Skip to main content
Top
Published in:

28-06-2023 | Aneurysm | Review Article

Early Outcomes of the Pipeline Vantage Flow Diverter

A Multicentre Study

Authors: Sara Sciacca, Ahmed Bassiouny, Nina Mansoor, Thais Minett, Parthiban Balasundaram, Juveria Siddiqui, Yogish Joshi, Shahram Derakhshani, Naga Kandasamy, Thomas Calvert Booth, Jeremy Lynch

Published in: Clinical Neuroradiology | Issue 4/2023

Login to get access

Abstract

Purpose

The recently introduced Pipeline Vantage Embolization Device with Shield Technology is the fourth generation of Pipeline flow diverter devices. Due to the relatively high rate of intraprocedural technical complications, modifications were subsequently made to the device after a limited release of the device in 2020. This study aimed to evaluate the safety and efficacy of the modified version of this device.

Methods

This was a multicentre retrospective series. The primary efficacy endpoint was aneurysm occlusion in the absence of retreatment. The primary safety endpoint was any neurological morbidity or death. Ruptured and unruptured aneurysms were included in the study.

Results

A total of 52 procedures were performed for 60 target aneurysms. Treatment was performed on 5 patients with ruptured aneurysms. The technical success rate was 98%. The mean clinical follow-up time was 5.5 months. In patients presenting with unruptured aneurysms there were no deaths, 3 (6.4%) major complications and 7 (13%) minor complications. In the five patients presenting with subarachnoid haemorrhage there were 2 (40%) major complications with 1 (20%) of these resulting in death, and 1 (20%) minor complication. Of the patients 29 (56%) had undergone 6‑monthly postprocedural angiographic imaging with a mean time of 6.6 months demonstrating that 83% of patients had achieved adequate occlusion (RROC1/2) of the aneurysm.

Conclusions

In this non-industry-sponsored study, the occlusion rates and safety outcomes were similar to those seen in previously published studies with flow diverter devices and earlier generation Pipeline devices. Modifications to the device appear to have improved ease of deployment.
Literature
1.
go back to reference Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D. The pipeline embolization device for the intracranial treatment of aneurysms trial. Ajnr Am J Neuroradiol. 2011;32:34–40.CrossRefPubMedPubMedCentral Nelson PK, Lylyk P, Szikora I, Wetzel SG, Wanke I, Fiorella D. The pipeline embolization device for the intracranial treatment of aneurysms trial. Ajnr Am J Neuroradiol. 2011;32:34–40.CrossRefPubMedPubMedCentral
2.
go back to reference Kallmes DF, Hanel R, Lopes D, Boccardi E, Bonafé A, Cekirge S, Fiorella D, Jabbour P, Levy E, McDougall C, Siddiqui A, Szikora I, Woo H, Albuquerque F, Bozorgchami H, Dashti SR, Delgado Almandoz JE, Kelly ME, Turner R, Woodward BK, Brinjikji W, Lanzino G, Lylykr P. International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study. AJNR Am J Neuroradiol. 2015;36:108–15.CrossRefPubMedPubMedCentral Kallmes DF, Hanel R, Lopes D, Boccardi E, Bonafé A, Cekirge S, Fiorella D, Jabbour P, Levy E, McDougall C, Siddiqui A, Szikora I, Woo H, Albuquerque F, Bozorgchami H, Dashti SR, Delgado Almandoz JE, Kelly ME, Turner R, Woodward BK, Brinjikji W, Lanzino G, Lylykr P. International retrospective study of the pipeline embolization device: a multicenter aneurysm treatment study. AJNR Am J Neuroradiol. 2015;36:108–15.CrossRefPubMedPubMedCentral
3.
go back to reference Martínez-Galdámez M, Lamin SM, Lagios KG, Liebig T, Ciceri EF, Chapot R, Stockx L, Chavda S, Kabbasch C, Faragò G, Nordmeyer H, Boulanger T, Piano M, Boccardi EP. Treatment of intracranial aneurysms using the pipeline flex embolization device with shield technology: angiographic and safety outcomes at 1‑year follow-up. J Neurointerv Surg. 2019;11:396–9.CrossRefPubMed Martínez-Galdámez M, Lamin SM, Lagios KG, Liebig T, Ciceri EF, Chapot R, Stockx L, Chavda S, Kabbasch C, Faragò G, Nordmeyer H, Boulanger T, Piano M, Boccardi EP. Treatment of intracranial aneurysms using the pipeline flex embolization device with shield technology: angiographic and safety outcomes at 1‑year follow-up. J Neurointerv Surg. 2019;11:396–9.CrossRefPubMed
4.
go back to reference Pereira VM, Kelly M, Vega P, Murias E, Yilmaz H, Erceg G, Pellaton A, Lovblad K, Krings T. New pipeline flex device: initial experience and technical nuances. J Neurointerv Surg. 2015;7:920–5.CrossRefPubMed Pereira VM, Kelly M, Vega P, Murias E, Yilmaz H, Erceg G, Pellaton A, Lovblad K, Krings T. New pipeline flex device: initial experience and technical nuances. J Neurointerv Surg. 2015;7:920–5.CrossRefPubMed
5.
go back to reference Atasoy D, Kandasamy N, Hart J, Lynch J, Yang SH, Walsh D, Tolias C, Booth TC. Outcome study of the pipeline embolization device with shield technology in unruptured aneurysms (PEDSU). AJNR Am J Neuroradiol. 2019;40:2094–101.PubMedPubMedCentral Atasoy D, Kandasamy N, Hart J, Lynch J, Yang SH, Walsh D, Tolias C, Booth TC. Outcome study of the pipeline embolization device with shield technology in unruptured aneurysms (PEDSU). AJNR Am J Neuroradiol. 2019;40:2094–101.PubMedPubMedCentral
6.
go back to reference Vollherbst DF, Cekirge HS, Saatci I, Baltacioglu F, Onal B, Koc O, Rautio R, Sinisalo M, Tomasello A, Vega P, Martínez-Galdámez M, Lynch J, Mendes Pereira V, Bendszus M, Möhlenbruch MA. First clinical multicenter experience with the new pipeline vantage flow diverter. J Neurointerv Surg. 2022;15(1):63–9.CrossRefPubMed Vollherbst DF, Cekirge HS, Saatci I, Baltacioglu F, Onal B, Koc O, Rautio R, Sinisalo M, Tomasello A, Vega P, Martínez-Galdámez M, Lynch J, Mendes Pereira V, Bendszus M, Möhlenbruch MA. First clinical multicenter experience with the new pipeline vantage flow diverter. J Neurointerv Surg. 2022;15(1):63–9.CrossRefPubMed
7.
go back to reference Wilson JTL, Hareendran A, Hendry A, Potter J, Bone I, Muir KW. Reliability of the modified Rankin scale across multiple raters: benefits of a structured interview. Stroke. 2005;36:777–81.CrossRefPubMed Wilson JTL, Hareendran A, Hendry A, Potter J, Bone I, Muir KW. Reliability of the modified Rankin scale across multiple raters: benefits of a structured interview. Stroke. 2005;36:777–81.CrossRefPubMed
8.
go back to reference Roy D, Milot G, Raymond J. Endovascular treatment of unruptured aneurysms. Stroke. 2001;32:1998–2004.CrossRefPubMed Roy D, Milot G, Raymond J. Endovascular treatment of unruptured aneurysms. Stroke. 2001;32:1998–2004.CrossRefPubMed
9.
go back to reference O’kelly CJ, Krings T, Fiorella D, Marotta TR. A novel grading scale for the angiographic assessment of intracranial aneurysms treated using flow diverting stents. Interv Neuroradiol. 2010;16:133–7.CrossRefPubMedPubMedCentral O’kelly CJ, Krings T, Fiorella D, Marotta TR. A novel grading scale for the angiographic assessment of intracranial aneurysms treated using flow diverting stents. Interv Neuroradiol. 2010;16:133–7.CrossRefPubMedPubMedCentral
10.
go back to reference Hanel RA, Kallmes DF, Lopes DK, Nelson PK, Siddiqui A, Jabbour P, Pereira VM, Szikora I, Zaidat OO, Bettegowda C, Colby GP, Mokin M, Schirmer C, et al. Prospective study on embolization of intracranial aneurysms with the pipeline device: the PREMIER study 1 year results. J Neurointerv Surg. 2020;12:62–6.CrossRefPubMed Hanel RA, Kallmes DF, Lopes DK, Nelson PK, Siddiqui A, Jabbour P, Pereira VM, Szikora I, Zaidat OO, Bettegowda C, Colby GP, Mokin M, Schirmer C, et al. Prospective study on embolization of intracranial aneurysms with the pipeline device: the PREMIER study 1 year results. J Neurointerv Surg. 2020;12:62–6.CrossRefPubMed
11.
go back to reference Becske T, Kallmes DF, Saatci I, McDougall CG, Szikora I, Lanzino G, Moran CJ, Woo HH, Lopes DK, Berez AL, Cher DJ, Siddiqui AH, Levy EI, Albuquerque FC, Fiorella DJ, Berentei Z, Marosfoi M, Cekirge SH, Nelson PK. Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology. 2013;267:858–68.CrossRefPubMed Becske T, Kallmes DF, Saatci I, McDougall CG, Szikora I, Lanzino G, Moran CJ, Woo HH, Lopes DK, Berez AL, Cher DJ, Siddiqui AH, Levy EI, Albuquerque FC, Fiorella DJ, Berentei Z, Marosfoi M, Cekirge SH, Nelson PK. Pipeline for uncoilable or failed aneurysms: results from a multicenter clinical trial. Radiology. 2013;267:858–68.CrossRefPubMed
12.
go back to reference Ebisudani Y, Sugiu K, Haruma J, Murai S, Hishikawa T, Hiramatsu M, Takahashi Y, Nishi K, Yamaoka Y, Sato Y, Kimura R, Edaki H, Date I. A case of delayed foreshortening and disconnection of pipeline embolization device detected by chronological cone-beam computed tomography fusion image. Acta Neurochir (Wien). 2022;164:2203–6.CrossRefPubMed Ebisudani Y, Sugiu K, Haruma J, Murai S, Hishikawa T, Hiramatsu M, Takahashi Y, Nishi K, Yamaoka Y, Sato Y, Kimura R, Edaki H, Date I. A case of delayed foreshortening and disconnection of pipeline embolization device detected by chronological cone-beam computed tomography fusion image. Acta Neurochir (Wien). 2022;164:2203–6.CrossRefPubMed
13.
go back to reference Heit JJ, Telischak NA, Do HM, Dodd RL, Steinberg GK, Marks MP. Pipeline embolization device retraction and foreshortening after internal carotid artery blister aneurysm treatment. Interv Neuroradiol. 2017;23:614–9.CrossRefPubMedPubMedCentral Heit JJ, Telischak NA, Do HM, Dodd RL, Steinberg GK, Marks MP. Pipeline embolization device retraction and foreshortening after internal carotid artery blister aneurysm treatment. Interv Neuroradiol. 2017;23:614–9.CrossRefPubMedPubMedCentral
14.
go back to reference Kallmes DF, Brinjikji W, Cekirge S, Fiorella D, Hanel RA, Jabbour P, Lopes D, Lylyk P, McDougall CG, Siddiqui A. Safety and efficacy of the pipeline embolization device for treatment of intracranial aneurysms: a pooled analysis of 3 large studies. J Neurosurg. 2017;127:775–80.CrossRefPubMed Kallmes DF, Brinjikji W, Cekirge S, Fiorella D, Hanel RA, Jabbour P, Lopes D, Lylyk P, McDougall CG, Siddiqui A. Safety and efficacy of the pipeline embolization device for treatment of intracranial aneurysms: a pooled analysis of 3 large studies. J Neurosurg. 2017;127:775–80.CrossRefPubMed
15.
go back to reference Pikis S, Mantziaris G, Mamalis V, Barkas K, Tsanis A, Lyra S, Karkoulias K, Petrosyan T, Archontakis E. Diffusion weighted image documented cerebral ischemia in the postprocedural period following pipeline embolization device with shield technology treatment of unruptured intracranial aneurysms: a prospective, single center study. J Neurointerv Surg. 2020;12:407–11.CrossRefPubMed Pikis S, Mantziaris G, Mamalis V, Barkas K, Tsanis A, Lyra S, Karkoulias K, Petrosyan T, Archontakis E. Diffusion weighted image documented cerebral ischemia in the postprocedural period following pipeline embolization device with shield technology treatment of unruptured intracranial aneurysms: a prospective, single center study. J Neurointerv Surg. 2020;12:407–11.CrossRefPubMed
16.
go back to reference Alpay K, Hinkka T, Lindgren AE, Isokangas JM, Raj R, Parkkola R, Sinisalo M, Numminen J, Pienimäki JP, Saari P, Seppänen J, Palosaari K, Rautio R. Flow diversion treatment for acutely ruptured aneurysms. J Neurointerv Surg. 2020;12:283–8.CrossRef Alpay K, Hinkka T, Lindgren AE, Isokangas JM, Raj R, Parkkola R, Sinisalo M, Numminen J, Pienimäki JP, Saari P, Seppänen J, Palosaari K, Rautio R. Flow diversion treatment for acutely ruptured aneurysms. J Neurointerv Surg. 2020;12:283–8.CrossRef
17.
go back to reference Maybaum J, Henkes H, Aguilar-Pérez M, Hellstern V, Gihr GA, Härtig W, Reisberg A, Mucha D, Schüngel MS, Brill R, Quäschling U, Hoffmann KT, Schob S. Flow diversion for reconstruction of Intradural vertebral artery dissecting aneurysms causing subarachnoid hemorrhage—a retrospective study from four neurovascular centers. Front Neurol. 2021;12:700164.CrossRefPubMedPubMedCentral Maybaum J, Henkes H, Aguilar-Pérez M, Hellstern V, Gihr GA, Härtig W, Reisberg A, Mucha D, Schüngel MS, Brill R, Quäschling U, Hoffmann KT, Schob S. Flow diversion for reconstruction of Intradural vertebral artery dissecting aneurysms causing subarachnoid hemorrhage—a retrospective study from four neurovascular centers. Front Neurol. 2021;12:700164.CrossRefPubMedPubMedCentral
18.
go back to reference Chalouhi N, Polifka A, Daou B, Kung D, Barros G, Tjoumakaris S, Gonzalez LF, et al. In-pipeline stenosis: incidence, predictors, and clinical outcomes. Neurosurgery. 2015;77:875–9. discussion 879.CrossRefPubMed Chalouhi N, Polifka A, Daou B, Kung D, Barros G, Tjoumakaris S, Gonzalez LF, et al. In-pipeline stenosis: incidence, predictors, and clinical outcomes. Neurosurgery. 2015;77:875–9. discussion 879.CrossRefPubMed
19.
go back to reference Wall A, Kugel H, Bachman R, Matuszewski L, Krämer S, Heindel W, Maintz D. 3.0 T vs. 1.5 T MR angiography: in vitro comparison of intravascular stent artifacts. J Magn Reson Imaging. 2005;22:772–9.CrossRefPubMed Wall A, Kugel H, Bachman R, Matuszewski L, Krämer S, Heindel W, Maintz D. 3.0 T vs. 1.5 T MR angiography: in vitro comparison of intravascular stent artifacts. J Magn Reson Imaging. 2005;22:772–9.CrossRefPubMed
20.
go back to reference Wang Y, Truong TN, Yen C, Bilecen D, Watts R, Trost DW, Prince MR. Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents. Magn Reson Med. 2003;49:972–6.CrossRefPubMed Wang Y, Truong TN, Yen C, Bilecen D, Watts R, Trost DW, Prince MR. Quantitative evaluation of susceptibility and shielding effects of nitinol, platinum, cobalt-alloy, and stainless steel stents. Magn Reson Med. 2003;49:972–6.CrossRefPubMed
21.
go back to reference Blum MB, Schmook M, Schernthaner R, Edelhauser G, Puchner S, Lammer J, Funovics MA. Quantification and detectability of in-stent stenosis with CT angiography and MR angiography in arterial stents in vitro. AJR Am J Roentgenol. 2007;189:1238–42.CrossRefPubMed Blum MB, Schmook M, Schernthaner R, Edelhauser G, Puchner S, Lammer J, Funovics MA. Quantification and detectability of in-stent stenosis with CT angiography and MR angiography in arterial stents in vitro. AJR Am J Roentgenol. 2007;189:1238–42.CrossRefPubMed
22.
go back to reference Starke RM, Thompson J, Pagani A, Choubey A, Wainwright JM, Wolf MF, Jahanbekam R, Girdhar G. Preclinical safety and efficacy evaluation of the pipeline vantage embolization device with shield technology. J Neurointerv Surg. 2020;12:981–6.CrossRefPubMedPubMedCentral Starke RM, Thompson J, Pagani A, Choubey A, Wainwright JM, Wolf MF, Jahanbekam R, Girdhar G. Preclinical safety and efficacy evaluation of the pipeline vantage embolization device with shield technology. J Neurointerv Surg. 2020;12:981–6.CrossRefPubMedPubMedCentral
23.
go back to reference Brinjikji W, Murad MH, Lanzino G, Cloft HJ, Kallmes DF. Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis. Stroke. 2013;44:442–7.CrossRefPubMed Brinjikji W, Murad MH, Lanzino G, Cloft HJ, Kallmes DF. Endovascular treatment of intracranial aneurysms with flow diverters: a meta-analysis. Stroke. 2013;44:442–7.CrossRefPubMed
24.
go back to reference Yang S‑H, Hampton T, Kandasamy N, Hart J, Ashmore J, Walsh DC, Tolias C, Booth TC. Outcome study of the pipeline embolization device for treatment of intracranial aneurysms at a single UK institution. Br J Neurosurg. 2017;31:661–7.CrossRefPubMed Yang S‑H, Hampton T, Kandasamy N, Hart J, Ashmore J, Walsh DC, Tolias C, Booth TC. Outcome study of the pipeline embolization device for treatment of intracranial aneurysms at a single UK institution. Br J Neurosurg. 2017;31:661–7.CrossRefPubMed
25.
go back to reference Ahmed SU, Mocco J, Zhang X, Kelly M, Doshi A, Nael K, De Leacy R. MRA versus DSA for the follow-up imaging of intracranial aneurysms treated using endovascular techniques: a meta-analysis. J Neurointerv Surg. 2019;11:1009–14.CrossRefPubMed Ahmed SU, Mocco J, Zhang X, Kelly M, Doshi A, Nael K, De Leacy R. MRA versus DSA for the follow-up imaging of intracranial aneurysms treated using endovascular techniques: a meta-analysis. J Neurointerv Surg. 2019;11:1009–14.CrossRefPubMed
Metadata
Title
Early Outcomes of the Pipeline Vantage Flow Diverter
A Multicentre Study
Authors
Sara Sciacca
Ahmed Bassiouny
Nina Mansoor
Thais Minett
Parthiban Balasundaram
Juveria Siddiqui
Yogish Joshi
Shahram Derakhshani
Naga Kandasamy
Thomas Calvert Booth
Jeremy Lynch
Publication date
28-06-2023
Publisher
Springer Berlin Heidelberg
Published in
Clinical Neuroradiology / Issue 4/2023
Print ISSN: 1869-1439
Electronic ISSN: 1869-1447
DOI
https://doi.org/10.1007/s00062-023-01314-x

Other articles of this Issue 4/2023

Clinical Neuroradiology 4/2023 Go to the issue