ABSTRACT

The introduction of computer-assisted navigation systems has played a significant role in assuring the integration and consistent intraoperative use of radiological information. We used a frameless stereotactic navigation system to treat 62 patients with a variety of skull base pathologies. The optoelectric appliance uses digital imaging information to locate surgical instruments in the operative area. The aim of this study was to evaluate the clinical accuracy, practicality, and impact of this navigation system on otolaryngological procedures. In conjunction with rigid head fixation and bone-anchored registration markers, the precision of registration was 0.8 mm and the accuracy of clinical measurements was less than 2 mm. With conventional fiducials and flexible head positioning, deviations were as large as 4.5 mm. The additional use of surface registration increased the precision of registration. Preoperative preparations took 15 to 35 minutes, depending on the complexity of the planning. Intraoperative computer support is an important aid to a surgeon's orientation, especially when a patient's anatomy is atypical. Navigation systems will likely improve the quality of surgery and facilitate training.

REFERENCES

  • 1 Freysinger W, Gunkel A R, Martin A, Bale R J, Vogele M, Thumfart W F. Advancing ear, nose, and throat computer-assisted surgery with the arm-based ISG viewing wand: the stereotactic suction tube.  Laryngoscope . 1997;  107 690-693
  • 2 Hassfeld S, Mühling J, Zöller J. Intraoperative navigation in oral and maxillofacial surgery.  Int J Oral Maxillofac Surg . 1995;  24 111-119
  • 3 Kamimura M, Ebara S, Itoh H, Tateiwa Y, Kinoshita T, Takaoka K. Accurate pedicle screw insertion under the control of a computer-assisted image guiding system: laboratory test and clinical study.  J Orthop Sci . 1999;  4 197-206
  • 4 Schlöndorff G, Mosges R, Meyer-Ebrecht B, Krybus W, Adams L. Ein neues Verfahren in der Kopf- und Halschirurgie [CAS (computer assisted surgery)].  HNO . 1989;  37 173-179
  • 5 Wirtz C R, Knauth M, Hassfeld S. Neuronavigation-first experiences with three different commercially available systems.  Zentralbl Neurochir . 1998;  59 14-22
  • 6 Fried M P, Morrison P R. Computer-augmented endoscopic sinus surgery.  Otolaryngol Clin North Am . 1998;  31 331-340
  • 7 Lenarz T. Computer aided surgery: comparison of different systems.  Skull Base Surg . 1997 (Suppl 2);  7 18
  • 8 Husstedt H, Heermann R, Becker H. Contribution of low-dose CT-scan protocols to the total positioning error in computer-assisted surgery.  Comput Aided Surg . 1999;  4 275-280
  • 9 Lenarz T, Heermann R. Editorial: Image-guided and computer-aided surgery in otology and neurotology: is there already a need for it?.  Am J Otol . 1999;  20 143-144
  • 10 Germano I M, Villalobos H, Silvers A, Post K D. Clinical use of the otical digitizer for intracranial neuronavigation.  Neurosurgery . 1999;  45 261-269
  • 11 Reinhardt H, Meyer H, Amrein E. A computer assisted device for the intra-operative CT-controlled localization of brain tumours.  Eur Surg Res . 1988;  20 51-58
  • 12 Heermann R, Enders S, Husstedt H, Lenarz T. Intraoperative navigation systems in anterior skull base surgery.  Skull Base Surg . 1997;  7(Suppl 2) 36
  • 13 Vorbeck F, Cartellieri M, Ehrenberger K, Imhof H. Experiences in intraoperative computer-aided navigation in ENT sinus surgery with the Aesculap navigation system.  Comput Aided Surg . 1998;  3 306-311
  • 14 Kikinis R, Gleason P L, Jolesz F A. Surgical planning using computer-assisted three-dimensional reconstructions. In: Taylor RH, Lavallée S, Burdea GC, Mösges R, eds. Computer Integrated Surgery, Technology and Clinical Applications Cambridge, Mass: MIT Press 1996: 147-154
  • 15 Gunkel A R, Freysinger W, Thumfart W F. Computer-assisted surgery in the frontal and maxillary sinus.  Laryngoscope . 1997;  107 631-633
  • 16 Fried M P, Kleefield J, Gopal H, Reardon E, Ho B T, Kuhn F A. Image-guided endoscopic surgery: results of accuracy and performance in a multicenter clinical study using an electromagnetic tracking system.  Laryngoscope . 1997;  107 594-601
  • 17 Edwards P J, Hawkes D J, Hill D L. Augmentation of reality using an operating microscope for otolaryngology and neurosurgical guidance.  J Image Guid Surg . 1995;  1 172-178
  • 18 Hauser R, Westermann B, Probst R. Non-invasive tracking of patient's head movements during computer-assisted intranasal microscopic surgery.  Laryngoscope . 1997;  107 491-499
  • 19 Neumann Jr M A, Pasquale-Niebles K, Bhuta T, Sillers M J. Image-guided transnasal endoscopic surgery of the paranasal sinuses and anterior skull base.  Am J Rhinol . 1999;  13 449-454
  • 20 Metson R B, Cosenza M J, Cunningham M J, Randolph G W. Physicians' experience with an optical image guidance system for sinus surgery.  Laryngoscope . 2000;  110 972-976
  • 21 Bale R J, Vogele M, Freysinger W. Minimally invasive head holder to improve the performance of frameless stereotactic surgery.  Laryngoscope . 1997;  107 373-377
  • 22 Caversaccio M, Lädrach K, Häusler R. Konzept eines rahmenlosen bildinteraktiven Navigationssystems für die Schädelbasis-, Nasen- und Nasennebenhöhlenchirurgie.  Otorhinolarngol Nova . 1997;  7 121-126
  • 23 Anon J B. Computer-aided endoscopic sinus surgery.  Laryngoscope . 1998;  108 949-961
  • 24 Casiano R R, Numa Jr A W. Efficacy of computed tomographic image-guided endoscopic sinus surgery in residency training programs.  Laryngoscope . 2000;  110 1277-1282
  • 25 Smith K R, Frank K J, Buchholz R D. The NeuroStation™-a highly accurate, minimally invasive solution to frameless stereotactic neurosurgery.  Comput Med Imaging Graph . 1994;  18 247-256
  • 26 Metson R, Cosenza M, Gliklich R E, Montgomery W W. The role of image-guidance systems for head and neck surgery.  Arch Otolaryngol Head Neck Surg . 1999;  125 1100-1104
  • 27 Gunkel A R, Vogele M, Martin A, Bale R J, Thumfart W F, Freysinger W. Computer aided surgery in the petrous bone.  Laryngoscope . 1999;  109 1793-1799
  • 28 Sargent E W, Buchholz R D. Middle cranial fossa surgery with image-guided instrumentation.  Otolaryngol Head Neck Surg . 1997;  117 131-134
  • 29 Olson G, Citardi M J. Image-guided functional endoscopic sinus surgery.  Otolaryngol Head Neck Surg . 2000;  123 188-194
  • 30 Klimek L, Wenzel M, Mösges R. Computer assisted orbital surgery.  Ophthalmic Surg . 1993;  24 411-417
  • 31 Mösges R, Klimek L. Computer-assisted surgery of the paranasal sinuses.  J Otolaryngol . 1993;  22 69-71
  • 32 Metson R, Gliklich R E, Cosenza M. A comparison of image guidance systems for sinus surgery.  Laryngoscope . 1998;  108 1164-1170
  • 33 Mann W, Klimek L. Indications for computer-assisted surgery in otolaryngology.  Comput Aided Surg . 1998;  3 202-204
  • 34 Anon J B, Klimek L, Mösges R, Zinreich S J. Computer-assisted endoscopic sinus surgery. An international review.  Otolaryngol Clin North Am . 1997;  30 389-401
    >