Biomechanical analysis of a magnesium plantar plate prototype system for the first tarsometatarsal joint fusion: a cadaveric study

Authors: Peng Zhou, Marx Ribeiro, Johannes Greven, Maximilian Praster, Jan-Marten Seitz, Simon Habicht, Frank Hildebrand, Elizabeth R. Balmayor, Philipp Lichte

Published in: Journal of Orthopaedic Surgery and Research | Issue 1/2024

Login to get access

Abstract

Background

Titanium plantar plates have proven successful in the fixation of the first tarsometatarsal arthrodesis (TMT). However, a second surgery is typically needed for implant removal, and potential adverse effects, carried by the conventional implantations, are not uncommon. The purpose of this study was to determine whether a novel magnesium-based plantar plate system provides similar fusion stability to a titanium-based plantar plate system under various loading conditions.

Methods

Six matched-pair human cadaveric specimens underwent TMT fusions using either a magnesium plantar plate system prototype or a titanium plantar plate system. Specimens were cyclically loaded with a force ranging from 5 N to 50 N for 5,000 cycles, and displacement was recorded. Axial stiffness (N/mm) was calculated from load-displacement curves. Each specimen was loaded to failure at a rate of 5 mm/min, and the ultimate load was recorded.

Results

No significant difference was found in the vertical displacement between Ti group and Mg group after 100 cycles (2.4 ± 1.0 mm vs. 1.3 ± 1.4 mm, p = 0.196), 500 cycles (3.3 ± 1.3 mm vs. 1.7 ± 1.7 mm, p = 0.142), 1,000 cycles (3.7 ± 1.5 mm vs. 1.9 ± 1.9 mm, p = 0.128), 2,500 cycles (4.2 ± 1.7 mm vs. 2.3 ± 2.2 mm, p = 0.172) and 5,000 cycles (4.5 ± 1.8 mm vs. 2.3 ± 3.3 mm, p = 0.125), Additionally, no significant differences were observed in initial stiffness (53.1 ± 19.2 N/mm vs. 82.2 ± 53.9 N/mm, p = 0.257), final stiffness (90.6 ± 48.9 N/mm vs. 120.0 ± 48.3 N/mm, p = 0.319), or maximum load-to-failure (259.8 ± 98.2 N vs. 323.9 ± 134.9 N, p = 0.369).

Conclusions

Based on the performed biomechanical testing, the magnesium plantar plate system provides mechanical stability equivalent to the titanium plantar plate system in fixation for the first TMT joint fusion.
Literature
1.
go back to reference Albrecht GH. The pathology and treatment of hallux valgus. Russk Vrach. 2023;10:14–9. Albrecht GH. The pathology and treatment of hallux valgus. Russk Vrach. 2023;10:14–9.
2.
go back to reference Truslow W. Metatarsus primus varus or hallux valgus? J Bone Joint Surg. 1925;7:98–108. Truslow W. Metatarsus primus varus or hallux valgus? J Bone Joint Surg. 1925;7:98–108.
3.
go back to reference Lapidus PW. A quarter of a century of experience with the operative correction of the metatarsus varus primus in hallux valgus. Bull Hosp Joint Dis. 1956;17(2):404–21.PubMed Lapidus PW. A quarter of a century of experience with the operative correction of the metatarsus varus primus in hallux valgus. Bull Hosp Joint Dis. 1956;17(2):404–21.PubMed
4.
go back to reference Bednarz PA, Manoli A 2. Modified lapidus procedure for the treatment of hypermobile hallux valgus. Foot Ankle Int. 2000;21(10):816–21. Bednarz PA, Manoli A 2. Modified lapidus procedure for the treatment of hypermobile hallux valgus. Foot Ankle Int. 2000;21(10):816–21.
5.
go back to reference Fuhrmann RA. Arthrodesis of the first tarsometatarsal joint for correction of the advanced splayfoot accompanied by a hallux valgus. Operative Orthopadie Und Traumatologie. 2005;17(2):195–210.CrossRefPubMed Fuhrmann RA. Arthrodesis of the first tarsometatarsal joint for correction of the advanced splayfoot accompanied by a hallux valgus. Operative Orthopadie Und Traumatologie. 2005;17(2):195–210.CrossRefPubMed
6.
go back to reference Gilardino MS, Chen E, Bartlett SP. Choice of internal rigid fixation materials in the treatment of facial fractures. Craniomaxillofacial Trauma Reconstruction. 2009;2(1):49–60.CrossRefPubMedPubMedCentral Gilardino MS, Chen E, Bartlett SP. Choice of internal rigid fixation materials in the treatment of facial fractures. Craniomaxillofacial Trauma Reconstruction. 2009;2(1):49–60.CrossRefPubMedPubMedCentral
7.
go back to reference Waehner M, Klos K, Polzer H, Ray R, Lorchan Lewis T, Waizy H. Lapidus Arthrodesis for correction of Hallux Valgus deformity: a systematic review and Meta-analysis. Foot Ankle Spec. 2024;19386400241233832. Waehner M, Klos K, Polzer H, Ray R, Lorchan Lewis T, Waizy H. Lapidus Arthrodesis for correction of Hallux Valgus deformity: a systematic review and Meta-analysis. Foot Ankle Spec. 2024;19386400241233832.
8.
go back to reference Sukegawa S, Masui M, Sukegawa-Takahashi Y, Nakano K, Takabatake K, Kawai H, Nagatsuka H, Furuki Y. Maxillofacial trauma surgery patients with Titanium Osteosynthesis miniplates: remove or not? J Craniofac Surg. 2020;31(5):1338–42.CrossRefPubMed Sukegawa S, Masui M, Sukegawa-Takahashi Y, Nakano K, Takabatake K, Kawai H, Nagatsuka H, Furuki Y. Maxillofacial trauma surgery patients with Titanium Osteosynthesis miniplates: remove or not? J Craniofac Surg. 2020;31(5):1338–42.CrossRefPubMed
9.
go back to reference Ambrose CG, Clanton TO. Bioabsorbable implants: review of clinical experience in orthopedic surgery. Ann Biomed Eng. 2004;32(1):171–7.CrossRefPubMed Ambrose CG, Clanton TO. Bioabsorbable implants: review of clinical experience in orthopedic surgery. Ann Biomed Eng. 2004;32(1):171–7.CrossRefPubMed
10.
11.
go back to reference Willbold E, Weizbauer A, Loos A, Seitz JM, Angrisani N, Windhagen H, Reifenrath J. Magnesium alloys: a stony pathway from intensive research to clinical reality. Different test methods and approval-related considerations. J Biomed Mater Res A. 2017;105(1):329–47.CrossRefPubMed Willbold E, Weizbauer A, Loos A, Seitz JM, Angrisani N, Windhagen H, Reifenrath J. Magnesium alloys: a stony pathway from intensive research to clinical reality. Different test methods and approval-related considerations. J Biomed Mater Res A. 2017;105(1):329–47.CrossRefPubMed
12.
go back to reference Hussain M, Khan SM, Shafiq M, Abbas N, Sajjad U, Hamid K. Advances in biodegradable materials: degradation mechanisms, mechanical properties, and biocompatibility for orthopedic applications. Heliyon. 2024;10(12):e32713. Hussain M, Khan SM, Shafiq M, Abbas N, Sajjad U, Hamid K. Advances in biodegradable materials: degradation mechanisms, mechanical properties, and biocompatibility for orthopedic applications. Heliyon. 2024;10(12):e32713.
13.
go back to reference Wang JL, Xu JK, Hopkins C, Chow DH, Qin L. Biodegradable magnesium-based implants in Orthopedics-A General Review and perspectives. Adv Sci (Weinheim Baden-Wurttemberg Germany). 2020;7(8):1902443. Wang JL, Xu JK, Hopkins C, Chow DH, Qin L. Biodegradable magnesium-based implants in Orthopedics-A General Review and perspectives. Adv Sci (Weinheim Baden-Wurttemberg Germany). 2020;7(8):1902443.
14.
go back to reference Zhang Y, Xu J, Ruan YC, Yu MK, O’Laughlin M, Wise H, Chen D, Tian L, Shi D, Wang J, Chen S, Feng JQ, Chow DH, Xie X, Zheng L, Huang L, Huang S, Leung K, Lu N, Zhao L, Li H, Zhao D, Guo X, Chan K, Witte F, Chan HC, Zheng Y, Qin L. Implant-derived magnesium induces local neuronal production of CGRP to improve bone-fracture healing in rats. Nat Med. 2016;22(10):1160–9.CrossRefPubMedPubMedCentral Zhang Y, Xu J, Ruan YC, Yu MK, O’Laughlin M, Wise H, Chen D, Tian L, Shi D, Wang J, Chen S, Feng JQ, Chow DH, Xie X, Zheng L, Huang L, Huang S, Leung K, Lu N, Zhao L, Li H, Zhao D, Guo X, Chan K, Witte F, Chan HC, Zheng Y, Qin L. Implant-derived magnesium induces local neuronal production of CGRP to improve bone-fracture healing in rats. Nat Med. 2016;22(10):1160–9.CrossRefPubMedPubMedCentral
15.
go back to reference Wang J, Xu J, Song B, Chow DH, Shu-Hang Yung P, Qin L. Magnesium (mg) based interference screws developed for promoting tendon graft incorporation in bone tunnel in rabbits. Acta Biomater. 2017;63:393–410.CrossRefPubMed Wang J, Xu J, Song B, Chow DH, Shu-Hang Yung P, Qin L. Magnesium (mg) based interference screws developed for promoting tendon graft incorporation in bone tunnel in rabbits. Acta Biomater. 2017;63:393–410.CrossRefPubMed
16.
go back to reference Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am Vol. 2011;93(11):1057–63.CrossRef Schreiber JJ, Anderson PA, Rosas HG, Buchholz AL, Au AG. Hounsfield units for assessing bone mineral density and strength: a tool for osteoporosis management. J Bone Joint Surg Am Vol. 2011;93(11):1057–63.CrossRef
17.
go back to reference Burchard R, Massa R, Soost C, Richter W, Dietrich G, Ohrndorf A, Christ HJ, Fritzen CP, Graw JA, Schmitt J. Biomechanics of common fixation devices for first tarsometatarsal joint fusion-a comparative study with synthetic bones. J Orthop Surg Res. 2018;13(1):176.CrossRefPubMedPubMedCentral Burchard R, Massa R, Soost C, Richter W, Dietrich G, Ohrndorf A, Christ HJ, Fritzen CP, Graw JA, Schmitt J. Biomechanics of common fixation devices for first tarsometatarsal joint fusion-a comparative study with synthetic bones. J Orthop Surg Res. 2018;13(1):176.CrossRefPubMedPubMedCentral
18.
go back to reference Kang H. Sample size determination and power analysis using the G*Power software. J Educational Evaluation Health Professions. 2021;18:17.CrossRef Kang H. Sample size determination and power analysis using the G*Power software. J Educational Evaluation Health Professions. 2021;18:17.CrossRef
19.
go back to reference Switaj PJ, Fuchs D, Alshouli M, Patwardhan AG, Voronov LI, Muriuki M, Havey RM, Kadakia AR. A biomechanical comparison study of a modern fibular nail and distal fibular locking plate in AO/OTA 44C2 ankle fractures. J Orthop Surg Res. 2016;11(1):100.CrossRefPubMedPubMedCentral Switaj PJ, Fuchs D, Alshouli M, Patwardhan AG, Voronov LI, Muriuki M, Havey RM, Kadakia AR. A biomechanical comparison study of a modern fibular nail and distal fibular locking plate in AO/OTA 44C2 ankle fractures. J Orthop Surg Res. 2016;11(1):100.CrossRefPubMedPubMedCentral
20.
go back to reference Stowers JM, Black AT, Kavanagh AM, Mata K, Eshetu T. Bone density by specific anatomic location using Hounsfield samples of the Lower extremity: a multi-center retrospective analysis. J foot Ankle Surgery: Official Publication Am Coll Foot Ankle Surg. 2023;62(1):80–4.CrossRef Stowers JM, Black AT, Kavanagh AM, Mata K, Eshetu T. Bone density by specific anatomic location using Hounsfield samples of the Lower extremity: a multi-center retrospective analysis. J foot Ankle Surgery: Official Publication Am Coll Foot Ankle Surg. 2023;62(1):80–4.CrossRef
21.
go back to reference Drummond D, Motley T, Kosmopoulos V, Ernst J. Stability of Locking Plate and Compression Screws for Lapidus Arthrodesis: a Biomechanical comparison of plate position. J foot Ankle Surgery: Official Publication Am Coll Foot Ankle Surg. 2018;57(3):466–70.CrossRef Drummond D, Motley T, Kosmopoulos V, Ernst J. Stability of Locking Plate and Compression Screws for Lapidus Arthrodesis: a Biomechanical comparison of plate position. J foot Ankle Surgery: Official Publication Am Coll Foot Ankle Surg. 2018;57(3):466–70.CrossRef
22.
go back to reference Klos K, Simons P, Schopp P, Schenk P, Kohler FC, Uddin A, Roth EK, Biedermann U, Hofmann GO, Lenz M. Biomechanical comparison of medio-plantar and plantar plate fixation for first tarsometatarsal joint arthrodesis. J Clin Med. 2023;12(12). Klos K, Simons P, Schopp P, Schenk P, Kohler FC, Uddin A, Roth EK, Biedermann U, Hofmann GO, Lenz M. Biomechanical comparison of medio-plantar and plantar plate fixation for first tarsometatarsal joint arthrodesis. J Clin Med. 2023;12(12).
23.
go back to reference Knutsen AR, Fleming JF, Ebramzadeh E, Ho NC, Warganich T, Harris TG, Sangiorgio SN. Biomechanical comparison of fixation devices for first metatarsocuneiform joint arthrodesis. Foot Ankle Spec. 2017;10(4):322–8.CrossRefPubMed Knutsen AR, Fleming JF, Ebramzadeh E, Ho NC, Warganich T, Harris TG, Sangiorgio SN. Biomechanical comparison of fixation devices for first metatarsocuneiform joint arthrodesis. Foot Ankle Spec. 2017;10(4):322–8.CrossRefPubMed
24.
go back to reference Klos K, Simons P, Hajduk AS, Hoffmeier KL, Gras F, Fröber R, Hofmann GO, Mückley T. Plantar versus dorsomedial locked plating for Lapidus arthrodesis: a biomechanical comparison. Foot Ankle Int. 2011;32(11):1081–5.CrossRefPubMed Klos K, Simons P, Hajduk AS, Hoffmeier KL, Gras F, Fröber R, Hofmann GO, Mückley T. Plantar versus dorsomedial locked plating for Lapidus arthrodesis: a biomechanical comparison. Foot Ankle Int. 2011;32(11):1081–5.CrossRefPubMed
25.
go back to reference Witte F. The history of biodegradable magnesium implants: a review. Acta Biomater. 2010;6(5):1680–92.CrossRefPubMed Witte F. The history of biodegradable magnesium implants: a review. Acta Biomater. 2010;6(5):1680–92.CrossRefPubMed
26.
go back to reference Staiger MP, Pietak AM, Huadmai J, Dias G. Magnesium and its alloys as orthopedic biomaterials: a review. Biomaterials. 2006;27(9):1728–34.CrossRefPubMed Staiger MP, Pietak AM, Huadmai J, Dias G. Magnesium and its alloys as orthopedic biomaterials: a review. Biomaterials. 2006;27(9):1728–34.CrossRefPubMed
27.
go back to reference Castellani C, Lindtner RA, Hausbrandt P, Tschegg E, Stanzl-Tschegg SE, Zanoni G, Beck S, Weinberg AM. Bone-implant interface strength and osseointegration: biodegradable magnesium alloy versus standard titanium control. Acta Biomater. 2011;7(1):432–40.CrossRefPubMed Castellani C, Lindtner RA, Hausbrandt P, Tschegg E, Stanzl-Tschegg SE, Zanoni G, Beck S, Weinberg AM. Bone-implant interface strength and osseointegration: biodegradable magnesium alloy versus standard titanium control. Acta Biomater. 2011;7(1):432–40.CrossRefPubMed
28.
go back to reference Fischer H, Schmidt-Bleek O, Orassi V, Wulsten D, Schmidt-Bleek K, Heiland M, Steffen C, Rendenbach C. Biomechanical comparison of WE43-based magnesium vs. titanium miniplates in a mandible fracture model in sheep. Materials (Basel, Switzerland). 2022;16(1). Fischer H, Schmidt-Bleek O, Orassi V, Wulsten D, Schmidt-Bleek K, Heiland M, Steffen C, Rendenbach C. Biomechanical comparison of WE43-based magnesium vs. titanium miniplates in a mandible fracture model in sheep. Materials (Basel, Switzerland). 2022;16(1).
29.
go back to reference Orassi V, Fischer H, Duda GN, Heiland M, Checa S, Rendenbach C. In Silico Biomechanical evaluation of WE43 magnesium plates for mandibular fracture fixation. Front Bioeng Biotechnol. 2021;9:803103.CrossRefPubMed Orassi V, Fischer H, Duda GN, Heiland M, Checa S, Rendenbach C. In Silico Biomechanical evaluation of WE43 magnesium plates for mandibular fracture fixation. Front Bioeng Biotechnol. 2021;9:803103.CrossRefPubMed
30.
go back to reference Kraus T, Fischerauer S, Treichler S, Martinelli E, Eichler J, Myrissa A, Zötsch S, Uggowitzer PJ, Löffler JF, Weinberg AM. The influence of biodegradable magnesium implants on the growth plate. Acta Biomater. 2018;66:109–17.CrossRefPubMed Kraus T, Fischerauer S, Treichler S, Martinelli E, Eichler J, Myrissa A, Zötsch S, Uggowitzer PJ, Löffler JF, Weinberg AM. The influence of biodegradable magnesium implants on the growth plate. Acta Biomater. 2018;66:109–17.CrossRefPubMed
31.
go back to reference Atkinson HD, Khan S, Lashgari Y, Ziegler A. Hallux valgus correction utilising a modified short scarf osteotomy with a magnesium biodegradable or titanium compression screws - a comparative study of clinical outcomes. BMC Musculoskelet Disord. 2019;20(1):334.CrossRefPubMedPubMedCentral Atkinson HD, Khan S, Lashgari Y, Ziegler A. Hallux valgus correction utilising a modified short scarf osteotomy with a magnesium biodegradable or titanium compression screws - a comparative study of clinical outcomes. BMC Musculoskelet Disord. 2019;20(1):334.CrossRefPubMedPubMedCentral
32.
go back to reference Windhagen H, Radtke K, Weizbauer A, Diekmann J, Noll Y, Kreimeyer U, Schavan R, Stukenborg-Colsman C, Waizy H. Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study. Biomed Eng Online. 2013;12:62.CrossRefPubMedPubMedCentral Windhagen H, Radtke K, Weizbauer A, Diekmann J, Noll Y, Kreimeyer U, Schavan R, Stukenborg-Colsman C, Waizy H. Biodegradable magnesium-based screw clinically equivalent to titanium screw in hallux valgus surgery: short term results of the first prospective, randomized, controlled clinical pilot study. Biomed Eng Online. 2013;12:62.CrossRefPubMedPubMedCentral
33.
go back to reference Plaass C, von Falck C, Ettinger S, Sonnow L, Calderone F, Weizbauer A, Reifenrath J, Claassen L, Waizy H, Daniilidis K, Stukenborg-Colsman C, Windhagen H. Bioabsorbable magnesium versus standard titanium compression screws for fixation of distal metatarsal osteotomies – 3 year results of a randomized clinical trial. J Orthop Science: Official J Japanese Orthop Association. 2018;23(2):321–7.CrossRef Plaass C, von Falck C, Ettinger S, Sonnow L, Calderone F, Weizbauer A, Reifenrath J, Claassen L, Waizy H, Daniilidis K, Stukenborg-Colsman C, Windhagen H. Bioabsorbable magnesium versus standard titanium compression screws for fixation of distal metatarsal osteotomies – 3 year results of a randomized clinical trial. J Orthop Science: Official J Japanese Orthop Association. 2018;23(2):321–7.CrossRef
34.
go back to reference Yu X, Zhao D, Huang S, Wang B, Zhang X, Wang W, Wei X. Biodegradable magnesium screws and vascularized iliac grafting for displaced femoral neck fracture in young adults. BMC Musculoskelet Disord. 2015;16:329.CrossRefPubMedPubMedCentral Yu X, Zhao D, Huang S, Wang B, Zhang X, Wang W, Wei X. Biodegradable magnesium screws and vascularized iliac grafting for displaced femoral neck fracture in young adults. BMC Musculoskelet Disord. 2015;16:329.CrossRefPubMedPubMedCentral
35.
go back to reference Antoniac I, Manescu Paltanea V, Antoniac A, Paltanea G. Magnesium-based alloys with adapted interfaces for bone implants and tissue engineering. Regenerative Biomaterials. 2023;10:rbad095.CrossRefPubMedPubMedCentral Antoniac I, Manescu Paltanea V, Antoniac A, Paltanea G. Magnesium-based alloys with adapted interfaces for bone implants and tissue engineering. Regenerative Biomaterials. 2023;10:rbad095.CrossRefPubMedPubMedCentral
36.
go back to reference Fu X, Wang GX, Wang CG, Li ZJ. Comparison between bioabsorbable magnesium and titanium compression screws for hallux valgus treated with distal metatarsal osteotomies: a meta-analysis. Joint Dis Relat Surg. 2023;34(2):289–97.CrossRef Fu X, Wang GX, Wang CG, Li ZJ. Comparison between bioabsorbable magnesium and titanium compression screws for hallux valgus treated with distal metatarsal osteotomies: a meta-analysis. Joint Dis Relat Surg. 2023;34(2):289–97.CrossRef
37.
go back to reference Hänzi AC, Gerber I, Schinhammer M, Löffler JF, Uggowitzer PJ. On the in vitro and in vivo degradation performance and biological response of new biodegradable Mg-Y-Zn alloys. Acta Biomater. 2010;6(5):1824–33.CrossRefPubMed Hänzi AC, Gerber I, Schinhammer M, Löffler JF, Uggowitzer PJ. On the in vitro and in vivo degradation performance and biological response of new biodegradable Mg-Y-Zn alloys. Acta Biomater. 2010;6(5):1824–33.CrossRefPubMed
38.
go back to reference Chakraborty Banerjee P, Al-Saadi S, Choudhary L, Harandi SE, Singh R. Magnesium implants: prospects and challenges. Materials (Basel, Switzerland). 2019;12(1). Chakraborty Banerjee P, Al-Saadi S, Choudhary L, Harandi SE, Singh R. Magnesium implants: prospects and challenges. Materials (Basel, Switzerland). 2019;12(1).
Metadata
Title
Biomechanical analysis of a magnesium plantar plate prototype system for the first tarsometatarsal joint fusion: a cadaveric study
Authors
Peng Zhou
Marx Ribeiro
Johannes Greven
Maximilian Praster
Jan-Marten Seitz
Simon Habicht
Frank Hildebrand
Elizabeth R. Balmayor
Philipp Lichte
Publication date
01-12-2024
Publisher
BioMed Central
Published in
Journal of Orthopaedic Surgery and Research / Issue 1/2024
Electronic ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-024-05208-7