Reduction of radiation and contrast agent exposure in a cryoballoon ablation procedure with integration of electromagnetic mapping and intracardiac echocardiography: a single center experience

Authors: Joyce Maalouf, Hoyle L. Whiteside, Ajay Pillai, Abdullah Omar, Adam Berman, Samir Saba, Haitham Hreibe

Published in: Journal of Interventional Cardiac Electrophysiology | Issue 3/2020

Login to get access

Abstract

Background

Pulmonary vein isolation (PVI) with cryoballoon ablation (CBA) is routinely guided by fluoroscopy and utilizes contrast injection to ensure catheter positioning and pulmonary vein occlusion. Non-fluoroscopic imaging techniques including electromagnetic mapping (EM) and intracardiac echocardiography (ICE) have demonstrated reduced fluoroscopy times and contrast exposure. Utilization of color flow Doppler to evaluate vein occlusion with the balloon has not been evaluated as an alternative to contrast injection. In this study we evaluate the effectiveness of cryoablation guided by EM and ICE along with color Doppler to achieve PVI.

Methods

We designed a retrospective cohort study comparing patients who were treated before and after implementation of EM (Carto 3, Biosense Webster) and ICE during CBA (AF Solutions, Medtronic). We analyzed patients receiving CBA with fluoroscopy plus EM and ICE (group 2; N = 24) versus fluoroscopy alone (group 1; N = 25). Procedural success was defined as freedom from atrial fibrillation or other atrial arrhythmias at 1 year post ablation. Primary outcomes were radiation time and contrast exposure.

Results

Procedural success was achieved in all cases. Total fluoroscopy time was reduced from 22.4 ± 9.8 min to 8.9 ± 5.1 min (P < 0.001) in patients receiving CBA guided by EM and ICE. Furthermore, exposure to contrast media was significantly lower at 75.4 ± 24.1 ml and 16.5 ± 21.1 ml (P ≤ 0.001) in group 1 and group 2, respectively. Neither the number of required cryotherapy treatments nor procedure duration was negatively impacted by the implementation of non-fluoroscopic techniques. The 1-year success rate was identical between both groups at 72% and 79%. There was no difference in complication rates.

Conclusion

This single-center cohort study demonstrates that CBA guided by EM and ICE can markedly reduce radiation and contrast exposure with excellent rates of acute PVI. This technique may be particularly effective in patients sensitive to intravenous contrast exposure.
Literature
1.
go back to reference Guhl EN, Adelstein E, Voigt A, Wang NC, Saba S, Jain SK. Impact of 3D mapping on procedural characteristics and outcomes in cryoballoon pulmonary vein isolation for atrial fibrillation. J Interv Card Electrophysiol. 2018;51(1):71–5.CrossRef Guhl EN, Adelstein E, Voigt A, Wang NC, Saba S, Jain SK. Impact of 3D mapping on procedural characteristics and outcomes in cryoballoon pulmonary vein isolation for atrial fibrillation. J Interv Card Electrophysiol. 2018;51(1):71–5.CrossRef
2.
go back to reference Kuck K-H, Brugada J, Furnkkranz A, Metzner A, Ouyang F, Chun KRJ, et al. Tondo C; FIRE AND ICE investigators. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl JMed. 2016;374(23):2235–45.CrossRef Kuck K-H, Brugada J, Furnkkranz A, Metzner A, Ouyang F, Chun KRJ, et al. Tondo C; FIRE AND ICE investigators. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl JMed. 2016;374(23):2235–45.CrossRef
3.
go back to reference Camm AJ, Lyp GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, Hohnloser SH, Hindricks G, Kirchhof P; ESC Committee for Practice Guidelines-CPG; Document Reviewers. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association. Europace 2012;14:1385–1413. Camm AJ, Lyp GY, De Caterina R, Savelieva I, Atar D, Hohnloser SH, Hohnloser SH, Hindricks G, Kirchhof P; ESC Committee for Practice Guidelines-CPG; Document Reviewers. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation—developed with the special contribution of the European Heart Rhythm Association. Europace 2012;14:1385–1413.
4.
go back to reference Lickfett L, Mahesh M, Vasamreddy C, Bradley D, Jayam V, Eldadah Z, et al. Radiation exposure during catheter ablation of atrial fibrillation. Circulation. 2004;110:3003–10.CrossRef Lickfett L, Mahesh M, Vasamreddy C, Bradley D, Jayam V, Eldadah Z, et al. Radiation exposure during catheter ablation of atrial fibrillation. Circulation. 2004;110:3003–10.CrossRef
5.
go back to reference Sunaga A, Masuda M, Asai M, Iida O, Okamoto S, Ishihara T, et al. Pressure monitoring predicts pulmonary vein occlusion in cryoballoon ablation. J Interv Card Electrophysiol. 2018;53(1):115–21.CrossRef Sunaga A, Masuda M, Asai M, Iida O, Okamoto S, Ishihara T, et al. Pressure monitoring predicts pulmonary vein occlusion in cryoballoon ablation. J Interv Card Electrophysiol. 2018;53(1):115–21.CrossRef
6.
go back to reference Chao TF, Lin YJ, Chang SL, Lo LW, Hu YF, Tuan TC, et al. Associations between renal function, atrial substrate properties and outcome of catheter ablation in patients with paroxysmal atrial fibrillation. Circ J. 2011;75:2326–32.CrossRef Chao TF, Lin YJ, Chang SL, Lo LW, Hu YF, Tuan TC, et al. Associations between renal function, atrial substrate properties and outcome of catheter ablation in patients with paroxysmal atrial fibrillation. Circ J. 2011;75:2326–32.CrossRef
7.
go back to reference Durán A, Kui Hian S, Miller DL, Le Heron J, Padovani R, Vano E. A summary of recommendations for occupational radiation protection in interventional cardiology. Catheter Cardiovasc Interv. 2013;81:562–7.CrossRef Durán A, Kui Hian S, Miller DL, Le Heron J, Padovani R, Vano E. A summary of recommendations for occupational radiation protection in interventional cardiology. Catheter Cardiovasc Interv. 2013;81:562–7.CrossRef
8.
go back to reference Heidbuchel H, Wittkampf FHM, Vano E, Ernst S, Schilling R, Picano E, et al. Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedure. Europace. 2014;16:946–64.CrossRef Heidbuchel H, Wittkampf FHM, Vano E, Ernst S, Schilling R, Picano E, et al. Practical ways to reduce radiation dose for patients and staff during device implantations and electrophysiological procedure. Europace. 2014;16:946–64.CrossRef
9.
go back to reference Khaykin Y, Oosthuizen R, Zarnett L, Essebag V, Parkash R, Seabrook C, et al. CARTO-guided vs. NavX-guided pulmonary vein antrum isolation and pulmonary vein antrum isolation performed without 3-D mapping: effect of the 3-D mapping system on procedure duration and fluoroscopy time. J Interv Card Electrophysiol. 2011;30(3):233–40.CrossRef Khaykin Y, Oosthuizen R, Zarnett L, Essebag V, Parkash R, Seabrook C, et al. CARTO-guided vs. NavX-guided pulmonary vein antrum isolation and pulmonary vein antrum isolation performed without 3-D mapping: effect of the 3-D mapping system on procedure duration and fluoroscopy time. J Interv Card Electrophysiol. 2011;30(3):233–40.CrossRef
10.
go back to reference Rubesch-Kütemeyer V, Molatta S, Vogt J, Gutleben KJ, Horstkotte D, Nölker G. Reduction of radiation exposure in cryoballoon ablation procedures: a single-Centre study applying intracardiac echocardiography and other radioprotective measures. Europace. 2017;19(6):947–53.PubMed Rubesch-Kütemeyer V, Molatta S, Vogt J, Gutleben KJ, Horstkotte D, Nölker G. Reduction of radiation exposure in cryoballoon ablation procedures: a single-Centre study applying intracardiac echocardiography and other radioprotective measures. Europace. 2017;19(6):947–53.PubMed
11.
go back to reference Nölker G, Heintze J, Gutleben KJ, Muntean B, Pütz V, Yalda A, et al. Cryoballoon pulmonary vein isolation supported by intracardiac echocardiography: integration of a nonfluoroscopic imaging technique in atrial fibrillation ablation. J Cardiovasc Electrophysiol. 2010;21:1325–30.CrossRef Nölker G, Heintze J, Gutleben KJ, Muntean B, Pütz V, Yalda A, et al. Cryoballoon pulmonary vein isolation supported by intracardiac echocardiography: integration of a nonfluoroscopic imaging technique in atrial fibrillation ablation. J Cardiovasc Electrophysiol. 2010;21:1325–30.CrossRef
12.
go back to reference Schmidt M, Daccarett M, Marschang H, Ritscher G, Turschner O, Brachmann J, et al. Intracardiac echocardiography improves procedural efficiency during cryoballoon ablation for atrial fibrillation: a pilot study. J Cardivasc Electrophysiol. 2010;21:1202–7.CrossRef Schmidt M, Daccarett M, Marschang H, Ritscher G, Turschner O, Brachmann J, et al. Intracardiac echocardiography improves procedural efficiency during cryoballoon ablation for atrial fibrillation: a pilot study. J Cardivasc Electrophysiol. 2010;21:1202–7.CrossRef
13.
go back to reference Bohó A, Misiková S, Spirny P, Komanová E, Sudzinová A, Hudák M, et al. Complications of circumferential pulmonary vein isolation using the cryoballoon technique: incidence and predictors. Int J Cardiol. 2014;171:217–23.CrossRef Bohó A, Misiková S, Spirny P, Komanová E, Sudzinová A, Hudák M, et al. Complications of circumferential pulmonary vein isolation using the cryoballoon technique: incidence and predictors. Int J Cardiol. 2014;171:217–23.CrossRef
14.
go back to reference Estner HL. Electrical isolation of pulmonary veins in patients with atrial fibrillation: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX(R)). Europace. 2006;8(8):583–7.CrossRef Estner HL. Electrical isolation of pulmonary veins in patients with atrial fibrillation: reduction of fluoroscopy exposure and procedure duration by the use of a non-fluoroscopic navigation system (NavX(R)). Europace. 2006;8(8):583–7.CrossRef
15.
go back to reference Rotter M. Reduction of fluoroscopy exposure and procedure duration during ablation of atrial fibrillation using a novel anatomical navigation system. Eur Heart J. 2005;26(14):1415–21.CrossRef Rotter M. Reduction of fluoroscopy exposure and procedure duration during ablation of atrial fibrillation using a novel anatomical navigation system. Eur Heart J. 2005;26(14):1415–21.CrossRef
16.
go back to reference Packer DL, Mark DB, Robb RA, Monahan KH, Bahnson TD, Poole JE, Noseworthy PA, Rosenberg YD, Jeffries N, Mitchell LB, Flaker GC, Pokushalov E, Romanov A, Bunch TJ, Noelker G, Ardashev A, Revishvili A, Wilber DJ, Cappato R, Kuck KH, Hindricks G, Davies DW, Kowey PR, Naccarelli GV, Reiffel JA, Piccini JP, Silverstein AP, Al-Khalidi HR, Lee KL; CABANA Investigators. Effect of catheter ablation vs antiarrhythmic drug therapy on mortality, stroke, bleeding, and cardiac arrest among patients with atrial fibrillation: The CABANA Randomized Clinical Trial. JAMA 2019. https://doi.org/10.1001/jama.2019.0693. Packer DL, Mark DB, Robb RA, Monahan KH, Bahnson TD, Poole JE, Noseworthy PA, Rosenberg YD, Jeffries N, Mitchell LB, Flaker GC, Pokushalov E, Romanov A, Bunch TJ, Noelker G, Ardashev A, Revishvili A, Wilber DJ, Cappato R, Kuck KH, Hindricks G, Davies DW, Kowey PR, Naccarelli GV, Reiffel JA, Piccini JP, Silverstein AP, Al-Khalidi HR, Lee KL; CABANA Investigators. Effect of catheter ablation vs antiarrhythmic drug therapy on mortality, stroke, bleeding, and cardiac arrest among patients with atrial fibrillation: The CABANA Randomized Clinical Trial. JAMA 2019. https://​doi.​org/​10.​1001/​jama.​2019.​0693.
17.
go back to reference Sharma A, Dhoot J, Wang J, Jones P, Gupta S, Wimmer AP. Outcomes following cryoballoon ablation for atrial fibrillation guided by pressure waveform monitoring without the routine use of pulmonary venography. J Interv Card Electrophysiol. 2017;49(1):75–82.CrossRef Sharma A, Dhoot J, Wang J, Jones P, Gupta S, Wimmer AP. Outcomes following cryoballoon ablation for atrial fibrillation guided by pressure waveform monitoring without the routine use of pulmonary venography. J Interv Card Electrophysiol. 2017;49(1):75–82.CrossRef
Metadata
Title
Reduction of radiation and contrast agent exposure in a cryoballoon ablation procedure with integration of electromagnetic mapping and intracardiac echocardiography: a single center experience
Authors
Joyce Maalouf
Hoyle L. Whiteside
Ajay Pillai
Abdullah Omar
Adam Berman
Samir Saba
Haitham Hreibe
Publication date
01-12-2020
Publisher
Springer US
Published in
Journal of Interventional Cardiac Electrophysiology / Issue 3/2020
Print ISSN: 1383-875X
Electronic ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-019-00667-z

Keynote series | Spotlight on menopause

Menopause can have a significant impact on the body, with effects ranging beyond the endocrine and reproductive systems. Learn about the systemic effects of menopause, so you can help patients in your clinics through the transition.   

Prof. Martha Hickey
Dr. Claudia Barth
Dr. Samar El Khoudary
Developed by: Springer Medicine
Watch now
Video

A quick guide to ECGs

Improve your ECG interpretation skills with this comprehensive, rapid, interactive course. Expert advice provides detailed feedback as you work through 50 ECGs covering the most common cardiac presentations to ensure your practice stays up to date. 

PD Dr. Carsten W. Israel
Developed by: Springer Medizin
Start the cases