16-05-2024 | CT Angiography | Head-Neck-ENT Radiology

A Comparative Study of AI-Based Automated and Manual Postprocessing of Head and Neck CT Angiography: An Independent External Validation with Multi-Vendor and Multi-Center Data

Authors: Kunhua Li, Yang Yang, Shengwen Niu, Yongwei Yang, Bitong Tian, Xinyue Huan, Dajing Guo

Published in: Neuroradiology

Login to get access

Abstract

Purpose

To externally validate the performance of automated postprocessing (AP) on head and neck CT Angiography (CTA) and compare it with manual postprocessing (MP).

Methods

This retrospective study included head and neck CTA-exams of patients from three tertiary hospitals acquired on CT scanners from five manufacturers. AP was performed by CerebralDoc. The image quality was assessed using Likert scales, and the qualitative and quantitative diagnostic performance of arterial stenosis and aneurysm, postprocessing time, and scanning radiation dose were also evaluated.

Results

A total of 250 patients were included. Among these, 55 patients exhibited significant stenosis (≥ 50%), and 33 patients had aneurysms, diagnosed using original CTA datasets and corresponding multiplanar reconstructions as the reference. While the scores of the V4 segment and the edge of the M1 segment on volume rendering (VR), as well as the C4 segment on maximum intensity projection (MIP), were significantly lower with AP compared to MP across vendors (all P < 0.05), most scores in AP demonstrated image quality that was either superior to or comparable with that of MP. Furthermore, the diagnostic performance of AP was either superior to or comparable with that of MP. Moreover, AP also exhibited advantages in terms of postprocessing time and radiation dose when compared to MP (P < 0.001).

Conclusion

The AP of CerebralDoc presents clear advantages over MP and holds significant clinical value. However, further optimization is required in the image quality of the V4 and M1 segments on VR as well as the C4 segment on MIP.
Appendix
Available only for authorised users
Literature
1.
go back to reference Powers WJ, Rabinstein AA, Ackerson T et al (2019) Guidelines for the Early Management of Patients With Acute Ischemic Stroke: 2019 Update to the 2018 Guidelines for the Early Management of Acute Ischemic Stroke: A Guideline for Healthcare Professionals From the American Heart Association/American Stroke Association. Stroke 50:e344–e418PubMedCrossRef Powers WJ, Rabinstein AA, Ackerson T et al (2019) Guidelines for the Early Management of Patients With Acute Ischemic Stroke: 2019 Update to the 2018 Guidelines for the Early Management of Acute Ischemic Stroke: A Guideline for Healthcare Professionals From the American Heart Association/American Stroke Association. Stroke 50:e344–e418PubMedCrossRef
2.
go back to reference Thompson BG, Brown RD, Amin-Hanjani S et al (2015) Guidelines for the Management of Patients With Unruptured Intracranial Aneurysms. Stroke 46:2368–2400PubMedCrossRef Thompson BG, Brown RD, Amin-Hanjani S et al (2015) Guidelines for the Management of Patients With Unruptured Intracranial Aneurysms. Stroke 46:2368–2400PubMedCrossRef
3.
go back to reference Wu S, Wu B, Liu M et al (2019) Stroke in China: advances and challenges in epidemiology, prevention, and management. The Lancet Neurology 18:394–405PubMedCrossRef Wu S, Wu B, Liu M et al (2019) Stroke in China: advances and challenges in epidemiology, prevention, and management. The Lancet Neurology 18:394–405PubMedCrossRef
4.
go back to reference Zhang L-J, Wu S-Y, Poon CS et al (2010) Automatic Bone Removal Dual-Energy CT Angiography for the Evaluation of Intracranial Aneurysms. J Comput Assist Tomogr 34:816–824PubMedCrossRef Zhang L-J, Wu S-Y, Poon CS et al (2010) Automatic Bone Removal Dual-Energy CT Angiography for the Evaluation of Intracranial Aneurysms. J Comput Assist Tomogr 34:816–824PubMedCrossRef
5.
go back to reference Saba L, Sanfilippo R, Montisci R, Mallarini G (2010) Assessment of Intracranial Arterial Stenosis with Multidetector Row CT Angiography: A Postprocessing Techniques Comparison. Am J Neuroradiol 31:874–879PubMedPubMedCentralCrossRef Saba L, Sanfilippo R, Montisci R, Mallarini G (2010) Assessment of Intracranial Arterial Stenosis with Multidetector Row CT Angiography: A Postprocessing Techniques Comparison. Am J Neuroradiol 31:874–879PubMedPubMedCentralCrossRef
6.
go back to reference Kaemmerer N, Brand M, Hammon M et al (2016) Dual-Energy Computed Tomography Angiography of the Head and Neck With Single-Source Computed Tomography. Invest Radiol 51:618–623PubMedCrossRef Kaemmerer N, Brand M, Hammon M et al (2016) Dual-Energy Computed Tomography Angiography of the Head and Neck With Single-Source Computed Tomography. Invest Radiol 51:618–623PubMedCrossRef
7.
go back to reference Lell MM, Ditt H, Panknin C et al (2008) Cervical CT Angiography Comparing Routine Noncontrast and a Late Venous Scan as Masks for Automated Bone Subtraction. Invest Radiol 43:27–32PubMedCrossRef Lell MM, Ditt H, Panknin C et al (2008) Cervical CT Angiography Comparing Routine Noncontrast and a Late Venous Scan as Masks for Automated Bone Subtraction. Invest Radiol 43:27–32PubMedCrossRef
8.
go back to reference Li Q, Lv F, Wei Y, Luo T, Xie P (2013) Automated Subtraction CT Angiography for Visualization of the Whole Brain Vasculature. Acad Radiol 20:1009–1014PubMedCrossRef Li Q, Lv F, Wei Y, Luo T, Xie P (2013) Automated Subtraction CT Angiography for Visualization of the Whole Brain Vasculature. Acad Radiol 20:1009–1014PubMedCrossRef
9.
go back to reference Fu F, Wei J, Zhang M et al (2020) Rapid vessel segmentation and reconstruction of head and neck angiograms using 3D convolutional neural network. Nat Commun 11:4829PubMedPubMedCentralCrossRef Fu F, Wei J, Zhang M et al (2020) Rapid vessel segmentation and reconstruction of head and neck angiograms using 3D convolutional neural network. Nat Commun 11:4829PubMedPubMedCentralCrossRef
10.
go back to reference Lell MM, Ruehm SG, Kramer M et al (2009) Cranial Computed Tomography Angiography With Automated Bone Subtraction. Invest Radiol 44:38–43PubMedCrossRef Lell MM, Ruehm SG, Kramer M et al (2009) Cranial Computed Tomography Angiography With Automated Bone Subtraction. Invest Radiol 44:38–43PubMedCrossRef
11.
go back to reference Fu F, Shan Y, Yang G et al (2023) Deep Learning for Head and Neck CT Angiography: Stenosis and Plaque Classification. Radiology 307:220996CrossRef Fu F, Shan Y, Yang G et al (2023) Deep Learning for Head and Neck CT Angiography: Stenosis and Plaque Classification. Radiology 307:220996CrossRef
12.
go back to reference Sheth SA, Giancardo L, Colasurdo M, Srinivasan VM, Niktabe A, Kan P (2023) Machine learning and acute stroke imaging. Journal of NeuroInterventional Surgery 15:195–199PubMedCrossRef Sheth SA, Giancardo L, Colasurdo M, Srinivasan VM, Niktabe A, Kan P (2023) Machine learning and acute stroke imaging. Journal of NeuroInterventional Surgery 15:195–199PubMedCrossRef
13.
go back to reference Li M, Ling R, Yu L et al (2023) Deep Learning Segmentation and Reconstruction for CT of Chronic Total Coronary Occlusion. Radiology 306:e221393PubMedCrossRef Li M, Ling R, Yu L et al (2023) Deep Learning Segmentation and Reconstruction for CT of Chronic Total Coronary Occlusion. Radiology 306:e221393PubMedCrossRef
14.
go back to reference Xing H, Zhang X, Nie Y et al (2022) A deep learning-based post-processing method for automated pulmonary lobe and airway trees segmentation using chest CT images in PET/CT. Quant Imaging Med Surg 12:4747–4757PubMedPubMedCentralCrossRef Xing H, Zhang X, Nie Y et al (2022) A deep learning-based post-processing method for automated pulmonary lobe and airway trees segmentation using chest CT images in PET/CT. Quant Imaging Med Surg 12:4747–4757PubMedPubMedCentralCrossRef
15.
go back to reference van der Ven JPG, van Genuchten W, Sadighy Z et al (2022) Multivendor Evaluation of Automated MRI Postprocessing of Biventricular Size and Function for Children With and Without Congenital Heart Defects. J Magn Reson Imaging 58:794–804PubMed van der Ven JPG, van Genuchten W, Sadighy Z et al (2022) Multivendor Evaluation of Automated MRI Postprocessing of Biventricular Size and Function for Children With and Without Congenital Heart Defects. J Magn Reson Imaging 58:794–804PubMed
16.
go back to reference Pain CD, Egan GF, Chen Z (2022) Deep learning-based image reconstruction and post-processing methods in positron emission tomography for low-dose imaging and resolution enhancement. Eur J Nucl Med Mol Imaging 49:3098–3118PubMedPubMedCentralCrossRef Pain CD, Egan GF, Chen Z (2022) Deep learning-based image reconstruction and post-processing methods in positron emission tomography for low-dose imaging and resolution enhancement. Eur J Nucl Med Mol Imaging 49:3098–3118PubMedPubMedCentralCrossRef
17.
go back to reference Luo N, He Y, Fan J et al (2022) Performance evaluation of deep learning-based post-processing and diagnostic reporting system for coronary CT angiography: a clinical comparative study. Chin Med J 135:2366–2368PubMedPubMedCentralCrossRef Luo N, He Y, Fan J et al (2022) Performance evaluation of deep learning-based post-processing and diagnostic reporting system for coronary CT angiography: a clinical comparative study. Chin Med J 135:2366–2368PubMedPubMedCentralCrossRef
18.
go back to reference Corrado PA, Wentland AL, Starekova J, Dhyani A, Goss KN, Wieben O (2022) Fully automated intracardiac 4D flow MRI post-processing using deep learning for biventricular segmentation. Eur Radiol 32:5669–5678PubMedCrossRef Corrado PA, Wentland AL, Starekova J, Dhyani A, Goss KN, Wieben O (2022) Fully automated intracardiac 4D flow MRI post-processing using deep learning for biventricular segmentation. Eur Radiol 32:5669–5678PubMedCrossRef
19.
go back to reference Ferreira PF, Martin RR, Scott AD et al (2020) Automating in vivo cardiac diffusion tensor postprocessing with deep learning–based segmentation. Magn Reson Med 84:2801–2814PubMedCrossRef Ferreira PF, Martin RR, Scott AD et al (2020) Automating in vivo cardiac diffusion tensor postprocessing with deep learning–based segmentation. Magn Reson Med 84:2801–2814PubMedCrossRef
20.
go back to reference Ekert K, Kloth C, Fritz J, Ioanoviciu SD, Horger M (2020) Improved Detection of Benign and Malignant Rib Lesions in the Routine Computed Tomography Workup of Oncological Patients Using Automated Unfolded Rib Image Postprocessing. Invest Radiol 55:84–90PubMedCrossRef Ekert K, Kloth C, Fritz J, Ioanoviciu SD, Horger M (2020) Improved Detection of Benign and Malignant Rib Lesions in the Routine Computed Tomography Workup of Oncological Patients Using Automated Unfolded Rib Image Postprocessing. Invest Radiol 55:84–90PubMedCrossRef
21.
go back to reference Shi Z, Zhao X, Zhu S et al (2023) Time-of-Flight Intracranial MRA at 3 T versus 5 T versus 7 T: Visualization of Distal Small Cerebral Arteries. Radiology 306:207–217PubMedCrossRef Shi Z, Zhao X, Zhu S et al (2023) Time-of-Flight Intracranial MRA at 3 T versus 5 T versus 7 T: Visualization of Distal Small Cerebral Arteries. Radiology 306:207–217PubMedCrossRef
22.
go back to reference Greve T, Sollmann N, Hock A, Zimmer C, Kirschke JS (2020) Novel Ultrafast Spiral Head MR Angiography Compared to Standard MR and CT Angiography. J Neuroimaging 31:45–56PubMedCrossRef Greve T, Sollmann N, Hock A, Zimmer C, Kirschke JS (2020) Novel Ultrafast Spiral Head MR Angiography Compared to Standard MR and CT Angiography. J Neuroimaging 31:45–56PubMedCrossRef
23.
go back to reference Greve T, Sollmann N, Hock A et al (2019) Highly accelerated time-of-flight magnetic resonance angiography using spiral imaging improves conspicuity of intracranial arterial branches while reducing scan time. Eur Radiol 30:855–865PubMedCrossRef Greve T, Sollmann N, Hock A et al (2019) Highly accelerated time-of-flight magnetic resonance angiography using spiral imaging improves conspicuity of intracranial arterial branches while reducing scan time. Eur Radiol 30:855–865PubMedCrossRef
24.
go back to reference Lell MM, Kramer M, Klotz E, Villablanca P, Ruehm SG (2009) Carotid Computed Tomography Angiography With Automated Bone Suppression. Invest Radiol 44:322–328PubMedCrossRef Lell MM, Kramer M, Klotz E, Villablanca P, Ruehm SG (2009) Carotid Computed Tomography Angiography With Automated Bone Suppression. Invest Radiol 44:322–328PubMedCrossRef
25.
go back to reference Slieker RC, van der Heijden AAWA, Siddiqui MK et al (2021) Performance of prediction models for nephropathy in people with type 2 diabetes: systematic review and external validation study. BMJ 374:n2134PubMedPubMedCentralCrossRef Slieker RC, van der Heijden AAWA, Siddiqui MK et al (2021) Performance of prediction models for nephropathy in people with type 2 diabetes: systematic review and external validation study. BMJ 374:n2134PubMedPubMedCentralCrossRef
26.
27.
go back to reference de Jong VMT, Rousset RZ, Antonio-Villa NE et al (2022) Clinical prediction models for mortality in patients with covid-19: external validation and individual participant data meta-analysis. BMJ 378:e069881PubMedCrossRef de Jong VMT, Rousset RZ, Antonio-Villa NE et al (2022) Clinical prediction models for mortality in patients with covid-19: external validation and individual participant data meta-analysis. BMJ 378:e069881PubMedCrossRef
28.
go back to reference Yang Y, Huan X, Guo D, Wang X, Niu S, Li K (2023) Performance of deep learning-based autodetection of arterial stenosis on head and neck CT angiography: an independent external validation study. Radiol Med (Torino) 128:1103–1115PubMedCrossRef Yang Y, Huan X, Guo D, Wang X, Niu S, Li K (2023) Performance of deep learning-based autodetection of arterial stenosis on head and neck CT angiography: an independent external validation study. Radiol Med (Torino) 128:1103–1115PubMedCrossRef
30.
go back to reference Yang W, Chen C, Yang Y et al (2023) Diagnostic performance of deep learning-based vessel extraction and stenosis detection on coronary computed tomography angiography for coronary artery disease: a multi-reader multi-case study. Radiol Med 128:307–315PubMedCrossRef Yang W, Chen C, Yang Y et al (2023) Diagnostic performance of deep learning-based vessel extraction and stenosis detection on coronary computed tomography angiography for coronary artery disease: a multi-reader multi-case study. Radiol Med 128:307–315PubMedCrossRef
31.
go back to reference Saver JL, Chapot R, Agid R et al (2020) Thrombectomy for Distal, Medium Vessel Occlusions. Stroke 51:2872–2884PubMedCrossRef Saver JL, Chapot R, Agid R et al (2020) Thrombectomy for Distal, Medium Vessel Occlusions. Stroke 51:2872–2884PubMedCrossRef
32.
go back to reference Boos J, Fang J, Heidinger BH, Raptopoulos V, Brook OR (2016) Dual energy CT angiography: pros and cons of dual-energy metal artifact reduction algorithm in patients after endovascular aortic repair. Abdom Radiol 42:749–758CrossRef Boos J, Fang J, Heidinger BH, Raptopoulos V, Brook OR (2016) Dual energy CT angiography: pros and cons of dual-energy metal artifact reduction algorithm in patients after endovascular aortic repair. Abdom Radiol 42:749–758CrossRef
33.
go back to reference Neyazi B, Swiatek VM, Skalej M et al (2020) Rupture risk assessment for multiple intracranial aneurysms: why there is no need for dozens of clinical, morphological and hemodynamic parameters. Ther Adv Neurol Disord 13:1756286420966159PubMedPubMedCentralCrossRef Neyazi B, Swiatek VM, Skalej M et al (2020) Rupture risk assessment for multiple intracranial aneurysms: why there is no need for dozens of clinical, morphological and hemodynamic parameters. Ther Adv Neurol Disord 13:1756286420966159PubMedPubMedCentralCrossRef
34.
go back to reference Malhotra A, Wu X, Forman HP, Matouk CC, Gandhi D, Sanelli P (2018) Management of Tiny Unruptured Intracranial Aneurysms. JAMA Neurol 75:27–34PubMedCrossRef Malhotra A, Wu X, Forman HP, Matouk CC, Gandhi D, Sanelli P (2018) Management of Tiny Unruptured Intracranial Aneurysms. JAMA Neurol 75:27–34PubMedCrossRef
35.
go back to reference Taki W (2011) Factors predicting retreatment and residual aneurysms at 1 year after endovascular coiling for ruptured cerebral aneurysms: Prospective Registry of Subarachnoid Aneurysms Treatment (PRESAT) in Japan. Neuroradiology 54:597–606PubMedCrossRef Taki W (2011) Factors predicting retreatment and residual aneurysms at 1 year after endovascular coiling for ruptured cerebral aneurysms: Prospective Registry of Subarachnoid Aneurysms Treatment (PRESAT) in Japan. Neuroradiology 54:597–606PubMedCrossRef
36.
go back to reference Yan C, Yu F, Zhang Y et al (2023) Multidelay Arterial Spin Labeling Versus Computed Tomography Perfusion in Penumbra Volume of Acute Ischemic Stroke. Stroke 54:1037–1045PubMedCrossRef Yan C, Yu F, Zhang Y et al (2023) Multidelay Arterial Spin Labeling Versus Computed Tomography Perfusion in Penumbra Volume of Acute Ischemic Stroke. Stroke 54:1037–1045PubMedCrossRef
37.
go back to reference Amukotuwa SA, Wu A, Zhou K, Page I, Brotchie P, Bammer R (2021) Distal Medium Vessel Occlusions Can Be Accurately and Rapidly Detected Using Tmax Maps. Stroke 52:3308–3317PubMedCrossRef Amukotuwa SA, Wu A, Zhou K, Page I, Brotchie P, Bammer R (2021) Distal Medium Vessel Occlusions Can Be Accurately and Rapidly Detected Using Tmax Maps. Stroke 52:3308–3317PubMedCrossRef
38.
go back to reference Brown RD, Broderick JP (2014) Unruptured intracranial aneurysms: epidemiology, natural history, management options, and familial screening. The Lancet Neurology 13:393–404PubMedCrossRef Brown RD, Broderick JP (2014) Unruptured intracranial aneurysms: epidemiology, natural history, management options, and familial screening. The Lancet Neurology 13:393–404PubMedCrossRef
39.
go back to reference Nam J-S, Jeon S-B, Jo J-Y et al (2019) Perioperative rupture risk of unruptured intracranial aneurysms in cardiovascular surgery. Brain 142:1408–1415PubMedCrossRef Nam J-S, Jeon S-B, Jo J-Y et al (2019) Perioperative rupture risk of unruptured intracranial aneurysms in cardiovascular surgery. Brain 142:1408–1415PubMedCrossRef
40.
go back to reference Yan Z, Yang X, Cheng K-T (2019) A Three-Stage Deep Learning Model for Accurate Retinal Vessel Segmentation. IEEE J Biomed Health Inform 23:1427–1436PubMedCrossRef Yan Z, Yang X, Cheng K-T (2019) A Three-Stage Deep Learning Model for Accurate Retinal Vessel Segmentation. IEEE J Biomed Health Inform 23:1427–1436PubMedCrossRef
41.
go back to reference Lell MM, Hinkmann F, Nkenke E et al (2010) Dual energy CTA of the supraaortic arteries: Technical improvements with a novel dual source CT system. Eur J Radiol 76:e6–e12PubMedCrossRef Lell MM, Hinkmann F, Nkenke E et al (2010) Dual energy CTA of the supraaortic arteries: Technical improvements with a novel dual source CT system. Eur J Radiol 76:e6–e12PubMedCrossRef
42.
go back to reference Deng K, Liu C, Ma R et al (2009) Clinical evaluation of dual-energy bone removal in CT angiography of the head and neck: comparison with conventional bone-subtraction CT angiography. Clin Radiol 64:534–541PubMedCrossRef Deng K, Liu C, Ma R et al (2009) Clinical evaluation of dual-energy bone removal in CT angiography of the head and neck: comparison with conventional bone-subtraction CT angiography. Clin Radiol 64:534–541PubMedCrossRef
43.
go back to reference Fabijańska A (2018) Segmentation of corneal endothelium images using a U-Net-based convolutional neural network. Artif Intell Med 88:1–13PubMedCrossRef Fabijańska A (2018) Segmentation of corneal endothelium images using a U-Net-based convolutional neural network. Artif Intell Med 88:1–13PubMedCrossRef
44.
go back to reference Nguyen-Huynh MN, Wintermark M, English J et al (2008) How Accurate Is CT Angiography in Evaluating Intracranial Atherosclerotic Disease? Stroke 39:1184–1188PubMedCrossRef Nguyen-Huynh MN, Wintermark M, English J et al (2008) How Accurate Is CT Angiography in Evaluating Intracranial Atherosclerotic Disease? Stroke 39:1184–1188PubMedCrossRef
45.
go back to reference Huang J, Degnan AJ, Liu Q et al (2012) Comparison of NASCET and WASID criteria for the measurement of intracranial stenosis using digital subtraction and computed tomography angiography of the middle cerebral artery. J Neuroradiol 39:342–345PubMedCrossRef Huang J, Degnan AJ, Liu Q et al (2012) Comparison of NASCET and WASID criteria for the measurement of intracranial stenosis using digital subtraction and computed tomography angiography of the middle cerebral artery. J Neuroradiol 39:342–345PubMedCrossRef
46.
go back to reference Westerlaan HE, van Dijk JMC, Jansen-van der Weide MC et al (2011) Intracranial Aneurysms in Patients with Subarachnoid Hemorrhage: CT Angiography as a Primary Examination Tool for Diagnosis—Systematic Review and Meta-Analysis. Radiology 258:134–145PubMedCrossRef Westerlaan HE, van Dijk JMC, Jansen-van der Weide MC et al (2011) Intracranial Aneurysms in Patients with Subarachnoid Hemorrhage: CT Angiography as a Primary Examination Tool for Diagnosis—Systematic Review and Meta-Analysis. Radiology 258:134–145PubMedCrossRef
47.
go back to reference Mannil M, Ramachandran J, Vittoria de Martini I et al (2017) Modified Dual-Energy Algorithm for Calcified Plaque Removal. Invest Radiol 52:680–685PubMedCrossRef Mannil M, Ramachandran J, Vittoria de Martini I et al (2017) Modified Dual-Energy Algorithm for Calcified Plaque Removal. Invest Radiol 52:680–685PubMedCrossRef
Metadata
Title
A Comparative Study of AI-Based Automated and Manual Postprocessing of Head and Neck CT Angiography: An Independent External Validation with Multi-Vendor and Multi-Center Data
Authors
Kunhua Li
Yang Yang
Shengwen Niu
Yongwei Yang
Bitong Tian
Xinyue Huan
Dajing Guo
Publication date
16-05-2024
Publisher
Springer Berlin Heidelberg
Published in
Neuroradiology
Print ISSN: 0028-3940
Electronic ISSN: 1432-1920
DOI
https://doi.org/10.1007/s00234-024-03379-y

Advances in Alzheimer's

Alzheimer's research and care is changing rapidly. Keep up with the latest developments from key international conferences, together with expert insights on how to integrate these advances into practice.

This content is intended for healthcare professionals outside of the UK.

Supported by:
  • Lilly
Developed by: Springer Healthcare IME
Learn more