Comparing standard- and low-dose CBCT in diagnosis and treatment decisions for impacted mandibular third molars: a non-inferiority randomised clinical study

Authors: Kuo Feng Hung, Andy Wai Kan Yeung, May Chun Mei Wong, Michael M. Bornstein, Yiu Yan Leung

Published in: Clinical Oral Investigations | Issue 12/2024

Login to get access

Abstract

Objective

This randomised clinical study aimed to assess the influence of low-dose cone-beam computed tomography (CBCT) on the visibility of the mandibular canal (MC) and its proximity to mandibular third molars (M3Ms) as assessed by general dental practitioners (GPs) and oral-maxillofacial surgeons (OMFSs), as well as its impact on their clinical decisions, when compared to standard-dose CBCT.

Methods

154 impacted M3Ms from 90 patients were randomly assigned to three groups for two CBCT exposures using one standard-dose (333 mGy×cm2) and one of the three investigated low-dose (78–131 mGy×cm2) protocols. Blinded assessments of the MC visibility, M3M-MC proximity, surgical approach, crown/root sectioning, and referral decisions, were made by GPs and OMFSs on the images separately. Pairwise comparisons for MC visibility between paired scans were evaluated using Wilcoxon signed rank test, followed by a non-inferiority test with non-inferiority margin of 0.5 on a four-point scale. Differences in other variables between paired scans were evaluated using Wilcoxon signed-rank or McNemar tests.

Results

The majority (78.5–99.3%) of MCs were clearly identified on standard-dose CBCT by all observers. Pairwise comparisons showed significant differences between paired scans only in MC visibility but not in the M3M-MC proximity or treatment decisions. The mean differences in MC visibility between paired scans ranged 0-0.22 with the upper bounds of the 95% confidence intervals (0.09–0.36) falling within the non-inferiority region.

Conclusions

The investigated low-dose CBCT protocols could provide acceptable image quality for the evaluation of impacted M3Ms in most cases. When compared to standard-dose CBCT, these low-dose CBCT images did not significantly affect the assessments of the M3M-MC proximity, treatment strategies, and patient management decisions made by GPs and OMFSs.

Clinical relevance

The low-dose protocols might be clinically acceptable for M3M management while greatly reducing radiation exposure.
Literature
1.
go back to reference Carter K, Worthington S (2016) Predictors of third molar impaction: a systematic review and meta-analysis. J Dent Res 95:267–276PubMedCrossRef Carter K, Worthington S (2016) Predictors of third molar impaction: a systematic review and meta-analysis. J Dent Res 95:267–276PubMedCrossRef
2.
go back to reference Pang SL, Yeung WKA, Hung KF, Hui L, Chung HZJ, Leung YY (2023) Third molar coronectomy vs total removal in second molar periodontal healing. Int Dent J S 0020–6539:00439–00432 Pang SL, Yeung WKA, Hung KF, Hui L, Chung HZJ, Leung YY (2023) Third molar coronectomy vs total removal in second molar periodontal healing. Int Dent J S 0020–6539:00439–00432
8.
go back to reference Matzen LH, Berkhout E (2019) Cone beam CT imaging of the mandibular third molar: a position paper prepared by the European Academy of DentoMaxilloFacial Radiology (EADMFR). Dentomaxillofac Radiol 48:20190039PubMedPubMedCentralCrossRef Matzen LH, Berkhout E (2019) Cone beam CT imaging of the mandibular third molar: a position paper prepared by the European Academy of DentoMaxilloFacial Radiology (EADMFR). Dentomaxillofac Radiol 48:20190039PubMedPubMedCentralCrossRef
9.
go back to reference Yeung AWK, Tanaka R, Jacobs R, Bornstein MM (2020) Awareness and practice of 2D and 3D diagnostic imaging among dentists in Hong Kong. Br Dent J 228:701–709PubMedCrossRef Yeung AWK, Tanaka R, Jacobs R, Bornstein MM (2020) Awareness and practice of 2D and 3D diagnostic imaging among dentists in Hong Kong. Br Dent J 228:701–709PubMedCrossRef
10.
go back to reference Hung KF, Hui L, Yeung AWK, Jacobs R, Leung YY, Bornstein MM (2022) An analysis of patient dose received during cone-beam computed tomography in relation to scan settings and imaging indications as seen in a dental institution in order to establish institutional diagnostic reference levels. Dentomaxillofac Radiol 51:20200529PubMedPubMedCentralCrossRef Hung KF, Hui L, Yeung AWK, Jacobs R, Leung YY, Bornstein MM (2022) An analysis of patient dose received during cone-beam computed tomography in relation to scan settings and imaging indications as seen in a dental institution in order to establish institutional diagnostic reference levels. Dentomaxillofac Radiol 51:20200529PubMedPubMedCentralCrossRef
11.
go back to reference Hol C, Hellén-Halme K, Torgersen G, Nilsson M, Møystad A (2015) How do dentists use CBCT in dental clinics? A Norwegian nationwide survey. Acta Odontol Scand 73:195–201PubMedCrossRef Hol C, Hellén-Halme K, Torgersen G, Nilsson M, Møystad A (2015) How do dentists use CBCT in dental clinics? A Norwegian nationwide survey. Acta Odontol Scand 73:195–201PubMedCrossRef
12.
go back to reference Hung K, Hui L, Yeung AWK, Scarfe WC, Bornstein MM (2020) Image retake rates of cone beam computed tomography in a dental institution. Clin Oral Investig 24:4501–4510PubMedCrossRef Hung K, Hui L, Yeung AWK, Scarfe WC, Bornstein MM (2020) Image retake rates of cone beam computed tomography in a dental institution. Clin Oral Investig 24:4501–4510PubMedCrossRef
13.
go back to reference Szalma J, Vajta L, Lovász BV, Kiss C, Soós B, Lempel E (2020) Identification of specific panoramic high-risk signs in impacted third molar cases in which cone beam computed tomography changes the treatment decision. J Oral Maxillofac Surg 78:1061–1070PubMedCrossRef Szalma J, Vajta L, Lovász BV, Kiss C, Soós B, Lempel E (2020) Identification of specific panoramic high-risk signs in impacted third molar cases in which cone beam computed tomography changes the treatment decision. J Oral Maxillofac Surg 78:1061–1070PubMedCrossRef
14.
go back to reference Wanzeler AMV, Silveira HLDD, Buligon RP, Corsetti A, Vieira HT, Arús NA, Vizzotto MB (2022) Can CBCT change the level of confidence of oral maxillofacial surgeons in mandibular third molar management? Braz Oral Res 36:e078PubMedCrossRef Wanzeler AMV, Silveira HLDD, Buligon RP, Corsetti A, Vieira HT, Arús NA, Vizzotto MB (2022) Can CBCT change the level of confidence of oral maxillofacial surgeons in mandibular third molar management? Braz Oral Res 36:e078PubMedCrossRef
15.
go back to reference Mendonça LM, Gaêta-Araujo H, Cruvinel PB, Tosin IW, Azenha MR, Ferraz EP, Oliveira-Santos C, Tirapelli C (2020) Can diagnostic changes caused by cone beam computed tomography alter the clinical decision in impacted lower third molar treatment plan? Dentomaxillofac Radiol 50:20200412PubMedPubMedCentralCrossRef Mendonça LM, Gaêta-Araujo H, Cruvinel PB, Tosin IW, Azenha MR, Ferraz EP, Oliveira-Santos C, Tirapelli C (2020) Can diagnostic changes caused by cone beam computed tomography alter the clinical decision in impacted lower third molar treatment plan? Dentomaxillofac Radiol 50:20200412PubMedPubMedCentralCrossRef
16.
go back to reference Liljeholm R, Kadesjö N, Benchimol D, Hellén-Halme K, Shi XQ (2017) Cone-beam computed tomography with ultra-low dose protocols for pre-implant radiographic assessment: an in vitro study. Eur J Oral Implantol 10:351–359PubMed Liljeholm R, Kadesjö N, Benchimol D, Hellén-Halme K, Shi XQ (2017) Cone-beam computed tomography with ultra-low dose protocols for pre-implant radiographic assessment: an in vitro study. Eur J Oral Implantol 10:351–359PubMed
17.
go back to reference Zaki IM, Hamed WM, Ashmawy MS (2021) Effect of CBCT dose reduction on the mandibular canal visibility: ex vivo comparative study. Oral Radiol 37:282–289PubMedCrossRef Zaki IM, Hamed WM, Ashmawy MS (2021) Effect of CBCT dose reduction on the mandibular canal visibility: ex vivo comparative study. Oral Radiol 37:282–289PubMedCrossRef
18.
go back to reference Cheung MC, Peters OA, Parashos P (2024) Global cone-beam computed tomography adoption, usage and scan interpretation preferences of dentists and endodontists. Int Endod J 57:133–145PubMedCrossRef Cheung MC, Peters OA, Parashos P (2024) Global cone-beam computed tomography adoption, usage and scan interpretation preferences of dentists and endodontists. Int Endod J 57:133–145PubMedCrossRef
19.
go back to reference Duong C, Zhu Q Jr, Aseltine RH, Kuo CL, da Cunha Godoy L, Kaufman B (2023) A survey on cone-beam computed tomography usage among endodontists in the United States. J Endod 49:1559–1564PubMedCrossRef Duong C, Zhu Q Jr, Aseltine RH, Kuo CL, da Cunha Godoy L, Kaufman B (2023) A survey on cone-beam computed tomography usage among endodontists in the United States. J Endod 49:1559–1564PubMedCrossRef
20.
go back to reference Piaggio G, Elbourne DR, Pocock SJ, Evans SJ, Altman DG, CONSORT Group (2012) Reporting of noninferiority and equivalence randomized trials: extension of the CONSORT 2010 statement. JAMA 308:2594–2604PubMedCrossRef Piaggio G, Elbourne DR, Pocock SJ, Evans SJ, Altman DG, CONSORT Group (2012) Reporting of noninferiority and equivalence randomized trials: extension of the CONSORT 2010 statement. JAMA 308:2594–2604PubMedCrossRef
21.
go back to reference Lee HJ, Kim JS, Lee JK, Lee HA, Pak S (2023) Ultra-low-dose hepatic multiphase CT using deep learning-based image reconstruction algorithm focused on arterial phase in chronic liver disease: a non-inferiority study. Eur J Radiol 159:110659PubMedCrossRef Lee HJ, Kim JS, Lee JK, Lee HA, Pak S (2023) Ultra-low-dose hepatic multiphase CT using deep learning-based image reconstruction algorithm focused on arterial phase in chronic liver disease: a non-inferiority study. Eur J Radiol 159:110659PubMedCrossRef
22.
go back to reference Park S, Yoon JH, Joo I, Yu MH, Kim JH, Park J, Kim SW, Han S, Ahn C, Kim JH, Lee JM (2022) Image quality in liver CT: low-dose deep learning vs standard-dose model-based iterative reconstructions. Eur Radiol 32:2865–2874PubMedCrossRef Park S, Yoon JH, Joo I, Yu MH, Kim JH, Park J, Kim SW, Han S, Ahn C, Kim JH, Lee JM (2022) Image quality in liver CT: low-dose deep learning vs standard-dose model-based iterative reconstructions. Eur Radiol 32:2865–2874PubMedCrossRef
23.
go back to reference Fryback DG, Thornbury JR (1991) The efficacy of diagnostic imaging. Med Decis Mak 11:88–94CrossRef Fryback DG, Thornbury JR (1991) The efficacy of diagnostic imaging. Med Decis Mak 11:88–94CrossRef
24.
go back to reference Meyer E, Labani A, Schaeffer M, Jeung MY, Ludes C, Meyer A, Roy C, Leyendecker P, Ohana M (2019) Wide-volume versus helical acquisition in unenhanced chest CT: prospective intra-patient comparison of diagnostic accuracy and radiation dose in an ultra-low-dose setting. Eur Radiol 29:6858–6866PubMedCrossRef Meyer E, Labani A, Schaeffer M, Jeung MY, Ludes C, Meyer A, Roy C, Leyendecker P, Ohana M (2019) Wide-volume versus helical acquisition in unenhanced chest CT: prospective intra-patient comparison of diagnostic accuracy and radiation dose in an ultra-low-dose setting. Eur Radiol 29:6858–6866PubMedCrossRef
25.
go back to reference Cederhag J, Iskanderani D, Alstergren P, Shi XQ, Hellén-Halme K (2023) Visibility of anatomical landmarks in the region of the mandibular third molar, a comparison between a low-dose and default protocol of CBCT. Acta Odontol Scand 81:449–455PubMed Cederhag J, Iskanderani D, Alstergren P, Shi XQ, Hellén-Halme K (2023) Visibility of anatomical landmarks in the region of the mandibular third molar, a comparison between a low-dose and default protocol of CBCT. Acta Odontol Scand 81:449–455PubMed
26.
go back to reference Pan S, Su JJ, Syed J, Moore C, Israel G, Singh D (2019) Reduced dose computed tomography: the effects of voltage reduction on density measurements of urolithiasis. J Endourol 33:682–686PubMedCrossRef Pan S, Su JJ, Syed J, Moore C, Israel G, Singh D (2019) Reduced dose computed tomography: the effects of voltage reduction on density measurements of urolithiasis. J Endourol 33:682–686PubMedCrossRef
27.
go back to reference Ludes C, Labani A, Severac F, Jeung MY, Leyendecker P, Roy C, Ohana M (2019) Ultra-low-dose unenhanced chest CT: prospective comparison of high kV/low mA versus low kV/high mA protocols. Diagn Interv Imaging 100:85–93PubMedCrossRef Ludes C, Labani A, Severac F, Jeung MY, Leyendecker P, Roy C, Ohana M (2019) Ultra-low-dose unenhanced chest CT: prospective comparison of high kV/low mA versus low kV/high mA protocols. Diagn Interv Imaging 100:85–93PubMedCrossRef
28.
go back to reference Kurihara C, Applegate KE, Jeong JH, Akahane K, Kang KW (2023) Radiological protection in human research ethics using a case study: toward update of the ICRP publication 62. J Radiol Prot 43:044002CrossRef Kurihara C, Applegate KE, Jeong JH, Akahane K, Kang KW (2023) Radiological protection in human research ethics using a case study: toward update of the ICRP publication 62. J Radiol Prot 43:044002CrossRef
30.
go back to reference Tadinada A, Schneider S, Yadav S (2017) Evaluation of the diagnostic efficacy of two cone beam computed tomography protocols in reliably detecting the location of the inferior alveolar nerve canal. Dentomaxillofac Radiol 46:20160389PubMedPubMedCentralCrossRef Tadinada A, Schneider S, Yadav S (2017) Evaluation of the diagnostic efficacy of two cone beam computed tomography protocols in reliably detecting the location of the inferior alveolar nerve canal. Dentomaxillofac Radiol 46:20160389PubMedPubMedCentralCrossRef
31.
go back to reference Waltrick KB, Nunes de Abreu Junior MJ, Corrêa M, Zastrow MD, Dutra VD (2013) Accuracy of linear measurements and visibility of the mandibular canal of cone-beam computed tomography images with different voxel sizes: an in vitro study. J Periodontol 84:68–77PubMedCrossRef Waltrick KB, Nunes de Abreu Junior MJ, Corrêa M, Zastrow MD, Dutra VD (2013) Accuracy of linear measurements and visibility of the mandibular canal of cone-beam computed tomography images with different voxel sizes: an in vitro study. J Periodontol 84:68–77PubMedCrossRef
33.
go back to reference Ahn S, Park SH, Lee KH (2013) How to demonstrate similarity by using noninferiority and equivalence statistical testing in radiology research. Radiology 267:328–338PubMedCrossRef Ahn S, Park SH, Lee KH (2013) How to demonstrate similarity by using noninferiority and equivalence statistical testing in radiology research. Radiology 267:328–338PubMedCrossRef
34.
go back to reference Savoldi F, Montalvao C, Hui L, Leung CKK, Jablonski NG, Tsoi JKH, Bornstein MM (2021) The Human Bone Collection of the Faculty of Dentistry at the University of Hong Kong: history and description of cranial and postcranial skeletal remains. Am J Phys Anthropol 175:718–730PubMedCrossRef Savoldi F, Montalvao C, Hui L, Leung CKK, Jablonski NG, Tsoi JKH, Bornstein MM (2021) The Human Bone Collection of the Faculty of Dentistry at the University of Hong Kong: history and description of cranial and postcranial skeletal remains. Am J Phys Anthropol 175:718–730PubMedCrossRef
35.
go back to reference Jasa GR, Shimizu M, Okamura K, Tokumori K, Takeshita Y, Weerawanich W, Yoshiura K (2017) Effects of exposure parameters and slice thickness on detecting clear and unclear mandibular canals using cone beam CT. Dentomaxillofac Radiol 46:20160315PubMedPubMedCentralCrossRef Jasa GR, Shimizu M, Okamura K, Tokumori K, Takeshita Y, Weerawanich W, Yoshiura K (2017) Effects of exposure parameters and slice thickness on detecting clear and unclear mandibular canals using cone beam CT. Dentomaxillofac Radiol 46:20160315PubMedPubMedCentralCrossRef
36.
go back to reference Brasil DM, Pauwels R, Coucke W, Haiter-Neto F, Jacobs R (2019) Image quality optimization of narrow detector dental computed tomography for paediatric patients. Dentomaxillofac Radiol 48:20190032PubMedPubMedCentralCrossRef Brasil DM, Pauwels R, Coucke W, Haiter-Neto F, Jacobs R (2019) Image quality optimization of narrow detector dental computed tomography for paediatric patients. Dentomaxillofac Radiol 48:20190032PubMedPubMedCentralCrossRef
37.
go back to reference Greese J, Diekhoff T, Sieper J, Schwenke C, Makowski MR, Poddubnyy D, Hamm B, Hermann KGA (2019) Detection of sacroiliitis by short-tau inversion recovery and T2-weighted turbo spin echo sequences: results from the SIMACT study. J Rheumatol 46:376–383PubMedCrossRef Greese J, Diekhoff T, Sieper J, Schwenke C, Makowski MR, Poddubnyy D, Hamm B, Hermann KGA (2019) Detection of sacroiliitis by short-tau inversion recovery and T2-weighted turbo spin echo sequences: results from the SIMACT study. J Rheumatol 46:376–383PubMedCrossRef
38.
go back to reference Foster N, Shaffrey C, Buchholz A, Turner R, Yang LZ, Niedzwiecki D, Goode A (2022) Image quality and dose comparison of 3 mobile intraoperative three-dimensional imaging systems in spine surgery. World Neurosurg 160:e142–e151PubMedPubMedCentralCrossRef Foster N, Shaffrey C, Buchholz A, Turner R, Yang LZ, Niedzwiecki D, Goode A (2022) Image quality and dose comparison of 3 mobile intraoperative three-dimensional imaging systems in spine surgery. World Neurosurg 160:e142–e151PubMedPubMedCentralCrossRef
39.
go back to reference Lagos de Melo LP, Queiroz PM, Moreira-Souza L, Nadaes MR, Santaella GM, Oliveira ML, Freitas DQ (2023) Influence of CBCT parameters on image quality and the diagnosis of vertical root fractures in teeth with metallic posts: an ex vivo study. Restor Dent Endod 48:e16PubMedPubMedCentralCrossRef Lagos de Melo LP, Queiroz PM, Moreira-Souza L, Nadaes MR, Santaella GM, Oliveira ML, Freitas DQ (2023) Influence of CBCT parameters on image quality and the diagnosis of vertical root fractures in teeth with metallic posts: an ex vivo study. Restor Dent Endod 48:e16PubMedPubMedCentralCrossRef
40.
go back to reference Lamira A, Mazzi-Chaves JF, Nicolielo LFP, Leoni GB, Silva-Sousa AC, Silva-Sousa YTC, Pauwels R, Buls N, Jacobs R, Sousa-Neto MD (2022) CBCT-based assessment of root canal treatment using micro-CT reference images. Imaging Sci Dent 52:245–258PubMedPubMedCentralCrossRef Lamira A, Mazzi-Chaves JF, Nicolielo LFP, Leoni GB, Silva-Sousa AC, Silva-Sousa YTC, Pauwels R, Buls N, Jacobs R, Sousa-Neto MD (2022) CBCT-based assessment of root canal treatment using micro-CT reference images. Imaging Sci Dent 52:245–258PubMedPubMedCentralCrossRef
41.
go back to reference Choi SY, Min JH, Kim JH, Park HJ, Kim YY, Han YE, Bae SH, Lee JH, Choi YH, Moon JE (2023) Interobserver variability and diagnostic performance in predicting malignancy of pancreatic intraductal papillary mucinous neoplasm with MRI. Radiology 308:e222463PubMedCrossRef Choi SY, Min JH, Kim JH, Park HJ, Kim YY, Han YE, Bae SH, Lee JH, Choi YH, Moon JE (2023) Interobserver variability and diagnostic performance in predicting malignancy of pancreatic intraductal papillary mucinous neoplasm with MRI. Radiology 308:e222463PubMedCrossRef
42.
go back to reference Gay-Escoda C, Sánchez-Torres A, Borrás-Ferreres J, Valmaseda-Castellón E (2022) Third molar surgical difficulty scales: systematic review and preoperative assessment form. Med Oral Patol Oral Cir Bucal 27:e68–e76PubMedCrossRef Gay-Escoda C, Sánchez-Torres A, Borrás-Ferreres J, Valmaseda-Castellón E (2022) Third molar surgical difficulty scales: systematic review and preoperative assessment form. Med Oral Patol Oral Cir Bucal 27:e68–e76PubMedCrossRef
43.
go back to reference Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef Landis JR, Koch GG (1977) The measurement of observer agreement for categorical data. Biometrics 33:159–174PubMedCrossRef
44.
go back to reference Neves FS, Souza TC, de-Azevedo-Vaz SL, Campos PSF, Bóscolo FN (2014) Influence of cone-beam computed tomography milliamperage settings on image quality of the mandibular third molar region. Oral Radiol 30:27–31CrossRef Neves FS, Souza TC, de-Azevedo-Vaz SL, Campos PSF, Bóscolo FN (2014) Influence of cone-beam computed tomography milliamperage settings on image quality of the mandibular third molar region. Oral Radiol 30:27–31CrossRef
46.
go back to reference Pakravan AH, Aghamiri SMR, Bamdadian T, Gholami M, Moshfeghi M (2019) Dosimetry of occupational radiation around panoramic X-ray apparatus. J Biomed Phys Eng 9:525–532 Pakravan AH, Aghamiri SMR, Bamdadian T, Gholami M, Moshfeghi M (2019) Dosimetry of occupational radiation around panoramic X-ray apparatus. J Biomed Phys Eng 9:525–532
51.
go back to reference Leung YY, Cheung LK (2010) Coronectomy as the treatment of choice in wisdom teeth showing radiographic signs of close proximity to inferior dental nerve. Ann R Australas Coll Dent Surg 20:93–94PubMed Leung YY, Cheung LK (2010) Coronectomy as the treatment of choice in wisdom teeth showing radiographic signs of close proximity to inferior dental nerve. Ann R Australas Coll Dent Surg 20:93–94PubMed
Metadata
Title
Comparing standard- and low-dose CBCT in diagnosis and treatment decisions for impacted mandibular third molars: a non-inferiority randomised clinical study
Authors
Kuo Feng Hung
Andy Wai Kan Yeung
May Chun Mei Wong
Michael M. Bornstein
Yiu Yan Leung
Publication date
01-12-2024
Publisher
Springer Berlin Heidelberg
Published in
Clinical Oral Investigations / Issue 12/2024
Print ISSN: 1432-6981
Electronic ISSN: 1436-3771
DOI
https://doi.org/10.1007/s00784-024-06022-5