Open Access 27-09-2024 | Hip-TEP | Original Paper

Reconstruction of severe acetabular defects (Paprosky type III A) in total hip arthroplasty using modular tantalum augments in combination with a cemented cup

Authors: David Spranz, Lisa-Marie Müller, Raphael Trefzer, Pit Hetto, Timo Nees, Tobias Renkawitz, Tilman Walker, Tobias Reiner

Published in: International Orthopaedics

Login to get access

Abstract

Purpose

Acetabular defect reconstruction can be a complex and challenging surgical procedure, with stable long-term fixation of the implants remaining the ultimate goal. The purpose of this study was (1) to evaluate the radiological and clinical outcome of complex acetabular reconstruction surgery with the use of modular tantalum TM augments in combination with cemented revision cups; (2) to investigate blood tantalum concentrations in these patients; and (3) to report complications and mechanisms of failure related to this procedure at mid-term follow-up (mean 4.5 years).

Methods

We retrospectively reviewed 29 patients (29 hips) with severe acetabular bone loss (Paprosky type III A) reconstructed using a modular tantalum TM augment in combination with a cemented cup. We evaluated the implant survival and the radiological and clinical outcomes after a mean follow-up of 4.5 years (SD 2.2; range 8.4 – 2.1 years) using patient reported outcome scores (PROMs). Blood samples were analysed regarding tantalum concentration and compared with a control group.

Results

The cumulative survival rate at 4.5 years with the endpoint “revision of the acetabular component for any reason” was 96.2% (95% Confidence Interval 75.7–99.5). The PROMs improved significantly up to the latest follow-up, and radiographic data showed only one patient with signs of initial implant migration with a broken screw and a change of the position of the augment and the cup. Mean blood tantalum concentrations were significantly higher in the study group (0.16 µg/L) compared to the control group (0.002 µg/L) (P < 0.001).

Conclusions

This study has demonstrated good mid-term (mean 4.5 years) clinical and radiological outcomes of modular tantalum TM augments in combination with a cemented cup for the reconstruction of major acetabular defects. Mean blood tantalum concentrations were increased in patients with stable tantalum implants compared to healthy controls.
Literature
1.
go back to reference Alqwbani M, Wang Z, Wang Q et al (2022) Porous tantalum shell and augment for acetabular defect reconstruction in revision total hip arthroplasty: a mid-term follow-up study. Int Orthop 46:1515–1520CrossRefPubMed Alqwbani M, Wang Z, Wang Q et al (2022) Porous tantalum shell and augment for acetabular defect reconstruction in revision total hip arthroplasty: a mid-term follow-up study. Int Orthop 46:1515–1520CrossRefPubMed
2.
go back to reference Beckmann NA, Bitsch RG, Schonhoff M et al (2020) Comparison of the primary stability of porous tantalum and titanium acetabular revision constructs. Materials (Basel) 13(7):1783CrossRefPubMed Beckmann NA, Bitsch RG, Schonhoff M et al (2020) Comparison of the primary stability of porous tantalum and titanium acetabular revision constructs. Materials (Basel) 13(7):1783CrossRefPubMed
3.
go back to reference Brooker AF, Bowerman JW, Robinson RA et al (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55:1629–1632CrossRefPubMed Brooker AF, Bowerman JW, Robinson RA et al (1973) Ectopic ossification following total hip replacement. Incidence and a method of classification. J Bone Joint Surg Am 55:1629–1632CrossRefPubMed
4.
go back to reference Bruggemann A, Mallmin H, Bengtsson M et al (2020) Safety of use of tantalum in total hip arthroplasty. J Bone Joint Surg Am 102:368–374CrossRefPubMed Bruggemann A, Mallmin H, Bengtsson M et al (2020) Safety of use of tantalum in total hip arthroplasty. J Bone Joint Surg Am 102:368–374CrossRefPubMed
5.
go back to reference Ciolli G, Mesnard G, Deroche E et al (2022) Is cemented dual-mobility cup a reliable option in primary and revision total hip arthroplasty: a systematic review. J Pers Med 13(1):81CrossRefPubMedPubMedCentral Ciolli G, Mesnard G, Deroche E et al (2022) Is cemented dual-mobility cup a reliable option in primary and revision total hip arthroplasty: a systematic review. J Pers Med 13(1):81CrossRefPubMedPubMedCentral
6.
go back to reference Ciriviri J, Nestorovski Z, Talevski D et al (2019) Treatment of acetabular defects with porous metal augments in revision hip surgery. Pril (Makedon Akad Nauk Umet Odd Med Nauki) 40:33–39PubMed Ciriviri J, Nestorovski Z, Talevski D et al (2019) Treatment of acetabular defects with porous metal augments in revision hip surgery. Pril (Makedon Akad Nauk Umet Odd Med Nauki) 40:33–39PubMed
7.
go back to reference Dwivedi C, Gokhale S, Khim HG et al (2017) Acetabular defect reconstruction with trabecular metal augments: study with minimum one-year follow-up. Hip Pelvis 29:168–175CrossRefPubMedPubMedCentral Dwivedi C, Gokhale S, Khim HG et al (2017) Acetabular defect reconstruction with trabecular metal augments: study with minimum one-year follow-up. Hip Pelvis 29:168–175CrossRefPubMedPubMedCentral
8.
go back to reference Eachempati KK, Malhotra R, Pichai S et al (2018) Results of trabecular metal augments in Paprosky IIIA and IIIB defects: A multicentre study. Bone Joint J 100-B:903–908CrossRefPubMed Eachempati KK, Malhotra R, Pichai S et al (2018) Results of trabecular metal augments in Paprosky IIIA and IIIB defects: A multicentre study. Bone Joint J 100-B:903–908CrossRefPubMed
9.
go back to reference Flecher X, Appy B, Parratte S et al (2017) Use of porous tantalum components in Paprosky two and three acetabular revision. A minimum five-year follow-up of fifty one hips. Int Orthop 41:911–916CrossRefPubMed Flecher X, Appy B, Parratte S et al (2017) Use of porous tantalum components in Paprosky two and three acetabular revision. A minimum five-year follow-up of fifty one hips. Int Orthop 41:911–916CrossRefPubMed
10.
go back to reference Han Q, Wang C, Chen H et al (2019) Porous Tantalum and Titanium in Orthopedics: A Review. ACS Biomater Sci Eng 5:5798–5824CrossRefPubMed Han Q, Wang C, Chen H et al (2019) Porous Tantalum and Titanium in Orthopedics: A Review. ACS Biomater Sci Eng 5:5798–5824CrossRefPubMed
11.
go back to reference Huang G, Pan ST, Qiu JX (2021) The clinical application of porous tantalum and its new development for bone tissue engineering. Materials (Basel) 14(10):2647CrossRefPubMedPubMedCentral Huang G, Pan ST, Qiu JX (2021) The clinical application of porous tantalum and its new development for bone tissue engineering. Materials (Basel) 14(10):2647CrossRefPubMedPubMedCentral
12.
go back to reference Kato H, Nakamura T, Nishiguchi S et al (2000) Bonding of alkali- and heat-treated tantalum implants to bone. J Biomed Mater Res 53:28–35CrossRefPubMed Kato H, Nakamura T, Nishiguchi S et al (2000) Bonding of alkali- and heat-treated tantalum implants to bone. J Biomed Mater Res 53:28–35CrossRefPubMed
13.
go back to reference Koob S, Scheidt S, Randau TM et al (2017) Biological downsizing: Acetabular defect reconstruction in revision total hip arthroplasty. Orthopade 46:158–167CrossRefPubMed Koob S, Scheidt S, Randau TM et al (2017) Biological downsizing: Acetabular defect reconstruction in revision total hip arthroplasty. Orthopade 46:158–167CrossRefPubMed
14.
go back to reference Levine BR, Sporer S, Poggie RA et al (2006) Experimental and clinical performance of porous tantalum in orthopedic surgery. Biomaterials 27:4671–4681CrossRefPubMed Levine BR, Sporer S, Poggie RA et al (2006) Experimental and clinical performance of porous tantalum in orthopedic surgery. Biomaterials 27:4671–4681CrossRefPubMed
15.
go back to reference O’neill CJ, Creedon SB, Brennan SA et al (2018) Acetabular revision using trabecular metal augments for Paprosky type 3 defects. J Arthroplasty 33:823–828CrossRefPubMed O’neill CJ, Creedon SB, Brennan SA et al (2018) Acetabular revision using trabecular metal augments for Paprosky type 3 defects. J Arthroplasty 33:823–828CrossRefPubMed
16.
go back to reference Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9:33–44CrossRefPubMed Paprosky WG, Perona PG, Lawrence JM (1994) Acetabular defect classification and surgical reconstruction in revision arthroplasty. A 6-year follow-up evaluation. J Arthroplasty 9:33–44CrossRefPubMed
17.
go back to reference Rodushkin I, Engstrom E, Stenberg A et al (2004) Determination of low-abundance elements at ultra-trace levels in urine and serum by inductively coupled plasma-sector field mass spectrometry. Anal Bioanal Chem 380:247–257CrossRefPubMed Rodushkin I, Engstrom E, Stenberg A et al (2004) Determination of low-abundance elements at ultra-trace levels in urine and serum by inductively coupled plasma-sector field mass spectrometry. Anal Bioanal Chem 380:247–257CrossRefPubMed
19.
go back to reference Ware J Jr, Kosinski M, Keller SD (1996) A 12-item short-form health survey: construction of scales and preliminary tests of reliability and validity. Med Care 34:220–233CrossRefPubMed Ware J Jr, Kosinski M, Keller SD (1996) A 12-item short-form health survey: construction of scales and preliminary tests of reliability and validity. Med Care 34:220–233CrossRefPubMed
20.
go back to reference Ying J, Cheng L, Li J et al (2023) Treatment of acetabular bone defect in revision of total hip arthroplasty using 3D printed tantalum acetabular augment. Orthop Surg 15:1264–1271CrossRefPubMedPubMedCentral Ying J, Cheng L, Li J et al (2023) Treatment of acetabular bone defect in revision of total hip arthroplasty using 3D printed tantalum acetabular augment. Orthop Surg 15:1264–1271CrossRefPubMedPubMedCentral
Metadata
Title
Reconstruction of severe acetabular defects (Paprosky type III A) in total hip arthroplasty using modular tantalum augments in combination with a cemented cup
Authors
David Spranz
Lisa-Marie Müller
Raphael Trefzer
Pit Hetto
Timo Nees
Tobias Renkawitz
Tilman Walker
Tobias Reiner
Publication date
27-09-2024
Publisher
Springer Berlin Heidelberg
Keywords
Hip-TEP
Hip-TEP
Published in
International Orthopaedics
Print ISSN: 0341-2695
Electronic ISSN: 1432-5195
DOI
https://doi.org/10.1007/s00264-024-06334-x