body) and 0.25 vs. 0.33 W/kg/μT2 (global head).

To increase receive sensitivity, we combined the whole-body transmit coil with a 16-element receive array. No changes in the S11 and S12 parameters of whole-body coil ports were observed using a network analyser (8712C, Hewlett Packard, Palo Alto, California) when the 16-channel array was inserted (regardless of its position). Adding the receive array led to a 2.6-fold increase in achieved SNR in phantom experiments, in comparison to the use of the whole-body coil in transceiver mode alone. This is comparable to the improvement seen on a 3T TIM Trio scanner (Siemens) between body coil 1H MRI and using half of a 32-channel cardiac receive array (Invivo, Gainesville, Florida). A further increase in SNR might be gained by placing another receive array beneath the volunteer. Active detuning of the body coil during signal acquisition could also potentially lead to an increase in SNR. Every increase in single-element SNR will also lead to an increase in the precision of the complex “weights” used in coil combination algorithms, e.g., WSVD. The final SNR will therefore increase both due to the increased single-element SNR, but even more because of the better complex “weights” estimation in the combination step [18].

On the other hand, the peak B1 achieved by the whole-body transmit coil in the phantom experiments at a depth of 10 cm was 3.5 times lower than that of our dedicated quadrature surface transmit coil as expected. However, the B1+ homogeneity of the whole-body transmit coil, allowing for short-TR scans with uniform Ernst-angle excitation or multi-echo readout [21] compensates for this. Furthermore, it may be feasible to drive the whole-body coil with a high-power 31P RF-amplifier, such as the 35 kW 123 MHz amplifiers used for 1H-MRI on Siemens 3T scanners. Dedicated low-loss transmit cabling could also be used. We estimate that these changes would give peak B1+ output comparable to our quadrature surface coil at 10 cm depth, while also retaining the coverage and uniformity advantages of the whole-body coil.

The in vivo data demonstrate that while our whole-body RF-coil in transceiver mode could be used to acquire PCr signal from skeletal muscles, its SNR is too low to allow detection of metabolites of lower concentration, e.g., 2,3-DPG. However, in combination with the 16-element receive array, good quality spectra and well-resolved maps of PCr and 2,3-DPG can be acquired in vivo. As expected, the PCr signal was localized to the skeletal muscle wall as well as the heart, while the 2,3-DPG signal was restricted only to the ventricular blood pools. This confirms the high spatial resolution of the acquired 31P-MRSI maps. The acquired in vivo 31P B1+ maps showed a high level of uniformity with a coefficient of variation <39% across all voxels with sufficient SNR. To demonstrate the use of this B1+ uniformity of the whole-body coil, we showed the effectiveness (>70%) of conventional amplitude-modulated saturation bands to suppress the signal from chest and abdominal muscles that can otherwise contaminate cardiac 31P-MRS data [8, 22].

In conclusion, we have designed, constructed and tested an easily removable, whole-body transmit RF-coil and its combination with 16-element receive array that is straightforward and comfortable to use for cardiac 31P-MR at 7T. This apparatus allowed us to measure in vivo the homogeneity of the 31P transmit field, confirming the results of electromagnetic field simulations. It allows us to record anatomically-consistent 31P metabolite maps, and to use saturation bands with amplitude-modulated pulses (and hence low SAR demands) to suppress signals from skeletal muscle. This combination of hardware is a step towards regionally-resolved, whole-heart cardiac 31P-MRS studies at 7T.

Acknowledgments

The authors would like to thank Karsten Wicklow and Ulrich Fontius from Siemens Healthcare GmbH for their excellent support.

References

  1. 1. Bottomley PA, Charles HC, Roemer PB, Flamig D, Engeseth H, Edelstein WA, et al. Human in vivo phosphate metabolite imaging with 31P NMR. Magnetic resonance in medicine. 1988;7(3):319–36. pmid:3205148
  2. 2. Valkovič L, Chmelík M, Krššák M. In-vivo 31P-MRS of skeletal muscle and liver: a way for non-invasive assessment of their metabolism. Anal Biochem. 2017;529:193–215. pmid:28119063
  3. 3. Neubauer S, Horn M, Cramer M, Harre K, Newell JB, Peters W, et al. Myocardial phosphocreatine-to-ATP ratio is a predictor of mortality in patients with dilated cardiomyopathy. Circulation. 1997;96(7):2190–6. pmid:9337189
  4. 4. Bottomley PA. NMR Spectroscopy of the human heart. In: Harris RK, Wasylishen RE, editors. Encyclopedia of magnetic resonance. Chichester: John Wiley; 2009.
  5. 5. Stoll VM, Clarke WT, Levelt E, Liu A, Myerson SG, Robson MD, et al. Dilated Cardiomyopathy: Phosphorus 31 MR Spectroscopy at 7 T. Radiology. 2016;281(2):409–17. pmid:27326664
  6. 6. Levelt E, Rodgers CT, Clarke WT, Mahmod M, Ariga R, Francis JM, et al. Cardiac energetics, oxygenation, and perfusion during increased workload in patients with type 2 diabetes mellitus. European heart journal. 2016;37(46):3461–9. pmid:26392437
  7. 7. Rider OJ, Francis JM, Ali MK, Holloway C, Pegg T, Robson MD, et al. Effects of catecholamine stress on diastolic function and myocardial energetics in obesity. Circulation. 2012;125(12):1511–9. pmid:22368152
  8. 8. Rodgers CT, Clarke WT, Snyder C, Vaughan JT, Neubauer S, Robson MD. Human cardiac 31P magnetic resonance spectroscopy at 7 Tesla. Magnetic resonance in medicine. 2014;72(2):304–15. pmid:24006267
  9. 9. Šedivý P, Kipfelsberger MC, Dezortová M, Krššák M, Drobný M, Chmelík M, et al. Dynamic 31P MR spectroscopy of plantar flexion: influence of ergometer design, magnetic field strength (3 and 7 T), and RF-coil design. Medical physics. 2015;42(4):1678–89. pmid:25832057
  10. 10. Purvis LAB, Clarke WT, Valkovič L, Levick C, Pavlides M, Barnes E, et al. Phosphodiester content measured in human liver by in vivo 31P magnetic resonance spectroscopy at 7T. Magnetic resonance in medicine. 2017; pmid:28244131
  11. 11. Valkovič L, Chmelík M, Ukropcová B, Heckmann T, Bogner W, Frollo I, et al. Skeletal muscle alkaline Pi pool is decreased in overweight-to-obese sedentary subjects and relates to mitochondrial capacity and phosphodiester content. Scientific reports. 2016;6:20087. pmid:26838588
  12. 12. Chmelík M, Považan M, Jiru F, Just Kukurová I, Dezortová M, Krššák M, et al. Flip-angle mapping of 31P coils by steady-state MR spectroscopic imaging. Journal of magnetic resonance imaging: JMRI. 2014;40(2):391–7. pmid:24925600
  13. 13. Valkovic L, Clarke WT, Purvis LA, Schaller B, Robson MD, Rodgers CT. Adiabatic excitation for 31P-MR spectroscopy in the human heart at 7T: A feasibility study. Magnetic resonance in medicine. 2016.
  14. 14. Valkovič L, Clarke WT, Purvis LA, Schaller B, Robson MD, Rodgers CT. Adiabatic excitation for 31P MR spectroscopy in the human heart at 7T: A feasibility study. Magnetic resonance in medicine. 2016; pmid:28000961
  15. 15. Loring J, van der Kemp WJ, Almujayyaz S, van Oorschot JW, Luijten PR, Klomp DW. Whole-body radiofrequency coil for 31P MRSI at 7T. NMR in biomedicine. 2016;29(6):709–20. pmid:27037615
  16. 16. Rodgers CT, Clarke WT, Berthel D, Neubauer S, Robson MD, editors. A 16-element receive array for human cardiac 31P MR spectroscopy at 7T. 22nd Joint Annual Meeting of ISMRM & ESMRMB; 2014; Milan, Italy.
  17. 17. Robson MD, Tyler DJ, Neubauer S. Ultrashort TE chemical shift imaging (UTE-CSI). Magnetic resonance in medicine. 2005;53(2):267–74. pmid:15678544
  18. 18. Rodgers CT, Robson MD. Coil combination for receive array spectroscopy: Are data-driven methods superior to methods using computed field maps? Magnetic resonance in medicine. 2016;75(2):473–87. pmid:25820303
  19. 19. Purvis LAB, Clarke WT, Biasiolli L, Valkovič L, Robson MD, Rodgers CT. OXSA: An open-source magnetic resonance spectroscopy analysis toolbox in MATLAB. PLoS ONE. 2017;12(9):e0185356. pmid:28938003
  20. 20. Luo Y, de Graaf RA, DelaBarre L, Tannus A, Garwood M. BISTRO: an outer-volume suppression method that tolerates RF field inhomogeneity. Magnetic resonance in medicine. 2001;45(6):1095–102. pmid:11378888
  21. 21. van der Kemp WJ, Boer VO, Luijten PR, Stehouwer BL, Veldhuis WB, Klomp DW. Adiabatic multi-echo 31P spectroscopic imaging (AMESING) at 7 T for the measurement of transverse relaxation times and regaining of sensitivity in tissues with short T2 values. NMR in biomedicine. 2013;26(10):1299–307. pmid:23553945
  22. 22. Schaller B, Clarke WT, Neubauer S, Robson MD, Rodgers CT. Suppression of skeletal muscle signal using a crusher coil: A human cardiac 31P-MR spectroscopy study at 7 tesla. Magnetic resonance in medicine. 2016;75(3):962–72. pmid:25924813